Quality Control of Synthetic Peptides using the MALDI-8030 Dual Polarity Benchtop MALDI-TOF Mass Spectrometer

Applications | 2022 | ShimadzuInstrumentation
MALDI, LC/TOF, LC/MS
Industries
Pharma & Biopharma
Manufacturer
Shimadzu

Summary

Significance of the Topic


Synthetic peptides are widely applied in biochemistry, immunology and medicine for diagnostics, therapeutics, antibody production and vaccine design. High purity and correct molecular identity are essential for their performance. MALDI-TOF mass spectrometry is a routine tool for peptide quality control, yet conventional positive ion analysis can lead to loss of labile modifications and complex spectra due to salt adducts.

Objectives and Overview of the Study


This study evaluates the Shimadzu MALDI-8030 dual polarity benchtop linear MALDI-TOF mass spectrometer for rapid quality control of synthetic peptides. The main goals are to compare positive and negative ion modes, demonstrate preservation of sensitive functional groups and assess spectral clarity through reduction of salt adduct interferences.

Methodology and Instrumentation


Samples:
  • Peptide D containing two labile sulfo groups on tyrosine residues.
  • Peptide A supplied in semi-pure and fully purified forms.
Sample Preparation:
  • Solutions in 1:1 acetonitrile/water with or without 0.1% trifluoroacetic acid (TFA) depending on ion mode.
  • Matrix: alpha-cyano-4-hydroxycinnamic acid (CHCA), 5 mg/mL in 1:1 acetonitrile/water, with or without 0.1% TFA.
Instrumentation:
  • Shimadzu MALDI-8030 dual polarity benchtop linear MALDI-TOF mass spectrometer.
  • Analysis performed in both positive and negative ion modes under standard operating conditions.

Main Results and Discussion


Peptide D:
  • Positive mode detection resulted in sequential loss of one and both sulfo groups; only desulfonated species were observed (m/z 2455 and 2375).
  • Negative mode preserved both sulfo modifications, allowing detection of the intact peptide at m/z 2532 with high mass accuracy and clear isotopic resolution; salt adducts were absent.
Peptide A (semi-pure and pure):
  • Positive mode spectra of semi-pure samples showed numerous sodium and potassium adducts of both impurities and target peptide, complicating interpretation.
  • Negative mode spectra exhibited a clean profile with minimal adduct formation, enabling straightforward identification of the peptide at m/z ~1730 in both semi-pure and pure forms.

Benefits and Practical Applications


The negative ion mode on a benchtop MALDI-TOF platform offers:
  • Improved detection of labile modifications by preventing in-source loss.
  • Simplified mass spectra through drastic reduction of salt adduct signals.
  • Faster, more reliable QC workflow for peptide manufacturing and research laboratories.

Future Trends and Potential Applications


Advancements may include coupling dual polarity MALDI-TOF with automated sample handling for high-throughput QC, extending negative mode analysis to other labile biomolecules, integration with ambient ionization techniques, and leveraging AI-driven spectral interpretation for real-time quality assessment.

Conclusion


Negative ion mode analysis on the MALDI-8030 benchtop MALDI-TOF mass spectrometer significantly enhances peptide quality control by retaining labile functional groups and producing cleaner spectra with minimal adduct interference. This approach streamlines purity assessment and molecular confirmation in both research and production settings.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Quality Control of Synthetic Peptides using the MALDI-8030 Dual Polarity Benchtop MALDI-TOF Mass Spectromete
MALDI-TOF Mass Spectrometry Analysis Application News MALDI-8030 Quality Control of Synthetic Peptides using the MALDI-8030 Dual Polarity Benchtop MALDI-TOF Mass Spectrometer S. Salivo (KRATOS ANALYTICAL LTD.) User Benefits  Easy, simple, negative ion quality control of peptides on affordable benchtop…
Key words
intensity, intensitypeptide, peptidenegative, negativemode, modeadducts, adductsadduct, adductsalt, saltpeptides, peptidescalculated, calculatedpure, puresynthetic, syntheticimpurity, impuritypositive, positivenews, newspotassium
Negative Mode Analysis of Synthetic Oligonucleotides using the MALDI-8030 Dual Polarity Benchtop MALDI-TOF Mass Spectromet
MALDI-TOF Mass Spectrometry Analysis Application News MALDI-8030 Negative Mode Analysis of Synthetic Oligonucleotides using the MALDI-8030 Dual Polarity Benchtop MALDI-TOF Mass Spectrometer S. Salivo (KRATOS ANALYTICAL LTD.) User Benefits  Simple analysis of oligonucleotides, in negative ion mode to reduce…
Key words
cftr, cftrprimer, primerethidium, ethidiumhomozygote, homozygoteallele, allelewild, wildheterozygote, heterozygotenegative, negativegene, geneoligonucleotides, oligonucleotidesmode, modepcr, pcrgel, gelnormal, normalelectrophoresis
Negative Mode Analysis of Synthetic Oligonucleotides using the MALDI-8030 Dual Polarity Benchtop MALDI-TOF Mass Spectrometer
MALDI-TOF Mass Spectrometry Analysis Application News MALDI-8030 Negative Mode Analysis of Synthetic Oligonucleotides using the MALDI-8030 Dual Polarity Benchtop MALDI-TOF Mass Spectrometer S. Salivo User Benefits  Simple analysis of oligonucleotides, in negative ion mode to reduce adducts, on an…
Key words
cftr, cftrprimer, primerethidium, ethidiumhomozygote, homozygoteallele, allelewild, wildheterozygote, heterozygotenegative, negativegene, geneoligonucleotides, oligonucleotidesmode, modepcr, pcrgel, gelnormal, normalelectrophoresis
Dual-Polarity Benchtop Linear MALDI-TOF Mass Spectrometer - MALDI-8030
MO473 Dual-Polarity Benchtop Linear MALDI-TOF Mass Spectrometer MALDI-8030 AFFORDABLE CLASSBest-in-class sensitivity and resolution • Easily detect analytes for your application Award-winning design • Based on the award-winning MALDI-8020 design Innovative electronics • Fast, high bit-depth transient recorder provides rapid data…
Key words
intensity, intensitymaldi, malditrueclean, truecleanevoo, evoolaser, laserdual, dualpolarity, polarityyour, yourbenchtop, benchtopaurasolution, aurasolutionintact, intactfleximass, fleximassmass, massmisidentification, misidentificationemstat
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike