Exemplary performance data for immunosuppressants

Technical notes | 2020 | Thermo Fisher ScientificInstrumentation
LC/MS, LC/MS/MS, LC/QQQ
Industries
Clinical Research
Manufacturer
Thermo Fisher Scientific

Summary

Significance of the Topic


Accurate measurement of immunosuppressant drugs such as cyclosporin A, everolimus, sirolimus and tacrolimus is critical for patient management in transplantation and autoimmune therapy. Liquid chromatography tandem mass spectrometry (LC-MS/MS) offers high specificity, sensitivity and throughput required for reliable in vitro diagnostics in clinical laboratories.

Objectives and Study Overview


This study evaluates the analytical performance of the Thermo Scientific Vanquish MD HPLC coupled with the TSQ Quantis MD triple-stage quadrupole mass spectrometer across multiple instrument combinations. The aim is to demonstrate linearity, accuracy and precision for whole blood quantification of four key immunosuppressants under conditions suitable for laboratory-developed test workflows.

Methodology and Instrumentation Used


Sample Preparation
  • Protein precipitation using zinc sulfate followed by transfer of supernatant to HPLC vials.
Chromatographic Conditions
  • Mobile phase A: 0.1% formic acid and 10 mM ammonium formate in water.
  • Mobile phase B: 0.1% formic acid and 10 mM ammonium formate in methanol.
Mass Spectrometry Settings
  • Run time: 3 minutes per injection.
  • Ionization: positive polarity heated electrospray ionization (HESI).
  • Detection: selected reaction monitoring (SRM) transitions with quantifier and qualifier ions.
Instrumentation
  • Vanquish MD HPLC system (Thermo Scientific) approved for in vitro diagnostic use.
  • TSQ Quantis MD Series triple quadrupole mass spectrometer (Thermo Scientific).
  • TraceFinder LDT software for data processing and quantitation.

Main Results and Discussion


Calibration and Linearity
  • Six-point calibration curves for each analyte showed correlation coefficients (R²) above 0.997 across all instrument pairings.
  • Deviation from assigned calibrator values remained below ±15%, meeting predefined acceptance criteria.
Precision and Accuracy
  • Within-run and between-day precision studies (n=5 to n=35) yielded coefficients of variation (CV) below 7.8% at low, medium and high control levels.
  • Recoveries of control samples were within ±15% of target values.
Chromatographic and Spectral Confirmation
  • Clear baseline separation achieved within a 3-minute window for all four drugs.
  • MS2 spectra displayed distinct quantifier and qualifier ion transitions, supporting compound confirmation.

Benefits and Practical Applications


The validated LC-MS/MS workflow enables fast, robust and reproducible measurement of immunosuppressants in clinical laboratories. High throughput and stringent in vitro diagnostic labeling ensure suitability for therapeutic drug monitoring, quality control and laboratory-developed assay implementation.

Future Trends and Potential Applications


Advances in software automation and artificial intelligence-driven method optimization are expected to further streamline LC-MS/MS assays. Expansion of quantitation panels to include additional metabolites and biomarkers will enhance patient monitoring capabilities. Integration with laboratory information systems and remote diagnostics may support decentralized testing models.

Conclusion


The combination of Vanquish MD HPLC and TSQ Quantis MD mass spectrometry delivers reliable performance for quantifying key immunosuppressants in whole blood. The method meets stringent accuracy and precision targets, providing a robust platform for clinical laboratories.

References


1. Clinical and Laboratory Standards Institute. Liquid Chromatography-Mass Spectrometry Methods; Approved Guideline. CLSI document C62-A. Wayne, PA; 2014.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Exemplary performance data of a Triple Quadrupole Mass Spectrometer in a Simulated Clinical LDT Workflow
Exemplary performance data of a Triple Quadrupole Mass Spectrometer in a Simulated Clinical LDT Workflow BC Cha, Mindy Gao, Kristine Van Natta, Terry Olney, Maurino Flora, John Glazier, Thermo Fisher Scientific, S355 River Oaks Parkway San Jose, CA 95134, USA…
Key words
qciii, qciiiqcv, qcvsirolimus, sirolimusqci, qcirsd, rsdcyclosporina, cyclosporinatacrolimus, tacrolimuseverolimus, everolimusestrone, estronetestosterone, testosteroneldt, ldtloq, loqimprecision, imprecisioncyclosporin, cyclosporincapability
Tomorrow’s quantitation: increased robustness for quantitation of immunosuppressant drugs in blood with the TSQ Fortis MS for clinical research
TECHNICAL NOTE 65206 Tomorrow’s quantitation: increased robustness for quantitation of immunosuppressant drugs in blood with the TSQ Fortis MS for clinical research Authors Neloni Wijeratne, Claudia Martins, Kristine Van Natta, Xiaolei Xie, Thermo Fisher Scientific, San Jose, CA Debadeep Bhattacharyya,…
Key words
fortis, fortiscleaning, cleaningimmunosuppressant, immunosuppressanttsq, tsqcyclosporin, cyclosporintriple, triplesirolimus, sirolimuseverolimus, everolimustacrolimus, tacrolimusquantitation, quantitationblood, bloodafter, afterdrugs, drugsbefore, beforequadrupole
Testing robustness: Immunosuppressant drugs in blood with a TSQ Quantis MS for clinical research
TECHNICAL NOTE 64969 Testing robustness: Immunosuppressant drugs in blood with a TSQ Quantis MS for clinical research Authors Kristine Van Natta, Neloni Wijeratne, Claudia Martins Thermo Fisher Scientific, San Jose, CA Keywords Immunosuppressant, tacrolimus, sirolimus, everolimus, cyclosporin A, TSQ Quantis…
Key words
quantis, quantistsq, tsqcyclosporin, cyclosporincalibrators, calibratorsblood, bloodmonitored, monitoredrobustness, robustnessimmunosuppressant, immunosuppressanttriple, tripleclinical, clinicalcalculated, calculatedquadrupole, quadrupolewere, weremass, massmartins
Quantification of immunosuppressants in human whole blood by liquid chromatography-tandem mass spectrometry for clinical research
TECHNICAL NOTE 64916 Quantification of immunosuppressants in human whole blood by liquid chromatography-tandem mass spectrometry for clinical research Authors Claudio De Nardi,1 Katharina Kern,2 Steffen Peters2 Thermo Fisher Scientific GmbH, Dreieich, Germany 2 RECIPE Chemicals + Instruments GmbH, Munich, Germany…