LCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Shim-pack GIST/GISS Series INSTRUCTION MANUAL

Manuals |  | ShimadzuInstrumentation
Consumables, LC columns
Industries
Manufacturer
Shimadzu

Summary

Significance of the topic


Modern HPLC demands columns that deliver consistent retention, high resolution, and long lifetimes. Shim-pack GIST GISS series columns address these needs across both reversed phase and normal phase separations.

Objectives and overview


This manual provides comprehensive instructions on installation, operation, and maintenance of Shim-pack GIST GISS columns to ensure optimal chromatographic performance and extended column life.

Instrumentation Used


  • Shim-pack GIST and GISS columns: C18, C18 AQ, C8, Phenyl, Phenyl Hexyl, NH2
  • PEEK tubing (0.25–0.3 mm ID) or stainless steel tubing for UHPLC (0.1–0.2 mm ID)
  • Column fittings: male nuts, ferrules, UHPLC specific fittings
  • Guard columns, 0.45 μm membrane filters, and Ghost Trap DS for particulate removal
  • HPLC or UHPLC system capable of operating pressures up to 80 MPa (analytical) and 130 MPa (prep)

Methodology and Instrumentation


Proper column installation requires matching the flow direction labeled on the column, minimizing dead volume with short, correct diameter tubing, and using compatible fittings. Columns undergo rigorous QC tests for reagent purity, packing uniformity, and chromatographic performance. Operating pH ranges (C18 and variants: pH1–10, Phenyl pH1–10, NH2 pH2–7.5) and temperatures (up to 60 °C at pH 2–7.5) vary by phase. Pressure limits depend on particle size and column length, with analytical columns rated to 50 MPa and UHPLC to 80 MPa.

Main results and discussion


Key considerations include avoiding dead volumes to prevent early peak tailing, degassing eluents to eliminate bubbles in the flow cell, and preventing column clogging from sample particulates and pump seal debris. Recommendations include pre filtration of mobile phase and samples, use of guard columns, and regular flushing with mutually miscible solvents when switching solvent polarity or storing columns. Proper valve operation speed is crucial to avoid pressure spikes.

Benefits and practical applications


Shim-pack GIST GISS columns deliver high resolution and reproducibility for pharmaceutical, bioanalytical, environmental, and food testing. Their broad pH and temperature compatibility simplifies method development and routine QA QC operations, extending instrument uptime.

Future trends and applications


Emerging advances focus on sub 2 μm particles for faster UHPLC separations at higher pressures, integration of online sample preparation devices, and greener mobile phase systems to reduce solvent consumption and waste.

Conclusion


Following the recommended installation, operating, and storage protocols for Shim-pack GIST GISS columns ensures stable baselines, sharp peaks, and extended column lifetimes, meeting diverse analytical demands.

References


  • Shimadzu Corporation Instruction Manual ERAD-0001-9007: Shim-pack GIST GISS Series Columns

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
High Performance Packed Column for HPLC - Scepter Series
ERAD_0001-9043A High Performance Packed Column for HPLC Shim-pack Scepter Series INSTRUCTION MANUAL ■ Introduction The column should be connected with male nuts. Ensure that the To maintain and maximize peak performance of Shim-pack Scepter series columns, and to ensure the…
Key words
scepter, sceptershim, shimpack, packcolumn, columnprecautions, precautionspfpp, pfppphenyl, phenylpressure, pressureguard, guardeluent, eluentperformance, performancecolumns, columnsinstalling, installingvalve, valvepreparative
Shim-pack HPLC Column Guidebook
C190-E206D High Performance Liquid Chromatography Columns Shim-pack HPLC Column Guidebook Shimadzu Liquid Chromatographs Nexera X3 model Nexera X3 is Shimadzu’s new flagship UHPLC model. Nexera X3 excels in high throughput environments requiring fast separations. The system easily handles small ID…
Key words
shim, shimpack, packcolumns, columnsparticle, particlesize, sizehplc, hplcuhplc, uhplccolumn, columnguidebook, guidebooklength, lengthseries, seriesyes, yesphase, phaseguard, guardmpa
Shim-pack HPLC Column Guidebook
C190-E206E High Performance Liquid Chromatography Columns Shim-pack HPLC Column Guidebook ™ Shimadzu Liquid Chromatographs Nexera X3 model Nexera X3 is Shimadzu’s new flagship UHPLC model. Nexera X3 excels in high throughput environments requiring fast separations. The system easily handles small…
Key words
shim, shimpack, packcolumns, columnsparticle, particlesize, sizehplc, hplcuhplc, uhplclength, lengthcolumn, columnguidebook, guidebookyes, yesmpa, mpaseries, seriesguard, guardphase
Separation Characteristics of the Shim-pack Reversed Phase Column Series
C190-E305 Separation Characteristics of the Shim-pack Reversed Phase Column Series Technical Report Reversed Phase Columns C4/C8/C18 Tomoaki Shimpo1 A b stra c t: Reversed phase mode is the most frequently used method of separation in high-performance liquid chromatography. This method…
Key words
responsiveness, responsivenesshydrophobic, hydrophobicshim, shimpack, packdifferences, differencescapacity, capacitybonds, bondsretention, retentionyes, yesbase, basehydrogen, hydrogenreversed, reversedphase, phaseods, odssize
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike