Determination of suppressor in acid copper baths by smartDT
Applications | | MetrohmInstrumentation
The precise measurement of suppressor concentration in acid copper plating baths is critical for maintaining deposit quality and process stability in electroplating industries. Traditional dilution titration methods require many reagent additions and prolonged analysis time. The smartDT approach optimizes addition steps, reducing both the number of operations and total assay duration while preserving analytical accuracy.
This application note evaluates a novel titration routine (smartDT) for rapid determination of suppressor levels in acid copper baths. The study compares classical dilution titration (DT) with smartDT regarding accuracy, number of additions, and total analysis time. Key performance metrics include recovery of a check standard and time saving per determination.
The method employs cyclic voltammetry semiautomated titration with a customized software algorithm to calculate variable addition volumes. The operator sets the largest and smallest addition steps; the software dynamically adjusts intermediate volumes based on the titration progress toward a fixed evaluation ratio (0.5). Peak potential is monitored at 0.2 V against a horizontal baseline.
Comparison of a 3 mL/L suppressor check standard yielded 3.02 mL/L by classic DT and 3.04 mL/L by smartDT (recoveries of 100.7% vs. 101.3%, respectively). The number of additions dropped from 14 to 8 with smartDT. Both workflows required 7 minutes for initial setup, but smartDT reduced titration time from 28 minutes to 16 minutes, delivering a total analysis time of 23 minutes versus 35 minutes—a 34% time saving.
smartDT demonstrates significant time savings and robust analytical performance for suppressor determination in acid copper baths. Its variable addition strategy streamlines titration workflows without compromising accuracy, offering a valuable tool for electroplating quality control laboratories.
Voltammetry/Coulometry
IndustriesEnergy & Chemicals
ManufacturerMetrohm
Summary
Significance of the Topic
The precise measurement of suppressor concentration in acid copper plating baths is critical for maintaining deposit quality and process stability in electroplating industries. Traditional dilution titration methods require many reagent additions and prolonged analysis time. The smartDT approach optimizes addition steps, reducing both the number of operations and total assay duration while preserving analytical accuracy.
Objectives and Study Overview
This application note evaluates a novel titration routine (smartDT) for rapid determination of suppressor levels in acid copper baths. The study compares classical dilution titration (DT) with smartDT regarding accuracy, number of additions, and total analysis time. Key performance metrics include recovery of a check standard and time saving per determination.
Methodology and Instrumentation
The method employs cyclic voltammetry semiautomated titration with a customized software algorithm to calculate variable addition volumes. The operator sets the largest and smallest addition steps; the software dynamically adjusts intermediate volumes based on the titration progress toward a fixed evaluation ratio (0.5). Peak potential is monitored at 0.2 V against a horizontal baseline.
Used Instrumentation
- 894 Professional CVS semiautomated titrator (viva).
- Rotating disk electrode (RDE) assembly.
- 2 mm platinum tip working electrode.
- Ag/AgCl/KCl (3 mol/L) reference electrode with KNO₃ bridge electrolyte (1 mol/L).
- Separate platinum rod counter electrode.
Key Results and Discussion
Comparison of a 3 mL/L suppressor check standard yielded 3.02 mL/L by classic DT and 3.04 mL/L by smartDT (recoveries of 100.7% vs. 101.3%, respectively). The number of additions dropped from 14 to 8 with smartDT. Both workflows required 7 minutes for initial setup, but smartDT reduced titration time from 28 minutes to 16 minutes, delivering a total analysis time of 23 minutes versus 35 minutes—a 34% time saving.
Benefits and Practical Applications
- Accelerated analysis throughput for routine bath control.
- Maintained or improved accuracy compared to classical methods.
- Lower reagent consumption and reduced operator workload.
- Easy integration into existing semiautomated systems.
Future Trends and Opportunities
- Expansion of smartDT algorithms to additional plating additives and bath chemistries.
- Integration with laboratory information management systems (LIMS) for real-time process feedback.
- Development of AI-enhanced titration control for predictive maintenance and adaptive sampling.
- Miniaturization and field-deployable versions for on-site quality control.
Conclusion
smartDT demonstrates significant time savings and robust analytical performance for suppressor determination in acid copper baths. Its variable addition strategy streamlines titration workflows without compromising accuracy, offering a valuable tool for electroplating quality control laboratories.
References
- VA Application Note V–202: Determination of suppressor in acid copper baths by smartDT.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Determination of suppressor with CVS using the calibration technique «smartDT» with dynamic addition volumes
|Metrohm|Applications
Application Bulletin 420/2 e Determination of suppressor with CVS using the calibration technique «smartDT» with dynamic addition volumes The Application Bulletin describes the determination of suppressor in acid copper baths by smartDT. The determination of suppressor with dilution titration (DT)…
Key words
smartdt, smartdtsuppressor, suppressorcvs, cvsaddition, additionvolume, volumevolumes, volumesdetermination, determinationcalibration, calibrationveff, veffvms, vmsviva, vivaevaluation, evaluationstandard, standardareanorm, areanormdynamic
Suppressor «Top Lucina α-M» in acid copper baths (Okuno Chemical Industries)
|Metrohm|Applications
VA Application Note No. V - 182 Title: Suppressor «Top Lucina α-M» in acid copper baths (Okuno Chemical Industries) Summary: Determination of suppressor «Top Lucina α-M» in acid copper baths by dilution titration (DT) using cyclic voltammetric stripping (CVS). Sample:…
Key words
lucina, lucinasuppressor, suppressorvertex, vertexcopper, coppertop, topelectrode, electrodebaths, bathscvs, cvspotential, potentialokuno, okunoacid, acidrde, rdeworking, workingelectroplating, electroplatingshaft
Suppressor «Cupraspeed» in acid copper baths (Atotech)
|Metrohm|Applications
VA Application Note No. V - 137 Title: Suppressor «Cupraspeed» in acid copper baths (Atotech) Summary: Determination of suppressor «Cupraspeed» in acid copper baths by dilution titration (DT) using cyclic voltammetric stripping (CVS). Sample: Acid copper electroplating bath Sample preparation:…
Key words
cupraspeed, cupraspeedsuppressor, suppressorvertex, vertexcopper, copperelectrode, electrodebaths, bathspotential, potentialatotech, atotechacid, acidrde, rdevoltammetric, voltammetricworking, workingelectroplating, electroplatingvms, vmscvs
Suppressor «Cupracid BL-CT» in acid copper baths (Atotech)
|Metrohm|Applications
VA Application Note No. V - 135 Title: Suppressor «Cupracid BL-CT» in acid copper baths (Atotech) Summary: Determination of suppressor «Cupracid BL-CT» in acid copper baths by dilution titration (DT) using cyclic voltammetric stripping (CVS). Sample: Acid copper electroplating bath…
Key words
cupracid, cupracidsuppressor, suppressorvertex, vertexcopper, copperelectrode, electrodebaths, bathspotential, potentialatotech, atotechacid, acidrde, rdeworking, workingelectroplating, electroplatingvms, vmscvs, cvsvirgin