Filter Validation Protocol - Agilent 850-DS Dissolution Sampling Station

Technical notes | 2013 | Agilent TechnologiesInstrumentation
Dissolution
Industries
Pharma & Biopharma
Manufacturer
Agilent Technologies

Summary

Importance of the Topic


Dissolution testing is a cornerstone of pharmaceutical development and quality control. Accurate filtration at the sampling stage ensures that undissolved particles and excipients do not interfere with downstream analytical methods, preserving column integrity and reliable quantification of the active pharmaceutical ingredient (API).

Objectives and Study Overview


This document outlines a validation protocol for the Whatman 850-DS 8-Channel Filter Plates used with the Agilent 850-DS Dissolution Sampling Station. The protocol aims to demonstrate that the multi-channel plates perform as effectively as individual 25 mm syringe filters in three critical aspects: filtration efficiency, leachability, and adsorption.

Methodology and Used Instrumentation


Methodology:
  1. Filter Selection: Choose filter material (PTFE, nylon, PES, or glass fiber) based on sample chemistry and analytical technique (HPLC or UV-Vis).
  2. Three Validation Tests:
    • Efficiency Test: Compare API recovery from 50 % standard with and without sonication; acceptance criterion ≤ 2 % difference.
    • Leachability Test: Analyze blank filtrates for UV-absorbing leachables; acceptance criterion ≤ 0.5 % of standard response.
    • Adsorption Test: Determine conditioning volume needed to achieve 98–102 % recovery; initial aliquots discarded accordingly.

Used Instrumentation:
  • Agilent 850-DS Dissolution Sampling Station with filter changer option.
  • Whatman 850-DS 8-Channel Filter Plates and corresponding 25 mm Puradisc syringe filters.
  • UV-Vis spectrophotometer or HPLC system for quantitative API analysis.

Main Results and Discussion


Validation confirmed that the multi-channel filter plates are functionally equivalent to individual syringe filters:
  • Efficiency: Removal of undissolved API achieved with ≤ 2 % variance after sonication.
  • Leachability: Blank filtrates showed no significant UV interference (< 0.5 % of standard).
  • Adsorption: After an initial discard volume, API recovery stabilized between 98 % and 102 %.

The compact plate design facilitates automated sample processing without altering internal filter components.

Benefits and Practical Applications


The validated filter plates offer:
  • High-throughput, automation-ready filtration.
  • Reduced manual handling and lower risk of cross-contamination.
  • Consistent protection of analytical columns from particulate damage.

Future Trends and Potential Applications


Potential developments include:
  • Integration with online dissolution platforms for real-time monitoring.
  • Application in screening of complex or poorly soluble formulations.
  • Customized membrane developments for specialized APIs.

Conclusion


The Agilent-GE Healthcare co-developed 8-Channel Filter Plates for the 850-DS station meet key validation criteria and can be adopted as equivalents to standard Puradisc filters, enhancing laboratory efficiency without compromising analytical quality.

Reference


No external literature references provided.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Merits of Automated Sampling in Dissolution Testing
Merits of Automated Sampling in Dissolution Testing
2014|Agilent Technologies|Technical notes
Merits of Automated Sampling in Dissolution Testing Technical Overview Introduction It is often said that there are four primary sources of potential error in any process, the so-called four Ms: Man, Method, Machine, and Materials. Any failure investigation must look…
Key words
dissolution, dissolutionsampling, samplingautosampler, autosamplertimepoints, timepointsdosage, dosagevessel, vesseltimepoint, timepointpaddle, paddleautosampling, autosamplinginitiates, initiateseach, eachmedia, mediaintegrity, integritysamples, samplescan
Guided Resource Map - Agilent 708-DS Dissolution Apparatus
Guided Resource Map Agilent 708-DS Dissolution Apparatus One platform for all dissolution needs The Agilent 708-DS Dissolution Apparatus provides a versatile and robust platform for standardizing dissolution testing in your lab. This instrument delivers consistent performance and results and complies…
Key words
sampling, samplingvolume, volumemedia, mediatimepoint, timepointdissolution, dissolutionlines, linespulling, pullingsample, sampleshafts, shaftsprime, primebasket, basketcollection, collectionminimum, minimumstops, stopstime
Cary 60 UV Dissolution Systems
Cary 60 UV Dissolution Systems
2012|Agilent Technologies|Brochures and specifications
Cary 60 UV Dissolution Systems THE STANDARD FOR MULTICELL OR FIBER OPTIC DISSOLUTION TESTING AGILENT CARY 60 UV-VIS SPECTROPHOTOMETER FIVE REASONS WHY THE CARY 60 SHOULD BE YOUR CHOICE FOR UV DISSOLUTION TESTING The Agilent Cary 60 UV-Visible Spectrophotometer provides…
Key words
dissolution, dissolutionoptic, opticfiber, fiberqualification, qualificationnewsletter, newsletterservices, servicesmulticell, multicellconfiguration, configurationddg, ddgonline, onlineworks, worksoffline, offlineflow, flowsystem, systemsupport
Titan3 and Target2 Chromatography Syringe Filters
Titan3 and Target2 Chromatography Syringe Filters
2018|Thermo Fisher Scientific|Brochures and specifications
Titan3 and Target2 Chromatography Syringe Filters Performance filtration solutions for chromatographic applications Sample preparation is a key stage in successful chromatography The high quality Thermo Scientific™ Titan3™ and Target2™ syringe filters ensure reliable elimination of both particles and micro organisms…
Key words
luer, luerpvdf, pvdfsyringe, syringefilters, filterstransparent, transparentcellulose, cellulosegmf, gmfpolypropylene, polypropylenemdl, mdlmembrane, membraneincompatibilities, incompatibilitiesnylon, nylonacetate, acetatebinding, bindingslip
Other projects
GCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike