Clarity Control Module SCHAMBECK RI2012

Manuals | 2024 | DataApexInstrumentation
Software
Industries
Other
Manufacturer
DataApex

Summary

Importance of the topic


Integrating a Schambeck RI2012 refractive index detector into a chromatography station ensures precise, automated control over key parameters such as temperature, signal polarity and zeroing routines.
Reliable detector management supports reproducible analyses in QA/QC, research and industrial laboratories.

Objectives and overview of the module


The module enables direct communication between Clarity software and Schambeck RI2012 (and related) detectors via RS-232.
It provides full control of detector settings from method files, records method parameters in the chromatogram header, and supports synchronized start and shutdown events.

Methodology and used instrumentation


The control module relies on the DataApex UNI Ruby driver script and a serial DB9F-DB9M cable for RS-232 communication.
Key steps include:
  • Adding the UNI Ruby control module in Clarity’s System Configuration.
  • Selecting the proper COM port and auto-detecting the detector.
  • Assigning digital inputs/outputs for external start and ready signals.
  • Configuring acquisition parameters such as sample rate, temperature setpoints, autozero and signal polarity.

Main results and discussion


The manual guides the user through:
  • Installation of the UNI driver script and hardware connections.
  • Clarity configuration to add the RI2012 as an LC or GC control module.
  • Method Setup screens for acquisition and advanced auxiliary signals.
  • Real-time monitoring of detector status, temperature, optical balance and output in the Device Monitor window.
  • Report generation with embedded detector method parameters.
  • Troubleshooting via COMMDRV.INI logging for diagnostic trace files.

Benefits and practical applications


Using this module streamlines routine HPLC analyses requiring refractive index detection by:
  • Centralizing instrument control within Clarity methods.
  • Ensuring consistent temperature and zeroing routines to stabilize baselines.
  • Embedding detector settings in data files for audit trails.
  • Facilitating automation with external start signals from autosamplers.

Future trends and possibilities for use


Advances may include network-enabled detectors for remote monitoring, cloud-based data logging, and tighter integration with laboratory information management systems (LIMS).
Enhanced scripting support could allow custom control sequences and improved error handling.

Conclusion


The Schambeck RI2012 Control Module provides robust, software-driven management of refractive index detectors in Clarity stations.
It enhances reproducibility, simplifies reporting and supports automated workflows critical for modern analytical laboratories.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Clarity Control Module ERC REFRACTOMAX 520
ERC REFRACTOMAX 520 Clarity Control Module ENG Code/Rev.: M190/90A Date: 2024-05-14 DataApex Ltd. Phone: +420 251 013 400 Petrzilkova 2583/13 [email protected] 158 00 Prague 5 www.dataapex.com Czech Republic ® are trademarks of DataApex Ltd. Microsoft® and WindowsTM are Clarity®, DataApex®…
Key words
clarity, clarityuni, unidataapex, dataapexdetector, detectordialog, dialogerc, ercautozero, autozerosetup, setupdet, detoutput, outputsignals, signalsdevice, devicetab, tabµriu, µriuautodetect
Clarity Control Module SHODEX RI-101
SHODEX RI-101 Clarity Control Module ENG Code/Rev.: M159/90C Date: 2024-02-14 DataApex Ltd. Phone: +420 251 013 400 Petrzilkova 2583/13 [email protected] 158 00 Prague 5 www.dataapex.com Czech Republic ® are trademarks of DataApex Ltd. Microsoft® and WindowsTM are Clarity®, DataApex® and…
Key words
clarity, claritydataapex, dataapexdetector, detectoruni, uniautozero, autozerodialog, dialogdet, detdevice, deviceoutput, outputsetup, setupsignals, signalstab, tabµriu, µriucommunication, communicationautodetect
Clarity Control Module SHODEX RI-501
SHODEX RI-501 Clarity Control Module ENG Code/Rev.: M212/90C Date: 2024-02-14 DataApex Ltd. Phone: +420 251 013 400 Petrzilkova 2583/13 [email protected] 158 00 Prague 5 www.dataapex.com Czech Republic ® are trademarks of DataApex Ltd. Microsoft® and WindowsTM are Clarity®, DataApex® and…
Key words
clarity, claritydetector, detectordataapex, dataapexuni, unidevice, deviceautozero, autozerodialog, dialogdet, detoutput, outputsetup, setupsignals, signalstab, tabmonitor, monitorµriu, µriubutton
Clarity Control Module RECIPE EC6000
RECIPE EC6000 Clarity Control Module ENG Code/Rev.: M216/80B Date: 10/27/2022 DataApex Ltd. Phone: +420 251 013 400 [email protected] www.dataapex.com Petrzilkova 2583/13 158 00 Prague 5 The Czech Republic ® are trademarks of DataApex Ltd. Microsoft ® and Windows TM are…
Key words
autozero, autozerodialog, dialogdigital, digitaluni, uniinput, inputclarity, claritysetup, setupdetector, detectorport, portdevice, devicetab, tabautodetect, autodetectcommunication, communicationcontrol, controlruby
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike