SEC and IEX for BioMolecules - Column Selection and Troubleshooting
Presentations | 2022 | Agilent TechnologiesInstrumentation
Size exclusion chromatography (SEC) and ion exchange chromatography (IEX) are essential techniques for the analysis of biomolecules under non-denaturing conditions. SEC separates molecules based on their hydrodynamic size, enabling aggregate and fragment analysis of monoclonal antibodies, polymers, and viral particles. IEX discriminates charge variants by exploiting electrostatic interactions between protein surface charges and charged stationary phases. Proper column selection, mobile phase design, and system setup are key to achieving high resolution, reproducibility, and minimal nonspecific interactions.
This application note provides a practical guide for selecting and troubleshooting SEC and IEX columns for biomolecular analysis. It reviews chromatographic mechanisms, outlines criteria for pore size and stationary-phase chemistry, and describes optimal mobile phase compositions and operating parameters. Case studies with protein standards and monoclonal antibodies illustrate resolution improvements. The document also offers troubleshooting tips for column care and instrument configuration.
SEC methodology relies on porous particles with carefully chosen pore diameters (100–2000 Å) to match analyte size ranges. Columns such as AdvanceBio SEC (130 Å, 300 Å) and Agilent Bio SEC (100 Å–2000 Å) minimize secondary interactions via hydrophilic coatings. Mobile phases typically consist of 50–150 mM phosphate buffers with 0–150 mM NaCl at pH 6.5–7.5. Operating conditions include flow rates of 0.1–2.0 mL/min (7.8 mm id) or 0.05–0.7 mL/min (4.6 mm id) at 20–30 °C.
IEX employs strong or weak anion (SAX/WAX) and cation (SCX/WCX) exchangers. Column families include Agilent Bio IEX (SAX, WAX, SCX, WCX), Bio mAb WCX, PL-SAX/PL-SCX (porous scale-up), and Bio-Monolith IEX (monolith). Gradient elution is achieved at constant pH (4.5–8.0) with increasing salt (400 mM–1 M NaCl). Typical LC systems: Agilent 1260 Infinity Bio-Inert LC with UV detection at 214–220 nm.
Successful biomolecule separations by SEC and IEX depend on matching column pore sizes and stationary-phase chemistries to analyte properties, designing mobile phases that suppress nonspecific interactions, and maintaining rigorous instrument and column care. Adoption of optimized gradients, particle technologies, and robust troubleshooting practices will support reliable analysis of complex biotherapeutics and biomacromolecules.
Consumables, LC columns, GPC/SEC
IndustriesManufacturerAgilent Technologies
Summary
Importance of the Topic
Size exclusion chromatography (SEC) and ion exchange chromatography (IEX) are essential techniques for the analysis of biomolecules under non-denaturing conditions. SEC separates molecules based on their hydrodynamic size, enabling aggregate and fragment analysis of monoclonal antibodies, polymers, and viral particles. IEX discriminates charge variants by exploiting electrostatic interactions between protein surface charges and charged stationary phases. Proper column selection, mobile phase design, and system setup are key to achieving high resolution, reproducibility, and minimal nonspecific interactions.
Objectives and Study Overview
This application note provides a practical guide for selecting and troubleshooting SEC and IEX columns for biomolecular analysis. It reviews chromatographic mechanisms, outlines criteria for pore size and stationary-phase chemistry, and describes optimal mobile phase compositions and operating parameters. Case studies with protein standards and monoclonal antibodies illustrate resolution improvements. The document also offers troubleshooting tips for column care and instrument configuration.
Methodology and Instrumentation
SEC methodology relies on porous particles with carefully chosen pore diameters (100–2000 Å) to match analyte size ranges. Columns such as AdvanceBio SEC (130 Å, 300 Å) and Agilent Bio SEC (100 Å–2000 Å) minimize secondary interactions via hydrophilic coatings. Mobile phases typically consist of 50–150 mM phosphate buffers with 0–150 mM NaCl at pH 6.5–7.5. Operating conditions include flow rates of 0.1–2.0 mL/min (7.8 mm id) or 0.05–0.7 mL/min (4.6 mm id) at 20–30 °C.
IEX employs strong or weak anion (SAX/WAX) and cation (SCX/WCX) exchangers. Column families include Agilent Bio IEX (SAX, WAX, SCX, WCX), Bio mAb WCX, PL-SAX/PL-SCX (porous scale-up), and Bio-Monolith IEX (monolith). Gradient elution is achieved at constant pH (4.5–8.0) with increasing salt (400 mM–1 M NaCl). Typical LC systems: Agilent 1260 Infinity Bio-Inert LC with UV detection at 214–220 nm.
Main Results and Discussion
- Pore size selection directly affects separation windows: 200 Å columns resolve antibody aggregates and fragments, while 300 Å columns suit intact mAbs and antibody–drug conjugates.
- Mobile phase pH and buffer concentration modulate peak symmetry and resolution for NISTmAb; optimal conditions around 150 mM phosphate at pH 6.8–7.4 yielded resolution >2.5 between dimer and monomer.
- IEX selectivity is governed by pH relative to protein pI: effective cation-exchange separations used pH 6.0–6.5, while anion exchange targeted pH 8.0; salt gradients of 0–750 mM NaCl achieved baseline separation of charge variants.
- Smaller particle sizes (1.7 µm) and shorter columns (50 mm) enable faster runs and higher efficiency but require optimized system tubing to minimize band broadening.
Benefits and Practical Applications of the Method
- High resolution analysis of aggregates, fragments, and charge variants improves biotherapeutic quality control.
- Hydrophilic column coatings and inert chemistries reduce sample loss from nonspecific interactions.
- Compatibility with MS allows seamless transition from UV-based profiling to mass analysis.
- Monolith and sub-2 µm phases facilitate high-throughput screening and process monitoring.
Future Trends and Applications
- Development of hybrid monolithic-particle columns for ultra-fast separations.
- Integration of multi-dimensional LC workflows combining SEC, IEX, and reversed-phase for comprehensive characterization.
- Automation of buffer preparation and inline filtration to enhance reproducibility.
- Enhanced bioinert hardware and consumables to extend column lifetime under high-salt conditions.
Conclusion
Successful biomolecule separations by SEC and IEX depend on matching column pore sizes and stationary-phase chemistries to analyte properties, designing mobile phases that suppress nonspecific interactions, and maintaining rigorous instrument and column care. Adoption of optimized gradients, particle technologies, and robust troubleshooting practices will support reliable analysis of complex biotherapeutics and biomacromolecules.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
A Chromatographic Separation of Biological Macromolecules
2022|Agilent Technologies|Presentations
A Chromatographic Separation of Biological Macromolecules Mohit P. Patel Applications Engineer LC Columns and Consumables Technical Support September 15, 2022 DE95811172 Agilent Restricted Outline LC techniques for biomolecule analysis Reversed phase • Commonly used, wide column choice and denaturing conditions.…
Key words
advancebio, advancebiobio, biorestricted, restrictedzorbax, zorbaxsax, saxmonolith, monolithagilent, agilentplrp, plrpreversed, reversedrrhd, rrhdphase, phasemab, mabmapping, mappingsec, secglycan
There is More to HPLC Than Reverse Phase - Column choices: Have you thought about the rest?
2019|Agilent Technologies|Presentations
There is More to HPLC Than Reverse Phase Column choices: Have you thought about the rest? Jean Lane Application Engineer December 12, 2019 DE.3031944444 Agilent Restricted Why Not to Use Reverse Phase Some reasons for trying an alternative C18 columns…
Key words
restricted, restrictedchiral, chiralagilent, agilentmonolith, monolithbio, bioporoshell, poroshellhic, hicprotein, proteiniex, iexphase, phaseadvancebio, advancebioaffinity, affinityzorbax, zorbaxexchange, exchangeinteractions
A Protein’s Journey
2020|Agilent Technologies|Presentations
A Protein’s Journey Rita K. Steed LC Application Engineer June 11, 2020 1 July 2, 2020 A Protein’s Journey DE.5077777778 Outline LC techniques for biomolecule analysis Reversed phase • Commonly used, wide column choice denaturing conditions. Can be used for…
Key words
journey, journeyprotein, proteinadvancebio, advancebiointeraction, interactionpore, poremedia, mediaamino, aminomab, mabphase, phasesize, sizespent, spentbuffer, bufferhic, hicvolume, volumeinterstitial
Agilent BioHPLC Column Selection Guide
2015|Agilent Technologies|Guides
For more information Learn more: www.agilent.com/chem/biohplc Buy online: www.agilent.com/chem/store U.S. and Canada 1-800-227-9770 [email protected] Agilent BioHPLC Column Selection Guide YOUR REFERENCE GUIDE TO THE ANALYSIS OF BIOPHARMACEUTICALS AND BIOMOLECULES Europe [email protected] Asia Pacific [email protected] This information is subject to change…
Key words
biohplc, biohplcbio, bioprotein, proteincolumns, columnsplrp, plrphplc, hplcsize, sizeproteins, proteinsadvancebio, advancebioanalysis, analysisuhplc, uhplcagilent, agilentzorbax, zorbaxpeptide, peptidemapping