Determination of transition metals by ion chromatography
Technical notes | 2017 | Thermo Fisher ScientificInstrumentation
Accurate determination of trace transition metals is critical in environmental monitoring, industrial quality control and biochemical research. Ion chromatography (IC) offers a sensitive and versatile approach to quantify multiple metal ions simultaneously. By employing chelating eluents and post-column derivatization, IC overcomes challenges associated with metal speciation and detection.
This study presents two IC methods for the simultaneous separation and quantification of common transition metals. Method A uses pyridine-2,6-dicarboxylic acid (PDCA) eluent to resolve eight cationic species, while Method B uses an oxalate buffer to separate five divalent metals. Both approaches utilize mixed-mode separation on a single IonPac CS5A column and post-column complexation with 4-(2-pyridylazo)resorcinol (PAR) for UV detection.
The IC separation is based on chelation of hydrated metal ions by anionic ligands, altering their net charge for cation or anion exchange:
Both methods require degassing and precaution against oxidation of eluent and PAR reagent.
Integration of high-pressure ion chromatography with mass spectrometry could further improve detection limits and speciation analysis. Automated sample preparation and miniaturized systems are expected to expand on-site monitoring capabilities. Development of new chelating eluents may allow broader coverage of metal species and oxidation states.
The described IC methods deliver robust, high-throughput quantification of key transition metals using a single mixed-mode column and post-column PAR derivatization. These approaches meet the demands of modern environmental and industrial laboratories by offering selectivity, sensitivity and operational simplicity.
Ion chromatography
IndustriesManufacturerThermo Fisher Scientific
Summary
Significance of the topic
Accurate determination of trace transition metals is critical in environmental monitoring, industrial quality control and biochemical research. Ion chromatography (IC) offers a sensitive and versatile approach to quantify multiple metal ions simultaneously. By employing chelating eluents and post-column derivatization, IC overcomes challenges associated with metal speciation and detection.
Objectives and overview
This study presents two IC methods for the simultaneous separation and quantification of common transition metals. Method A uses pyridine-2,6-dicarboxylic acid (PDCA) eluent to resolve eight cationic species, while Method B uses an oxalate buffer to separate five divalent metals. Both approaches utilize mixed-mode separation on a single IonPac CS5A column and post-column complexation with 4-(2-pyridylazo)resorcinol (PAR) for UV detection.
Methodology
The IC separation is based on chelation of hydrated metal ions by anionic ligands, altering their net charge for cation or anion exchange:
- Method A: 7.0 mM PDCA, 66 mM KOH, 74 mM formic acid, 5.6 mM K₂SO₄; flow 1.2 mL/min; injection 50 µL.
- Method B: 8.0 mM oxalic acid, 50 mM KOH, 100 mM tetramethylammonium hydroxide; flow 1.2 mL/min; injection 50 µL.
- Post-column derivatization: 0.5 mM PAR in diluent; reaction coil 375 µL; reagent flow 0.7 mL/min; detection at 530 nm.
Both methods require degassing and precaution against oxidation of eluent and PAR reagent.
Used Instrumentation
- Thermo Scientific Dionex DX 500 or ICS-5000+ HPIC system with GP40 Gradient Pump, AD20 Absorbance Detector, LC20 Enclosure, and PC10 Postcolumn Pneumatic Delivery Package.
- Dionex IonPac CS5A analytical column with CG5A guard.
Main results and discussion
- Method A achieved baseline resolution of Fe³⁺, Cu²⁺, Ni²⁺, Zn²⁺, Co²⁺, Cd²⁺, Mn²⁺ and Fe²⁺ in under 15 minutes.
- Method B separated Pb²⁺, Cu²⁺, Co²⁺, Zn²⁺ and Ni²⁺ within 12 minutes; cadmium and manganese coeluted under these conditions.
- Iron accumulation in Method B necessitated periodic acid cleaning: 1 M HCl wash for 60 minutes followed by re-equilibration.
- Post-column PAR derivatization provided sensitive absorbance detection for metals not amenable to conductivity measurement.
Benefits and practical applications
- Simultaneous multi-metal analysis reduces sample throughput time.
- Mixed-mode separation on one column simplifies instrument setup.
- Post-column derivatization enhances detection sensitivity and selectivity.
- Applicable to environmental waters, industrial effluents and process streams.
Future trends and potential applications
Integration of high-pressure ion chromatography with mass spectrometry could further improve detection limits and speciation analysis. Automated sample preparation and miniaturized systems are expected to expand on-site monitoring capabilities. Development of new chelating eluents may allow broader coverage of metal species and oxidation states.
Conclusion
The described IC methods deliver robust, high-throughput quantification of key transition metals using a single mixed-mode column and post-column PAR derivatization. These approaches meet the demands of modern environmental and industrial laboratories by offering selectivity, sensitivity and operational simplicity.
Reference
- Thermo Fisher Scientific Technical Note TN72544-EN (2017).
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Determination of Transition Metals by Ion Chromatography
|Thermo Fisher Scientific|Applications
Technical Note 10 Determination of Transition Metals by Ion Chromatography INTRODUCTION Metal ions can exist in several different forms. The factors which determine the form of the metal ion are the extent of complexation and the oxidation state. In many…
Key words
metals, metalschelating, chelatingcomplexed, complexedmetpac, metpacpdca, pdcatransition, transitionhydrated, hydratedagents, agentsoxalic, oxalicligands, ligandsweakly, weaklypostcolumn, postcolumncomplexes, complexeseluent, eluentanionic
Ion Pac CS5A Column & MetPac Reagents
2001|Thermo Fisher Scientific|Technical notes
columns IonPac® CS5A Column & MetPac™ Reagents The IonPac CS5A is a high resolution ion-exchange column for the determination of transition and lanthanide metals. The MetPac PDCA and Oxalic Acid Eluent Concentrates are ready-todilute reagents that improve ease-of-use and ensure…
Key words
metpac, metpacpostcolumn, postcolumndiluent, diluentreagent, reagentchelation, chelationpdca, pdcaeluent, eluentmetals, metalstransition, transitioniron, ironcobalt, cobaltconcentrates, concentrateszinc, zincnickel, nickelcopper
Determination of Transition Metals in Serum and Whole Blood by Ion Chromatography
2001|Thermo Fisher Scientific|Applications
Application Note 108 Determination of Transition Metals in Serum and Whole Blood by Ion Chromatography INTRODUCTION The determination of transition metals in physiological fluids is of considerable interest in clinical chemistry. In recent years several studies have linked the concentrations…
Key words
metpac, metpacmetals, metalspostcolumn, postcolumneluent, eluenttransition, transitionpdca, pdcaoxalic, oxalicwhole, wholereagent, reagentserum, serumblood, bloodacid, acidphysiological, physiologicalzinc, zincdiluent
Determination of transition metals at ppt levels in high-purity water and SC2 (D-clean) baths
2017|Thermo Fisher Scientific|Applications
APPLICATION NOTE 131 Determination of transition metals at ppt levels in high-purity water and SC2 (D-clean) baths Authors Mark Laikhtman and Jeff Rohrer Thermo Fisher Scientific, Sunnyvale, CA Keywords Ion chromatography, IC, copper, zinc, absorbance, IonPac CS5A, iron, nickel, PAR,…
Key words
pdca, pdcadqp, dqpmetals, metalspar, parpostcolumn, postcolumntransition, transitiondionex, dionexconcentrator, concentratorminutes, minutespump, pumptubing, tubingreservoir, reservoirreagents, reagentsstandards, standardsoutlined