Clarity Control Module RECIPE EC6000
Manuals | 2024 | DataApexInstrumentation
The Recipe EC6000 amperometric detector provides high sensitivity for electrochemical detection in liquid chromatography. Integrating this detector with Clarity chromatography station streamlines workflow, enhances data fidelity, and centralizes instrument control, which is essential for quality-focused laboratories engaged in pharmaceutical, environmental and food safety analyses.
This manual guides users through installation, configuration and operation of the Recipe EC6000 control module within the Clarity software environment. It explains how to establish serial communication, configure detector parameters, monitor performance and integrate method settings into chromatographic workflows.
The Recipe EC6000 detector communicates via an RS-232 serial interface. Required hardware includes a straight DB9F-DB9M cable and gender changer. Installation involves adding the Recipe EC6000 driver in Clarity’s System Configuration, assigning the detector to an instrument tab and performing AutoDetect on the chosen COM port. Detector parameters such as working potential (±2000 mV), current measurement range (10 pA to 200 nA), digital filter settings (0.01 to 5 Hz) and offset adjustments are defined in the Method Setup – Acquisition dialog. Hardware Configuration and Device Monitor windows allow verification of communication settings and manual control functions (Autozero, Reset). Troubleshooting is supported by enabling detailed COM port logging via the COMMDRV.INI file.
The manual demonstrates successful integration of the Recipe EC6000 detector into the Clarity software, enabling full remote control of detector settings and capturing of method details alongside chromatographic data. The step-by-step configuration ensures reliable communication and reproducible baseline performance. The approach reduces manual intervention and errors by embedding autozero routines and offset calibration directly into analytical methods.
Ongoing developments may include expanded digital I/O for synchronization with autosamplers, enhanced scripting support for custom detector functions, remote web-based monitoring dashboards and integration with laboratory information management systems (LIMS). Advancements in digital communication standards could further improve data throughput and reliability.
The Recipe EC6000 Control Module manual provides a comprehensive framework for configuring and operating the amperometric detector within Clarity. By following the outlined procedures, laboratories can achieve robust detector control, traceable method documentation and streamlined analytical processes.
No external references were listed in the original document.
Software
IndustriesOther
ManufacturerRECIPE, DataApex
Summary
Significance of the topic
The Recipe EC6000 amperometric detector provides high sensitivity for electrochemical detection in liquid chromatography. Integrating this detector with Clarity chromatography station streamlines workflow, enhances data fidelity, and centralizes instrument control, which is essential for quality-focused laboratories engaged in pharmaceutical, environmental and food safety analyses.
Aims and overview of the manual
This manual guides users through installation, configuration and operation of the Recipe EC6000 control module within the Clarity software environment. It explains how to establish serial communication, configure detector parameters, monitor performance and integrate method settings into chromatographic workflows.
Methodology and instrumentation
The Recipe EC6000 detector communicates via an RS-232 serial interface. Required hardware includes a straight DB9F-DB9M cable and gender changer. Installation involves adding the Recipe EC6000 driver in Clarity’s System Configuration, assigning the detector to an instrument tab and performing AutoDetect on the chosen COM port. Detector parameters such as working potential (±2000 mV), current measurement range (10 pA to 200 nA), digital filter settings (0.01 to 5 Hz) and offset adjustments are defined in the Method Setup – Acquisition dialog. Hardware Configuration and Device Monitor windows allow verification of communication settings and manual control functions (Autozero, Reset). Troubleshooting is supported by enabling detailed COM port logging via the COMMDRV.INI file.
Used Instrumentation
- Recipe EC6000 amperometric detector (DataApex)
- Clarity chromatography station with LC/GC Control module
- RS-232 serial cable DB9F-DB9M and DB9F-DB9F gender changer
- PC serial COM port or MultiCOM adapter
Main results and discussion
The manual demonstrates successful integration of the Recipe EC6000 detector into the Clarity software, enabling full remote control of detector settings and capturing of method details alongside chromatographic data. The step-by-step configuration ensures reliable communication and reproducible baseline performance. The approach reduces manual intervention and errors by embedding autozero routines and offset calibration directly into analytical methods.
Benefits and practical applications
- Centralized control of detector parameters within Clarity methods
- Automated baseline zeroing and real-time monitoring
- Custom naming for multi-detector setups
- Automatic inclusion of detector settings in method reports
- Enhanced troubleshooting via COM port logging
Future trends and opportunities
Ongoing developments may include expanded digital I/O for synchronization with autosamplers, enhanced scripting support for custom detector functions, remote web-based monitoring dashboards and integration with laboratory information management systems (LIMS). Advancements in digital communication standards could further improve data throughput and reliability.
Conclusion
The Recipe EC6000 Control Module manual provides a comprehensive framework for configuring and operating the amperometric detector within Clarity. By following the outlined procedures, laboratories can achieve robust detector control, traceable method documentation and streamlined analytical processes.
Reference
No external references were listed in the original document.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Clarity Control Module SHODEX RI-101
2024|DataApex|Manuals
SHODEX RI-101 Clarity Control Module ENG Code/Rev.: M159/90C Date: 2024-02-14 DataApex Ltd. Phone: +420 251 013 400 Petrzilkova 2583/13 [email protected] 158 00 Prague 5 www.dataapex.com Czech Republic ® are trademarks of DataApex Ltd. Microsoft® and WindowsTM are Clarity®, DataApex® and…
Key words
clarity, claritydataapex, dataapexdetector, detectoruni, uniautozero, autozerodialog, dialogdevice, devicedet, detoutput, outputsetup, setupsignals, signalstab, tabµriu, µriucommunication, communicationautodetect
Clarity Control Module SCHAMBECK RI2012
2024|DataApex|Manuals
SCHAMBECK RI2012 Clarity Control Module ENG Code/Rev.: M182/90C Date: 2024-02-14 DataApex Ltd. Phone: +420 251 013 400 Petrzilkova 2583/13 [email protected] 158 00 Prague 5 www.dataapex.com Czech Republic ® are trademarks of DataApex Ltd. Microsoft® and WindowsTM are Clarity®, DataApex® and…
Key words
uni, unidataapex, dataapexclarity, claritydetector, detectortemperature, temperaturedialog, dialogsetup, setupautodetect, autodetectschambeck, schambeckdevice, devicesetting, settingautodetected, autodetectedshutdown, shutdownset, setpolarity
Clarity Control Module ERC REFRACTOMAX 520
2024|DataApex|Manuals
ERC REFRACTOMAX 520 Clarity Control Module ENG Code/Rev.: M190/90A Date: 2024-05-14 DataApex Ltd. Phone: +420 251 013 400 Petrzilkova 2583/13 [email protected] 158 00 Prague 5 www.dataapex.com Czech Republic ® are trademarks of DataApex Ltd. Microsoft® and WindowsTM are Clarity®, DataApex®…
Key words
clarity, clarityuni, unidataapex, dataapexdetector, detectordialog, dialogerc, ercautozero, autozerosetup, setupdet, detoutput, outputsignals, signalsdevice, devicetab, tabµriu, µriucommunication
Clarity Control Module KONTRON 432
2024|DataApex|Manuals
KONTRON 432 Clarity Control Module ENG Code/Rev.: M181/90C Date: 2024-02-14 DataApex Ltd. Phone: +420 251 013 400 Petrzilkova 2583/13 [email protected] 158 00 Prague 5 www.dataapex.com Czech Republic ® are trademarks of DataApex Ltd. Microsoft® and WindowsTM are Clarity®, DataApex® and…
Key words
clarity, clarityaufs, aufssetup, setupdataapex, dataapexdialog, dialoguni, uniaux, auxinvoked, invokeddevice, devicewavelength, wavelengthcommunication, communicationautozero, autozeroautodetect, autodetectbutton, buttontable