Applications | 2007 | Agilent TechnologiesInstrumentation
The stability and performance of polypropylene homopolymers depend critically on phenolic antioxidants and slip additives such as erucamide. Routine monitoring of these compounds ensures product durability under heat, light and oxygen exposure. Rapid and reliable analysis supports quality control in polymer manufacturing and helps prevent material degradation.
This work set out to adapt the ASTM D6042 method for faster, high-throughput analysis on an Agilent 1200 Rapid Resolution LC (RRLC) system. Key goals included reducing run times and solvent use without sacrificing chromatographic resolution. The study also explored the use of a dedicated method translator to migrate conventional HPLC protocols to modern sub-2 µm columns.
A standard mixture containing Tinuvin P, erucamide, Irganox 3114, Irganox 1010, vitamin E, Irganox 1076 and Irgafos 168 was prepared at 200 µg/mL in isopropanol. Polypropylene homopolymer extracts were obtained via ultrasonic extraction with a 1:1 methylene chloride/cyclohexane solvent blend. Initial separations used a ZORBAX Eclipse XDB-C8 4.6×150 mm, 5 µm column under a water/acetonitrile gradient. Solvent incompatibilities at high injection volumes were addressed by reducing sample introduction to 5 µL.
Comparison of C18 and C8 chemistries revealed that C8 columns offered improved peak shape and retention under the selected mobile phase conditions. Use of the Agilent method translator enabled three conversion modes: simple slope-matched transfer, speed-optimized at maximum system pressure, and resolution-optimized with shallower gradient ramps. Transitioning from a 150 mm column to 50 mm sub-2 µm columns reduced analysis times from 10 min to under 3 min while preserving or enhancing resolution. Spiked polymer extracts achieved baseline separation of all target additives in as little as 1.4 min.
The optimized RRLC methods deliver:
Further advances are expected in ultra-high-pressure LC systems and sub-2 µm stationary phases, offering even shorter run times and higher peak capacities. Integration with mass spectrometry detectors will expand qualitative profiling capabilities. Machine-learning algorithms may soon automate gradient design and method translation, streamlining method development across diverse polymer formulations.
This study demonstrates that combining Agilent RRLC hardware with a method translation utility enables fast, high-resolution analysis of antioxidants and slip additives in polypropylene. The approach significantly improves laboratory productivity, reduces environmental impact, and maintains reliable quantitation in QA/QC workflows.
Consumables, HPLC, LC columns
IndustriesEnergy & Chemicals
ManufacturerAgilent Technologies
Summary
Significance of the topic
The stability and performance of polypropylene homopolymers depend critically on phenolic antioxidants and slip additives such as erucamide. Routine monitoring of these compounds ensures product durability under heat, light and oxygen exposure. Rapid and reliable analysis supports quality control in polymer manufacturing and helps prevent material degradation.
Objectives and Overview
This work set out to adapt the ASTM D6042 method for faster, high-throughput analysis on an Agilent 1200 Rapid Resolution LC (RRLC) system. Key goals included reducing run times and solvent use without sacrificing chromatographic resolution. The study also explored the use of a dedicated method translator to migrate conventional HPLC protocols to modern sub-2 µm columns.
Methodology
A standard mixture containing Tinuvin P, erucamide, Irganox 3114, Irganox 1010, vitamin E, Irganox 1076 and Irgafos 168 was prepared at 200 µg/mL in isopropanol. Polypropylene homopolymer extracts were obtained via ultrasonic extraction with a 1:1 methylene chloride/cyclohexane solvent blend. Initial separations used a ZORBAX Eclipse XDB-C8 4.6×150 mm, 5 µm column under a water/acetonitrile gradient. Solvent incompatibilities at high injection volumes were addressed by reducing sample introduction to 5 µL.
Used Instrumentation
- Agilent 1200 Series RRLC: G1379B degasser, G1312B binary pump SL, G1367C autosampler SL, G1316B column compartment, G1315C DAD detector.
- ZORBAX RRHT columns (Eclipse XDB-C18 and XDB-C8) in various formats: 4.6×150 mm (5 µm), 4.6×100 mm (3.5 µm), 4.6×50 mm (1.8 µm), 3.0×100 mm (3.5 µm) and 3.0×50 mm (1.8 µm).
- UV/VIS diode array detection at 200 nm with a 3 mm, 2 µL flow cell.
Main Results and Discussion
Comparison of C18 and C8 chemistries revealed that C8 columns offered improved peak shape and retention under the selected mobile phase conditions. Use of the Agilent method translator enabled three conversion modes: simple slope-matched transfer, speed-optimized at maximum system pressure, and resolution-optimized with shallower gradient ramps. Transitioning from a 150 mm column to 50 mm sub-2 µm columns reduced analysis times from 10 min to under 3 min while preserving or enhancing resolution. Spiked polymer extracts achieved baseline separation of all target additives in as little as 1.4 min.
Benefits and Practical Applications
The optimized RRLC methods deliver:
- Up to tenfold faster analyses compared to conventional HPLC.
- Lower solvent consumption and waste generation.
- Seamless migration of legacy methods via an intuitive translator tool.
- Robust detection of low-level antioxidants and slip additives in complex polymer matrices.
Future Trends and Applications
Further advances are expected in ultra-high-pressure LC systems and sub-2 µm stationary phases, offering even shorter run times and higher peak capacities. Integration with mass spectrometry detectors will expand qualitative profiling capabilities. Machine-learning algorithms may soon automate gradient design and method translation, streamlining method development across diverse polymer formulations.
Conclusion
This study demonstrates that combining Agilent RRLC hardware with a method translation utility enables fast, high-resolution analysis of antioxidants and slip additives in polypropylene. The approach significantly improves laboratory productivity, reduces environmental impact, and maintains reliable quantitation in QA/QC workflows.
References
- ASTM D6042-04, Standard Test Method for Determination of Phenolic Antioxidants and Erucamide Slip Additives in Polypropylene Homopolymer Formulations Using Liquid Chromatography.
- Woodman, M., Improving the Effectiveness of Method Translation for Fast and High Resolution Separations.
- Woodman, M., Screening and Qualitative Identification of Antioxidant Polymer Additives by HPLC with UV/VIS and APCI-MS Detection.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Agilent 1260 Infinity Binary LC Optimization Guide
2010|Agilent Technologies|Guides
Agilent 1260 Infinity Binary LC Optimization Guide Authors: Angelika Gratzfeld-Huesgen, Michael Frank, Christian Gotenfels Table of Contents Introduction 5 1. Theory of using smaller particles in liquid chromatography 6 1.1. Benefits of columns packed with sub-2-micron particles 12 Frictional heating…
Key words
application, applicationcolumn, columnresolution, resolutionrrht, rrhtdelay, delayrapid, rapidsheet, sheetdata, datavolume, volumeagilent, agilentcolumns, columnsdatasheet, datasheetfast, fastzorbax, zorbaxthreaded
Polymers and Rubbers Application Compendium
2010|Agilent Technologies|Guides
MATERIALS TESTING AND RESEARCH SOLUTIONS FROM AGILENT Polymers and Rubbers Application Compendium AGILENT TECHNOLOGIES MATERIALS TESTING AND RESEARCH SOLUTIONS From the extraction of raw materials through the development, manufacturing and utilization of advanced materials, to material reuse and recycling, Agilent…
Key words
kfm, kfmsurface, surfaceforce, forceafm, afmmodulus, modulustopography, topographyimages, imagesimaging, imagingpolymer, polymerscratch, scratchsample, samplepotential, potentialprobe, probefrom, frommeasurements
ANALYSIS FOR EXTRACTABLE AND LEACHABLE COMPOUNDS FROM POLYMERIC MATERIALS
2016|Waters|Posters
ANALYSIS FOR EXTRACTABLE AND LEACHABLE COMPOUNDS FROM POLYMERIC MATERIALS Gordon Fujimoto1, Sarah Dowd1, Baiba Cabovska2, Marian Twohig2 1 Waters Corporation, Beverly, MA, USA; 2Waters Corporation, Milford, MA, USA INTRODUCTION Extractable and leachable compounds, which are MRM TRANSITIONS MS Conditions MS…
Key words
antioxidant, antioxidantstabilizer, stabilizerunispray, unisprayplasticizer, plasticizerxevo, xevoleachable, leachableextractable, extractablepolymeric, polymerictinuvin, tinuvinpreservative, preservativecompounds, compoundsbht, bhttandem, tandemionization, ionizationmaterials
High throughput analysis of polymer stabilizers with the Agilent 6420 Triple Quadrupole LC/MS
2014|Agilent Technologies|Others
materials analysis High throughput analysis of polymer stabilizers with the Agilent 6420 Triple Quadrupole LC/MS Solutions for Your Analytical Business Markets and Applications Programs Authors Susanne Beißmann, Michael Reisinger, Christian W. Klampfl, Wolfgang Buchberger Johannes Kepler University Linz, Austria. Andreas…
Key words
polymer, polymeradditives, additivespolyolefin, polyolefinmrm, mrmappi, appianalysis, analysismarkets, marketsbuchberger, buchbergerapci, apciprograms, programsfragmentor, fragmentorvoltage, voltagebusiness, businesskepler, keplerwere