Post-column chemistry for improved optical absorption detection
Posters | | MetrohmInstrumentation
Trace-level quantification of analytes by UV/VIS detection is essential in environmental monitoring, pharmaceutical quality control and industrial process analysis. Direct absorption measurements often suffer from poor sensitivity or selectivity when analytes lack strong chromophores or are present in complex matrices. Post-column derivatization remedies this by generating intense chromophores after chromatographic separation, enhancing detection without degrading resolution.
This work examines a versatile flow-through reactor for automated post-column chemistry, demonstrating four key applications: amino acid profiling via ninhydrin reaction (USP), bromate detection through iodide derivatization (EPA 326.0), silicate analysis with molybdate complexation and UV detection, and hexavalent chromium quantification using diphenylcarbazide (EPA 218.6). The reactor’s temperature, flow rates and geometry were systematically optimized to maximize derivatization efficiency and analytical performance.
Separations were conducted with ion chromatography coupled to a professional reactor capable of 25–120 °C, adjustable volume and minimal dead space. Key instrumentation included:
• Amino acids: Ninhydrin derivatization at 120 °C achieved a detection limit of 0.25 µmol/L on a Metrosep Amino Acids column.
• Bromate: Iodide-based post-column reaction at 25 °C enabled quantification down to 1 µg/L in drinking water samples on Metrosep A Supp 16.
• Silicate: Molybdate complex formation at 30 °C provided sensitive detection at 360 nm for 50 mg/L standards using a Hamilton PRP-X100 column.
• Chromate(VI): Diphenylcarbazide reaction at optimized temperatures (up to 55 °C) yielded reliable measurement of 1 mg/L Cr(VI) with minimal interference.
The reactor’s customizable geometry and precise thermal control delivered high derivatization yields, reproducibility and compatibility with challenging matrices.
Future developments may include integration of post-column reactors with mass spectrometers and high-resolution detectors, advancing detection capabilities. Novel microreactor designs and green derivatization reagents promise faster reactions and reduced reagent consumption. AI-driven method optimization and real-time monitoring will streamline workflows and enhance analytical throughput.
The presented flow-through reactor offers flexible, automated post-column chemistry for enhanced UV/VIS detection across diverse analytes. Its robust design, precise temperature control and low dead volume preserve chromatographic performance while achieving trace-level sensitivity, supporting applications in environmental, pharmaceutical and industrial analysis.
Ion chromatography
IndustriesEnvironmental, Pharma & Biopharma
ManufacturerMetrohm
Summary
Significance of the Topic
Trace-level quantification of analytes by UV/VIS detection is essential in environmental monitoring, pharmaceutical quality control and industrial process analysis. Direct absorption measurements often suffer from poor sensitivity or selectivity when analytes lack strong chromophores or are present in complex matrices. Post-column derivatization remedies this by generating intense chromophores after chromatographic separation, enhancing detection without degrading resolution.
Study Objectives and Overview
This work examines a versatile flow-through reactor for automated post-column chemistry, demonstrating four key applications: amino acid profiling via ninhydrin reaction (USP), bromate detection through iodide derivatization (EPA 326.0), silicate analysis with molybdate complexation and UV detection, and hexavalent chromium quantification using diphenylcarbazide (EPA 218.6). The reactor’s temperature, flow rates and geometry were systematically optimized to maximize derivatization efficiency and analytical performance.
Methodology and Instrumentation
Separations were conducted with ion chromatography coupled to a professional reactor capable of 25–120 °C, adjustable volume and minimal dead space. Key instrumentation included:
- 850 Professional IC
- 889 IC Sample Center
- 872 Extension Module IC Pump
- 886 Professional Reactor
- 887 Professional UV/VIS Detector
Main Results and Discussion
• Amino acids: Ninhydrin derivatization at 120 °C achieved a detection limit of 0.25 µmol/L on a Metrosep Amino Acids column.
• Bromate: Iodide-based post-column reaction at 25 °C enabled quantification down to 1 µg/L in drinking water samples on Metrosep A Supp 16.
• Silicate: Molybdate complex formation at 30 °C provided sensitive detection at 360 nm for 50 mg/L standards using a Hamilton PRP-X100 column.
• Chromate(VI): Diphenylcarbazide reaction at optimized temperatures (up to 55 °C) yielded reliable measurement of 1 mg/L Cr(VI) with minimal interference.
The reactor’s customizable geometry and precise thermal control delivered high derivatization yields, reproducibility and compatibility with challenging matrices.
Benefits and Practical Applications
- Broad applicability to both inorganic and organic analytes
- High sensitivity and low detection limits
- Minimal dead volume preserving chromatographic resolution
- Robust flow path resistant to extreme pH and solvents
- Automated operation reducing manual errors
Future Trends and Opportunities
Future developments may include integration of post-column reactors with mass spectrometers and high-resolution detectors, advancing detection capabilities. Novel microreactor designs and green derivatization reagents promise faster reactions and reduced reagent consumption. AI-driven method optimization and real-time monitoring will streamline workflows and enhance analytical throughput.
Conclusion
The presented flow-through reactor offers flexible, automated post-column chemistry for enhanced UV/VIS detection across diverse analytes. Its robust design, precise temperature control and low dead volume preserve chromatographic performance while achieving trace-level sensitivity, supporting applications in environmental, pharmaceutical and industrial analysis.
References
- U.S. Pharmacopeia (USP): Post-column Ninhydrin Derivatization for Amino Acids Analysis
- U.S. EPA Method 326.0: Bromate Determination by Post-Column Iodide Reaction
- U.S. EPA Method 218.6: Hexavalent Chromium Determination via Diphenylcarbazide
- Metrohm Application Note 8.000.6063EN: Post-Column Chemistry Reactor Specifications
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
947 Professional UV/VIS Detector, 943 Reactor, 943 Thermostat Vario
2019|Metrohm|Brochures and specifications
947 Professional UV/VIS Detector Vario 943 Professional Reactor Vario 943 Professional Thermostat Vario UV/VIS detection – an alternative detection method in ion chromatography UV/VIS Detection – catch the color 02 The 947 Professional UV/VIS Detector Vario enables reliable and accurate…
Key words
reactor, reactorlamp, lampvario, variovis, vispostcolumn, postcolumnprecolumn, precolumnbromate, bromateprofessional, professionalmau, mausilicate, silicateintensity, intensityintelligent, intelligentderivatization, derivatizationchromium, chromiummeasuring
Analysis of water samples and water constituents with Metrohm instruments
2021|Metrohm|Guides
Monograph Analysis of water samples and water constituents with Metrohm instruments Peter A. Bruttel and Nadine Seifert, revised by Dr. Markus Läubli, Uwe Loyall, and Lucia Meier Preface Water is essential for all living organisms – without water no life…
Key words
remarks, remarksliterature, literaturemetrohm, metrohmpotential, potentialelectrode, electrodemode, modegeneral, generalinstrument, instrumenthardness, hardnesswater, waterprofessional, professionalinstruments, instrumentsexample, exampletitrant, titrantmmol
944 Professional UV/VIS Detector, 943 Reactor, 943 Thermostat Vario
2013|Metrohm|Brochures and specifications
944 Professional UV/VIS Detector Vario 943 Professional Reactor Vario 943 Professional Thermostat Vario UV/VIS detection – an alternative detection method in ion chromatography UV/VIS Detektion – catch the color 02 The 944 Professional UV/VIS Detector Vario enables reliable and accurate…
Key words
reactor, reactorvario, variopostcolumn, postcolumnvis, visprecolumn, precolumnbromate, bromateprofessional, professionalintelligent, intelligentderivatization, derivatizationmeasuring, measuringunsur, unsurthermostat, thermostatanaly, analydetektion, detektiontransfer
Trace-Level Determination of Bromate in Water by Ion Chromatography with Optimized Post-Column Derivatization and UV/vis Detection as Triiodide
2007|Metrohm|Applications
2 Food and Beverage ADVERTISING SUPPLEMENT THE APPLICATION NOTEBOOK – SEPTEMBER 2007 Trace-Level Determination of Bromate in Water by Ion Chromatography with Optimized Post-Column Derivatization and UV/vis Detection as Triiodide Achim Rumi,1 Rob Proost2 and Alfred Steinbach1 1Metrohm D uring…
Key words
iodide, iodidemolybdate, molybdatetriiodide, triiodidesulphuric, sulphuricbromate, bromateadvertising, advertisingpost, postpotassium, potassiumammonium, ammoniumeluent, eluentsupplement, supplementcolumn, columnbeverage, beveragevis, visreaction