Small Molecule System Suitability (Evaluation) Test for LC-MS/MS
Posters | 2022 | Thermo Fisher Scientific | ASMSInstrumentation
Liquid chromatography–tandem mass spectrometry (LC-MS/MS) is a cornerstone technique for trace-level quantitation in pharmaceutical, environmental and clinical laboratories. Routine system suitability tests are essential to confirm that all modules (pump, autosampler, column and mass spectrometer) perform within defined criteria before sample analysis. A rapid, integrated approach reduces downtime, ensures data quality and supports regulatory compliance.
This work presents a small molecule system suitability evaluation (SMSS) test using a pre-formulated standard to assess the entire LC-MS/MS workflow in under 5 minutes. Key goals include:
The SMSS protocol employs a 25 µL aliquot of Pierce™ SMSS standard in a silanized vial. A 0.1 µL injection is repeated 10 times using a 5-minute binary gradient (water/0.1% formic acid and methanol) on a UHPLC system. Mass spectrometric detection uses heated electrospray ionization with at least two selected-reaction monitoring transitions per compound, covering m/z 75–1222 in both positive and negative modes. Automated software (TraceFinder™ or Chromeleon™) processes data, applies pass/fail criteria based on peak area, retention time precision and linearity, then generates a summary report.
The SMSS test provides module-specific performance metrics:
The SMSS workflow offers:
Ongoing developments may include cloud-based data aggregation for multi-site performance comparison, integration of additional chemistries or larger panels of analytes, and enhanced software analytics for predictive maintenance of LC-MS/MS systems.
The Pierce™ SMSS test provides a fast, user-friendly approach to assess LC-MS/MS system readiness. By combining minimal sample handling, defined acceptance criteria and automated reporting, laboratories can ensure reliable instrument performance and maintain confidence in routine analytical results.
LC/MS, LC/MS/MS, LC/QQQ
IndustriesOther
ManufacturerThermo Fisher Scientific
Summary
Significance of the Topic
Liquid chromatography–tandem mass spectrometry (LC-MS/MS) is a cornerstone technique for trace-level quantitation in pharmaceutical, environmental and clinical laboratories. Routine system suitability tests are essential to confirm that all modules (pump, autosampler, column and mass spectrometer) perform within defined criteria before sample analysis. A rapid, integrated approach reduces downtime, ensures data quality and supports regulatory compliance.
Objectives and Study Overview
This work presents a small molecule system suitability evaluation (SMSS) test using a pre-formulated standard to assess the entire LC-MS/MS workflow in under 5 minutes. Key goals include:
- Demonstrate a simple, fast protocol with minimal sample preparation.
- Define acceptance criteria for each hardware module and monitor carryover, precision and sensitivity.
- Enable automated data processing and reporting for routine and long-term performance tracking.
Methodology and Instrumentation
The SMSS protocol employs a 25 µL aliquot of Pierce™ SMSS standard in a silanized vial. A 0.1 µL injection is repeated 10 times using a 5-minute binary gradient (water/0.1% formic acid and methanol) on a UHPLC system. Mass spectrometric detection uses heated electrospray ionization with at least two selected-reaction monitoring transitions per compound, covering m/z 75–1222 in both positive and negative modes. Automated software (TraceFinder™ or Chromeleon™) processes data, applies pass/fail criteria based on peak area, retention time precision and linearity, then generates a summary report.
Used Instrumentation
- Vanquish™ Flex or Horizon Binary UHPLC system with temperature-controlled autosampler and column compartment
- Hypersil GOLD™ 2.1 × 50 mm column
- TSQ Altis™ Plus, TSQ Quantis™ Plus or TSQ Fortis™ Plus triple quadrupole mass spectrometer
- Thermo Scientific™ TraceFinder™ and Chromeleon™ software for data acquisition and reporting
Key Results and Discussion
The SMSS test provides module-specific performance metrics:
- Pump precision: Ten-injection overlay of pump pressure profiles and retention times exhibit excellent reproducibility.
- Autosampler accuracy: Linear response across injection volumes (0.1–0.5 µL) with %RSD <5% demonstrates volume precision.
- Mass spectrometer sensitivity: All analytes exceed intensity thresholds with RSD <10% in positive and negative modes.
- Carryover assessment: Blank injections before and after the highest-volume SMSS spike show carryover below 0.01%.
- Ion ratio consistency: Product-ion ratios comply with EU Council Directive 2002/657/EC confirmatory criteria.
- Long-term robustness: A single vial supports ~200 consecutive injections (~16 hours) with stable response for key analytes.
Benefits and Practical Applications
The SMSS workflow offers:
- Rapid system check in under 5 minutes without complex sample prep.
- Comprehensive evaluation of pump, autosampler, column, MS and software integration.
- Automated pass/fail reporting for consistent QC documentation.
- Capability for intra- and inter-laboratory performance tracking and benchmarking.
Future Trends and Applications
Ongoing developments may include cloud-based data aggregation for multi-site performance comparison, integration of additional chemistries or larger panels of analytes, and enhanced software analytics for predictive maintenance of LC-MS/MS systems.
Conclusion
The Pierce™ SMSS test provides a fast, user-friendly approach to assess LC-MS/MS system readiness. By combining minimal sample handling, defined acceptance criteria and automated reporting, laboratories can ensure reliable instrument performance and maintain confidence in routine analytical results.
References
- European Commission. Council Directive 96/23/EC as amended by 2002/657/EC on the performance of analytical methods and interpretation of results.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Mass Spectrometry Applications for Food Safety Analysis
2012|Thermo Fisher Scientific|Guides
A p pli ca t i o n N o teb o ok Mass Spectrometry Applications for Food Safety Analysis • Pesticides • Drug Residues • Mycotoxins • Marine Toxins • Additional Contaminants • Natural Compounds and Food Additives Mass…
Key words
pos, posmass, massscientific, scientificthermo, thermoturboflow, turboflowqual, qualaldrich, aldrichion, ionsigma, sigmamethod, methodtsq, tsqsrm, srmquant, quantwere, werepesticide
Development of Small Molecule System Suitability Standard (SMSS) for routine monitoring of the performance of a liquid chromatography/mass spectrometry (LC-MS) system
2022|Thermo Fisher Scientific|Posters
Development of Small Molecule System Suitability Standard (SMSS) for routine monitoring of the performance of a liquid chromatography/mass spectrometry (LC-MS) system Jae Choi1, Bhavin Patel1, Alan Atkins2, Nicole Zehethofer3, Charles Maxey4, Claudia Martins4, Kay Opperman1, Ryan Bomgarden1; 1Thermo Fisher Scientific,…
Key words
smss, smsssuitability, suitabilitylot, lotpiercetm, piercetmmolecule, moleculesystem, systemsmall, smallconsistency, consistencyday, daystandard, standardroutine, routinerobust, robustsst, sstmonitoring, monitoringaccessed
An integrated LC-MS system performance evaluation test for peptide mapping and monitoring
2021|Thermo Fisher Scientific|Applications
APPLICATION NOTE 000336 An integrated LC-MS system performance evaluation test for peptide mapping and monitoring Authors: Hao Yang1, Melissa Sato1, Royston Quintyn2, Catharina Crone3, Julia Kraegenbring3 Kai Scheffler4, Min Du5, Thermo Fisher Scientific, San Jose, CA, US 1 2 Thermo…
Key words
slhtlfgdelc, slhtlfgdelcattributes, attributeseworkflow, eworkflowsystem, systempeptide, peptidesequence, sequenceperformance, performanceorbitrap, orbitrapmonitored, monitoredcheck, checktemplate, templatevanquish, vanquishhpyfyapellyyank, hpyfyapellyyankiaevek, iaevekshc
Usability Study of a New HPLC, a New Tandem MS and a New Data Processing Software
2015|Thermo Fisher Scientific|Posters
Poster N o te 6 43 75 Usability Study of a New HPLC, a New Tandem MS and a New Data Processing Software Jason Lai, Jia Wang, Brad Hart, Kristine Van Natta, Marta Kozak, Jorge Valdivia, Haibo Wang, Erkan Diri…
Key words
hesi, hesipos, posapci, apcineg, negprobe, probealprazolam, alprazolamreserpine, reserpineion, ionconfirming, confirmingquantifying, quantifyingcrashed, crashedpolarity, polarityeluting, elutingloading, loadingchloramphenicol