Determination of Hydrofluoric Acid by Aluminium Titration
Applications | | MetrohmInstrumentation
Hydrofluoric acid is a critical etching reagent in semiconductor and materials processing. Accurate quantification of HF in industrial etchant mixtures ensures process safety, product quality and regulatory compliance. Thermometric titration using aluminum nitrate provides a rapid, sensitive and selective approach for HF determination in complex acidic matrices.
This application note presents a method for measuring HF concentration in mixed acid etchant solutions. The primary goals are to:
The technique relies on an exothermic precipitation reaction at pH 4.5 where Al³⁺ reacts with F⁻ in the presence of Na⁺ and K⁺ to form insoluble NaK₂AlF₆. Endpoint detection is achieved thermometrically by monitoring solution temperature change.
Reagents:
Key titration settings:
Sample preparation involves diluting approximately 30 mL of etchant, neutralizing to faint pink with NaOH, adding buffer reagent and initiating the automated titration.
Analysis of a synthetic etch solution yielded an average HF concentration of 1.78 % w/w with a standard deviation of 0.01 % (n=8). The exothermic temperature rise and its second derivative provided a clear and reproducible endpoint. The precipitation of NaK₂AlF₆ ensured high selectivity for fluoride in the presence of strong acids.
Advances in sensor miniaturization and data analytics may further improve thermometric endpoint detection. Integration with online process monitoring could enable real-time HF control. Adaptation of this method to continuous flow or microfluidic platforms may expand its use in high-throughput laboratories.
The thermometric titration method using Al(NO₃)₃ titrant provides a precise, selective and efficient approach for determining hydrofluoric acid in mixed etchant solutions. Its robust performance and ease of automation make it well suited for industrial QA/QC applications.
Thermo Fisher Scientific. Application Note No. H-087: Determination of Hydrofluoric Acid by Aluminium Titration.
Titration
IndustriesEnergy & Chemicals
ManufacturerMetrohm
Summary
Importance of the topic
Hydrofluoric acid is a critical etching reagent in semiconductor and materials processing. Accurate quantification of HF in industrial etchant mixtures ensures process safety, product quality and regulatory compliance. Thermometric titration using aluminum nitrate provides a rapid, sensitive and selective approach for HF determination in complex acidic matrices.
Study objectives and overview
This application note presents a method for measuring HF concentration in mixed acid etchant solutions. The primary goals are to:
- Develop a reliable thermometric titration protocol using standard Al(NO₃)₃ titrant.
- Establish operating parameters that yield precise and reproducible endpoints.
- Demonstrate method performance on a synthetic etch solution.
Methodology and instrumentation
The technique relies on an exothermic precipitation reaction at pH 4.5 where Al³⁺ reacts with F⁻ in the presence of Na⁺ and K⁺ to form insoluble NaK₂AlF₆. Endpoint detection is achieved thermometrically by monitoring solution temperature change.
Instrumentation used
- Thermometric titrator head (Model 859 Titrotherm)
- Magnetic stirrer unit (Model 802)
Reagents and operating parameters
Reagents:
- 1 mol/L Al(NO₃)₃ titrant, standardized against anhydrous NaF.
- Ionic strength and buffer reagent: mixture of potassium acetate, sodium acetate and glacial acetic acid.
- 5 mol/L NaOH for sample neutralization.
- 0.1 % phenolphthalein indicator in ethanol.
Key titration settings:
- Titrant flow rate: 2 mL/min
- Single exothermic endpoint
- Data smoothing factor: 45
- Stirring speed: setting 10 on 802 stirrer
- Equilibration delay before titration: 60 s
Sample preparation involves diluting approximately 30 mL of etchant, neutralizing to faint pink with NaOH, adding buffer reagent and initiating the automated titration.
Key results and discussion
Analysis of a synthetic etch solution yielded an average HF concentration of 1.78 % w/w with a standard deviation of 0.01 % (n=8). The exothermic temperature rise and its second derivative provided a clear and reproducible endpoint. The precipitation of NaK₂AlF₆ ensured high selectivity for fluoride in the presence of strong acids.
Benefits and practical applications
- Rapid and automated analysis reduces operator time and potential errors.
- High selectivity avoids interference from other acid components.
- Suitable for routine QA/QC in semiconductor etching and metal finishing industries.
- Minimal sample preparation and robust buffer system ensure reproducibility.
Future trends and potential applications
Advances in sensor miniaturization and data analytics may further improve thermometric endpoint detection. Integration with online process monitoring could enable real-time HF control. Adaptation of this method to continuous flow or microfluidic platforms may expand its use in high-throughput laboratories.
Conclusion
The thermometric titration method using Al(NO₃)₃ titrant provides a precise, selective and efficient approach for determining hydrofluoric acid in mixed etchant solutions. Its robust performance and ease of automation make it well suited for industrial QA/QC applications.
References
Thermo Fisher Scientific. Application Note No. H-087: Determination of Hydrofluoric Acid by Aluminium Titration.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Analysis of zirconium acetate
|Metrohm|Applications
Thermo. Titr. Application Note No. H-092 Title: Analysis of zirconium acetate Scope: Automated determination of the zirconium content of zirconium acetate, as well as other zirconium compounds which can be rendered soluble as zirconium acetate. Principle: An aliquot of zirconium…
Key words
zirconium, zirconiumdose, dosekoac, koactitration, titrationnaoac, naoachoac, hoacbuffer, bufferacetate, acetatesolution, solutionendpoints, endpointspre, precurve, curvesmoothing, smoothingaliquot, aliquotderivative
Determination of Ferric and Cupric Ions in Copper Refining Solutions
|Metrohm|Applications
Thermo. Titr. Application Note No. H-070 Title: Determination of Ferric and Cupric Ions in Copper Refining Solutions Scope: Determination of Fe 3+ and Cu2+ in copper refining solutions by thermometric titration. It was found that the conventional approach of masking…
Key words
dsf, dsftitration, titrationsmoothing, smoothingimediately, imediatelycupric, cuprichead, headtitre, titreferric, ferricfactor, factorglacial, glacialrefining, refiningtitrant, titrantpermit, permitclicking, clickingcopper
Determination of Aluminium by Fluoride Titration
|Metrohm|Applications
Thermo. Titr. Application Note No. H-053 Title: Determination of Aluminium by Fluoride Titration Scope: Determination of aluminium in acidic, basic and neutral solutions; including alum, etching solutions and aluminate solutions. Principle: Titration with standard NaF solution in the presence of…
Key words
alum, alumtitration, titrationsolution, solutionaluminate, aluminatebayer, bayernaf, nafliquor, liquoraluminium, aluminiumdsf, dsfendpoints, endpointscurve, curvewaiting, waitingsmoothing, smoothingaliquots, aliquotsspent
Standardization of Sodium Fluoride for Aluminium Titrations
|Metrohm|Applications
Thermo. Titr. Application Note No. H-057 Title: Standardization of Sodium Fluoride for Aluminium Titrations Scope: Standardization of sodium fluoride titrant for determination of aluminium. Principle: Sodium fluoride titrant is standardized with a solution of aluminium prepared from high purity metal…
Key words
titrant, titrantstandardization, standardizationfluoride, fluoridenaf, nafaluminium, aluminiumsodium, sodiumtitration, titrationmmole, mmolebuffer, buffertitrate, titratedsf, dsftitrations, titrationsintercept, interceptsmoothing, smoothingstandardized