Epoxide equivalents in epoxy resin

Applications | 2020 | MetrohmInstrumentation
Titration
Industries
Energy & Chemicals
Manufacturer
Metrohm

Summary

Importance of the Topic


The epoxy equivalent weight (EEW) of resin formulations directly impacts their reactivity, curing behavior and the mechanical properties of resulting coatings or composites. Accurate determination of epoxide content is essential for quality control in manufacturing and ensures consistent performance in applications ranging from adhesives to protective coatings.

Objectives and Study Overview


This study presents a robust analytical protocol for quantifying epoxy equivalents in resins according to EN ISO 3001 and ASTM D1652. The goal is to demonstrate the advantages of automated potentiometric titration over manual approaches, focusing on reproducibility, precision and ease of operation.

Methodology and Instrumentation


The analysis is based on the reaction of in-situ generated hydrobromic acid (HBr) with epoxy groups. Key steps include:
  • Dissolution of the sample in chloroform or methylene chloride.
  • Addition of tetraethylammonium bromide (TEABr) solution and glacial acetic acid.
  • Titration with standardized perchloric acid, using a nonaqueous potentiometric endpoint detection.

Used Instrumentation:
  • 905 Titrando system with magnetic stirrer
  • Solvotrode easyClean electrode for nonaqueous titrations
  • 900 Touch Control for user interface

Main Results and Discussion


The automated titration provided clear, well-defined titration curves and highly reproducible results. For a two-component adhesive resin, the determined EEW was 186.35 (n=3) with a relative standard deviation of 0.98%. The precision achieved highlights the benefit of automated endpoint detection and consistent stirring conditions.

Benefits and Practical Applications


  • High reproducibility and accuracy suitable for routine quality control
  • Elimination of manual endpoint estimation reduces operator bias
  • Rapid analysis with minimal sample preparation
  • Compliance with international standards EN ISO 3001 and ASTM D1652

Future Trends and Potential Applications


Emerging developments may include inline monitoring of epoxy content during production, miniaturized titration cells for reduced reagent consumption and integration with laboratory information management systems (LIMS). Advances in green solvents and reagent recycling could further enhance sustainability.

Conclusion


Automated potentiometric titration using the 905 Titrando and Solvotrode easyClean provides a straightforward, reliable and standard-compliant method for determining epoxy equivalent weight. The high precision and operational convenience make it ideal for industrial QC laboratories.

Reference


  • EN ISO 3001: Plastics – Determination of epoxide equivalent weight.
  • ASTM D1652: Standard Test Method for Epoxy Equivalent Weight by Potentiometric Titration.
  • Metrohm AG Application Note AN-T-136: Epoxide equivalents in epoxy resin.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Isocyanate content of polyurethane raw materials
AN-T-167 Isocyanate content of polyurethane raw materials Titration according to EN ISO 14896 for unsaturated polyester resin and polyurethane resin Summary Polyurethane (PU) is a class of very important polymers due to its flexibility and insulating properties. It is used…
Key words
polyurethane, polyurethanemetrohm, metrohmupr, uprpolyester, polyesterresin, resinunsaturated, unsaturatedpur, purusb, usbelectrode, electrodenco, ncoeasyclean, easycleanitrode, itroderaw, rawisocyanate, isocyanatestirrers
Nicotine content in tobacco
Nicotine content in tobacco
2020|Metrohm|Applications
AN-T-175 Nicotine content in tobacco Cost-efficient method using non-aqueous titration Summary Nicotine is an N-containing alkaloid that stimulates the parasympathetic nervous system, and is both highly addictive and dangerous to human health. Its toxic threshold value lies between 30–60 mg…
Key words
nicotine, nicotinemetrohm, metrohmtobacco, tobaccotitration, titrationchloride, chlorideselective, selectiveperchloric, perchloricomnis, omnisglacial, glacialbarium, bariumrel, reltitrator, titratorequivalence, equivalencesituations, situationsleaves
Automated titration of the hydroxyl number according to ASTM E1899 and EN ISO 4629-2
Application Bulletin 322/2 e Automated titration of the hydroxyl number according to ASTM E1899 and EN ISO 4629-2 Branch General analytical chemistry; organic chemistry; plastics Keywords Titration; hydroxyl number; automation; potentiometric titration; polyols; resins; paints; lubricants; ASTM E1899; EN ISO…
Key words
hydroxyl, hydroxyltitrant, titranttitration, titrationtsi, tsititer, titernumber, numberastm, astmdet, detequivalence, equivalenceaccording, accordingsample, samplesolvotrode, solvotrodeuntil, untildetermination, determinationweight
Determination of carbonyl compounds by potentiometric titration
Application Bulletin AB-042/2 e Determination of carbonyl compounds by potentiometric titration Branch General analytical chemistry; organic petrochemistry, biofuels; food, stimulants, flavors; paints, lacquers, solvents chemistry; beverages, Method 1 – Water soluble carbonyl compounds Instruments  Titrator with DET U and…
Key words
carbonyl, carbonyltitration, titrationpotentiometric, potentiometrictitrant, titranttiter, titerblank, blankequivalence, equivalencesolution, solutiondet, detreaction, reactiondetermination, determinationsample, samplecompounds, compoundsuntil, untilconsumption
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike