How Does the Biotage Horizon 5000 Perform?

Technical notes | 2017 | BiotageInstrumentation
Sample Preparation
Industries
Manufacturer
Biotage

Summary

Importance of the Topic


Solid phase extraction (SPE) is essential for trace-level environmental and drinking water analysis. The Biotage Horizon 5000 automates disk-based SPE aiming to deliver ultra-low background, high recovery, and consistent performance to meet regulatory standards at microgram per liter concentrations.

Study Objectives and Overview


This evaluation compares the Horizon 5000 to its predecessor, focusing on background contamination, analyte recovery, precision, and operational robustness over extended use. A full suite of ~100 semivolatile compounds in drinking water matrices was assessed.

Methodology and Instrumentation


Key aspects:
  • Monitoring phthalate levels in blanks over three months to verify low system background
  • Measuring recoveries of a low-level spike for ~100 analytes in replicate samples across one month
  • Calculating relative standard deviations for precision assessment
  • Comparing daily average recoveries between Horizon 5000 and SPE-DEX 4790 systems

Used Instrumentation


Primary instruments:
  • Biotage Horizon 5000 automated SPE-DEX 5000 system
  • SPE-DEX 4790 reference system
  • Gas chromatograph with appropriate detectors for semivolatile analyte quantification

Main Results and Discussion


Blank phthalate levels remained below 0.1 µg/L (EPA 525.2 limit) throughout three months, confirming minimal background contamination. Analyte recoveries consistently fell within 70–130% over a month, and precision (RSD) was below 30% for low-level spikes. Comparative tests showed equivalent daily average recoveries between Horizon 5000 and the older 4790 model, demonstrating maintained or improved performance.

Benefits and Practical Applications


  • Ultra-low background enables reliable measurement of trace contaminants
  • Robust performance over extended use reduces downtime
  • High accuracy and precision across diverse compound classes
  • Seamless upgrade path for laboratories using SPE-DEX 4790, with additional software and hardware enhancements

Future Trends and Opportunities


Further developments may include integration with liquid chromatography-mass spectrometry workflows, expanded analyte libraries, greener solvent protocols, real-time monitoring and optimization using advanced software, and increased automation for high-throughput analysis.

Conclusion


The Biotage Horizon 5000 demonstrates reliable low-background operation, excellent recovery and precision for a broad suite of semivolatile analytes, and performance comparable to its predecessor with added functional enhancements. Its robust design supports consistent, high-quality SPE in environmental and drinking water testing laboratories.

References


  • Extraction of a Full Suite of Semivolatile Compounds from Drinking Water using Automated Solid Phase Extraction, AN1141703_01, 2017

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Atlantic® ReadyDisk - Ready to go SPE Disks for Water Extraction
Atlantic ReadyDisk | Page 1 ® Atlantic ReadyDisk ® Ready to go SPE Disks for Water Extraction Atlantic ReadyDisks are solid phase extraction (SPE) disks, pre-assembled into disposable holders and are used for extracting organic and semivolatile organic compounds from…
Key words
atlantic, atlanticreadydisk, readydiskreadydisks, readydisksdvb, dvbextracting, extractingmedia, mediaextraction, extractiondisk, diskconstructed, constructedholder, holderphase, phaseinteractions, interactionsunpredictable, unpredictablepolar, polarreversed
Atlantic® & Pacific® Disk Selection Guide
Atlantic & Pacific Disk Selection Guide | Page 1 ® ® Atlantic & Pacific Disk Selection Guide ® Compounds »» Pesticides »» »» »» »» »» »» »» bisphenol A general SVOCs PDBEs herbicides PAHs carbonyls dioxins furans »» Pesticides…
Key words
atlantic, atlanticepa, epagrease, greasedisk, diskspe, spefurans, furanspdbes, pdbeswastewater, wastewaterhlb, hlbdioxins, dioxinsdisks, diskspesticides, pesticidessvocs, svocspahs, pahsextractable
PAH Analysis in Drinking Water Using Disk SPE Automated with the SPE-DEX 5000
Application Note AN119-HOR PAH ANalysis in Drinking Water Using Disk SPE Automated with the SPE-DEX 5000 Page 1 PAH Analysis in Drinking Water using Disk PAHAutomated Analysis in with Drinking Water using SPE the SPE-DEX 5000Disk SPE Automated with the…
Key words
mecl, meclbenzo, benzodrinking, drinkingwater, wateroff, offfluoranthene, fluoranthenehorizon, horizonpahs, pahstime, timeanthracene, anthracenebiodegrade, biodegradepyrene, pyrenelse, lseembraced, embracedprominence
Operating the Biotage Horizon 5000 Software
Operating the Biotage Horizon 5000 Software | Page 1 ® Operating the Biotage Horizon 5000 Software ® (Previously known as SPE-DEX 5000) Software for the SPE-DEX ® 5000 The SPE-DEX 5000 software controls all operations of the system. It has…
Key words
run, runmessage, messagepause, pauseoperation, operationclicked, clickedmethod, methodoperator, operatorlines, lineschosen, chosenoperators, operatorsdevelopment, developmentroutine, routinefeatures, featuresbutton, buttonsimple