Installation Instruction for ProfIC Vario 1 Anion
Manuals | | MetrohmInstrumentation
Ion chromatography
IndustriesManufacturerMetrohm
Summary
Importance of the Topic
Ion chromatography with sequential suppression provides a fully automated and reliable approach to routine anion analysis in environmental, industrial, and quality control laboratories. Standardized installation and configuration of the ProfIC Vario 1 Anion system ensure consistent results, reduce downtime, and simplify operator training.Objectives and Study Overview
This application bulletin presents a step-by-step guide for installing and configuring the ProfIC Vario 1 Anion system. It covers hardware assembly, software setup, method importation, system purge, and an example evaluation of standard anions to demonstrate performance metrics.Methodology and Instrumentation
- Core System: 940 Professional IC Vario 1 Anion with sequential suppression
- Sample Handling: 858 Professional Sample Processor controlled via USB
- Software: MagIC Net 3.1 for instrument configuration, method creation, and data evaluation
- Optional Modules:
- Liquid Handling Station (LQH) for needle rinsing and dilution
- Eluent Production Module (941) for automated eluent preparation
- Key Installation Steps:
- Install MagIC Net software and drivers
- Assemble detector, suppressor, sample processor, and tubing according to the device schematic
- Configure instruments and define eluent, suppressor solution, column, and rack positions in MagIC Net
- Purge the system, rinse suppressor and column, then equilibrate until a stable baseline is reached
Main Results and Discussion
An example mixture of fluoride, chloride, nitrite, bromide, nitrate, sulfate, and phosphate (5 mg/L each) was run using a Metrosep A Supp 15 - 150/4.0 column at 45 °C. Key performance data include:- Precision (10 injections of 2 mg/L standards): relative standard deviations from 0.1 % (sulfate) to 1.3 % (bromide)
- Calibration (1–20 mg/L, quadratic fit): correlation coefficients ≥ 0.99994 for all ions
Benefits and Practical Applications
- Fully automated workflow minimizes manual intervention and contamination risk
- Flexible configuration supports a wide range of anions and custom method development
- MagIC Net integration enables streamlined data handling and compliance with regulatory standards
- Optional modules enhance throughput through automated eluent preparation and sample rinsing
Future Trends and Potential Uses
Continued integration with automated eluent modules, advanced suppressor technologies, and real-time data analytics will further simplify ion chromatographic workflows. Leveraging cloud-based software and artificial intelligence for predictive maintenance and method optimization represents a promising direction for next-generation analytical platforms.Conclusion
The ProfIC Vario 1 Anion system delivers a robust and user-friendly solution for routine anion analysis. A systematic installation and configuration protocol combined with MagIC Net software ensures high precision, reproducibility, and ease of operation across diverse laboratory environments.References
No references were provided in the original text.Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Installation Instruction for ProfIC Vario 2 Anion
|Metrohm|Manuals
Application Bulletin 365 Installation Instruction for ProfIC Vario 2 Anion ProfIC Vario 2 Anion is a Professional IC Vario system for the fully automatic determination of anions or cations with sequential suppression with Metrohm Inline Ultrafiltration. «ProfIC Vario 2» safely…
Key words
msm, msmstation, stationrinsing, rinsingultrafiltration, ultrafiltrationregeneration, regenerationhandling, handlingconnected, connectedrotor, rotorperistaltic, peristalticcapillary, capillarylabeled, labeledstoppers, stoppersexemplary, exemplarysuppressor, suppressormcs
Installation Instruction for ProfIC Vario 10 Anion
|Metrohm|Manuals
Application Bulletin 386 Installation Instruction for ProfIC Vario 10 Anion ProfIC Vario 10 – Anion is a Professional IC Vario system with Inline Neutralization of the sample. When strongly acidic or alkaline samples are involved, Inline Neutralization is recommended. With…
Key words
msm, msmconnected, connectedstation, stationrinsing, rinsingregeneration, regenerationhandling, handlingrotor, rotorlqh, lqhlabeled, labeledcollector, collectorsuppressor, suppressormcs, mcsnow, nowcapillary, capillarywaste
Installation Instruction for ProfIC Vario 8 Anion
|Metrohm|Manuals
Application Bulletin 381 Installation Instruction for ProfIC Vario 8 Anion The ProfIC Vario 8 Anion is a Professional IC Vario system with Inline Sample Preconcentration and Matrix Elimination for anions or cations with sequential suppression. «ProfIC Vario 8» safely masters…
Key words
msm, msmstation, stationrinsing, rinsingconnected, connectedregeneration, regenerationhandling, handlingrotor, rotorcapillary, capillarylabeled, labeledsuppressor, suppressormcs, mcssolution, solutionpump, pumpliquid, liquidperistaltic
Installation Instruction for ProfIC Vario 4 Anion
|Metrohm|Manuals
Application Bulletin 370 Installation Instruction for ProfIC Vario 4 Anion ProfIC Vario 4 Anion is a Professional IC Vario system with Inline Dilution for the fully automatic determination of anions and cations with sequential suppression in concentrated solutions. «ProfIC Vario…
Key words
msm, msmrinsing, rinsingconnected, connectedrotor, rotorperistaltic, peristaltictubing, tubinglqh, lqhprocessor, processorpump, pumpcapillary, capillarylabeled, labeledsuppressor, suppressorstation, stationmcs, mcsoutlet