Installation Instruction for a High Pressure Gradient System
Manuals | | MetrohmInstrumentation
Ion chromatography with high-pressure gradient elution has become a cornerstone technique for the selective, reproducible, and fully automated determination of anions and cations in diverse matrices. By precisely controlling eluent composition via two or more high-pressure pumps, laboratories benefit from enhanced separation efficiency, shorter analysis times, and improved sensitivity. This flexibility is critical for environmental monitoring, food safety, pharmaceutical quality control, and industrial process analytics, where reliable quantification of ionic species underpins decision making and regulatory compliance.
This Application Bulletin describes the installation, configuration, and operation of a High-Pressure Gradient System built around the Metrohm 940 Professional IC Vario, 858 Professional Sample Processor, and MagIC Net software. Key goals are to guide users through hardware and software setup, ensure correct fluidic connections, demonstrate method import and gradient configuration in MagIC Net, and illustrate an exemplary analysis of organic acids using a high-pressure carbonate/bicarbonate gradient.
The high-pressure gradient system relies on two synchronized high-pressure pumps to blend Eluent A (6.4 mmol/L Na2CO3 + 2.0 mmol/L NaHCO3) with ultrapure water (Eluent B). Gradient segments are defined in MagIC Net to vary eluent composition from 10 % to 90 % A over user-defined ramps.
An exemplary analysis of 14 organic acids (e.g., lactate, acetate, propionate, malate) using a 90-minute high-pressure gradient achieved excellent calibration linearity (R2 between 0.999758 and 0.999999) and recoveries near 100 %. Repeatability (n=9) showed relative standard deviations below 3.5 % for all analytes. The gradient returned to initial conditions for thorough equilibration between runs, avoiding retention time shifts. The chemical suppressor operating in automatic stepping mode delivered stable baselines and minimized background noise for high sensitivity detection.
As analytical demands evolve, high-pressure gradient IC systems will integrate stronger software-driven method development, AI-assisted parameter optimization, and expanded multi-pump configurations for ternary or quaternary eluents. Miniaturized fluidics, faster detectors, and real-time data analytics will accelerate throughput in environmental monitoring, metabolomics, and process control. Connectivity enhancements will enable remote system diagnostics and predictive maintenance.
The installation and configuration protocol presented here enable robust, high-pressure gradient ion chromatography with Metrohm IC instruments. Through careful hardware setup, software configuration in MagIC Net, and method optimization, users can achieve high accuracy, precision, and reproducibility for complex ionic analyses. The modular platform supports seamless expansion and automation for diverse laboratory needs, ensuring long-term analytical performance and productivity.
Ion chromatography
IndustriesManufacturerMetrohm
Summary
Importance of the Topic
Ion chromatography with high-pressure gradient elution has become a cornerstone technique for the selective, reproducible, and fully automated determination of anions and cations in diverse matrices. By precisely controlling eluent composition via two or more high-pressure pumps, laboratories benefit from enhanced separation efficiency, shorter analysis times, and improved sensitivity. This flexibility is critical for environmental monitoring, food safety, pharmaceutical quality control, and industrial process analytics, where reliable quantification of ionic species underpins decision making and regulatory compliance.
Objectives and Overview
This Application Bulletin describes the installation, configuration, and operation of a High-Pressure Gradient System built around the Metrohm 940 Professional IC Vario, 858 Professional Sample Processor, and MagIC Net software. Key goals are to guide users through hardware and software setup, ensure correct fluidic connections, demonstrate method import and gradient configuration in MagIC Net, and illustrate an exemplary analysis of organic acids using a high-pressure carbonate/bicarbonate gradient.
Methodology and Used Instrumentation
The high-pressure gradient system relies on two synchronized high-pressure pumps to blend Eluent A (6.4 mmol/L Na2CO3 + 2.0 mmol/L NaHCO3) with ultrapure water (Eluent B). Gradient segments are defined in MagIC Net to vary eluent composition from 10 % to 90 % A over user-defined ramps.
- Primary Instrumentation
- 940 Professional IC Vario ONE/SeS/HPG (Article No. 2.940.1440)
- IC Conductivity Detector (2.850.9010)
- 858 Professional Sample Processor – Pump version (2.858.0020)
- MagIC Net 3.1 Professional Software (6.6059.312)
- Columns and Suppression
- Metrosep A Supp 7 - 250/4.0 and guard column
- MSM-HC rotor A for chemical suppression with automatic stepping
- Accessories for Eluent Handling
- Accessory Kit Vario/Flex Basic and ONE (6.5000.000, 6.5000.010)
- Eluent bottles, mixing coil, filters, adapters, and tubing for high-pressure gradient
- Optional Modules
- Liquid Handling Station (LQH) for needle rinsing and dilution
- Eluent Production Module (941) for automated eluent preparation
Main Results and Discussion
An exemplary analysis of 14 organic acids (e.g., lactate, acetate, propionate, malate) using a 90-minute high-pressure gradient achieved excellent calibration linearity (R2 between 0.999758 and 0.999999) and recoveries near 100 %. Repeatability (n=9) showed relative standard deviations below 3.5 % for all analytes. The gradient returned to initial conditions for thorough equilibration between runs, avoiding retention time shifts. The chemical suppressor operating in automatic stepping mode delivered stable baselines and minimized background noise for high sensitivity detection.
Benefits and Practical Applications
- Fully automated high-pressure gradient elution reduces manual intervention and potential for error.
- Wide eluent composition range (10 %–90 %) allows tailored resolution of complex ionic mixtures.
- Chemical suppression with MSM rotors ensures low background and stable baselines.
- Integration of sample processor guarantees reproducible injection timing, crucial for gradient methods.
- Optional liquid handling and eluent production modules further streamline workflow in routine laboratories.
Future Trends and Applications
As analytical demands evolve, high-pressure gradient IC systems will integrate stronger software-driven method development, AI-assisted parameter optimization, and expanded multi-pump configurations for ternary or quaternary eluents. Miniaturized fluidics, faster detectors, and real-time data analytics will accelerate throughput in environmental monitoring, metabolomics, and process control. Connectivity enhancements will enable remote system diagnostics and predictive maintenance.
Conclusion
The installation and configuration protocol presented here enable robust, high-pressure gradient ion chromatography with Metrohm IC instruments. Through careful hardware setup, software configuration in MagIC Net, and method optimization, users can achieve high accuracy, precision, and reproducibility for complex ionic analyses. The modular platform supports seamless expansion and automation for diverse laboratory needs, ensuring long-term analytical performance and productivity.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Installation Instruction for ProfIC Vario 1 Anion
|Metrohm|Manuals
Application Bulletin 359 Installation Instruction for ProfIC Vario 1 Anion ProfIC Vario 1 Anion is a Professional IC Vario system for the fully automatic determination of anions or cations with sequential suppression. «ProfIC Vario 1» safely masters all routine tasks…
Key words
msm, msmrinsing, rinsingstation, stationcapillary, capillaryperistaltic, peristalticregeneration, regenerationconnected, connectedrotor, rotorpump, pumplabeled, labeledexemplary, exemplarysuppressor, suppressormcs, mcshandling, handlingdesignation
Installation instruction for ProfIC Vario 1 AnCat
|Metrohm|Manuals
Application Bulletin 361 Installation instruction for ProfIC Vario 1 AnCat ProfIC Vario 1 AnCat is a very compact and professional IC system for the fully automatic determination of anions with sequential suppression and cations. With «ProfIC Vario 1 AnCat» anions…
Key words
msm, msmrinsing, rinsingstation, stationcation, cationanion, anionconnect, connectprocessor, processorregeneration, regenerationconnected, connectedsuppressor, suppressormcs, mcsnow, nowhandling, handlingcapillary, capillarydosino
Installation Instruction for ProfIC Vario 4 AnCat
|Metrohm|Manuals
Application Bulletin 372 Installation Instruction for ProfIC Vario 4 AnCat ProfIC Vario 4 AnCat is a Professional IC Vario system with Inline Dilution for the fully automatic determination of anions and cations simultaneously in concentrated solutions. «ProfIC Vario 4» can…
Key words
msm, msmrinsing, rinsingmcs, mcsconnected, connectedrotor, rotornitrite, nitriteprocessor, processorperistaltic, peristalticregeneration, regenerationstandard, standardbromide, bromidefluoride, fluoridenitrate, nitrateexemplary, exemplarysuppressor
Installation Instruction for ProfIC Vario 4 Anion
|Metrohm|Manuals
Application Bulletin 370 Installation Instruction for ProfIC Vario 4 Anion ProfIC Vario 4 Anion is a Professional IC Vario system with Inline Dilution for the fully automatic determination of anions and cations with sequential suppression in concentrated solutions. «ProfIC Vario…
Key words
msm, msmrinsing, rinsingconnected, connectedrotor, rotorperistaltic, peristaltictubing, tubinglqh, lqhprocessor, processorcapillary, capillarypump, pumplabeled, labeledsuppressor, suppressorstation, stationmcs, mcsoutlet