LCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Agilent G6475A and 6495D LC/TQ Site Preparation Checklist

Manuals | 2023 | Agilent TechnologiesInstrumentation
LC/MS, LC/MS/MS, LC/QQQ
Industries
Manufacturer
Agilent Technologies

Summary

Significance of Topic


Ensuring that triple quadrupole LC/MS systems are installed under optimal conditions is critical to achieving reliable performance, longevity, and return on investment.

Objectives and Study Overview


This guide aims to outline the essential laboratory preparations for installing Agilent G6475A and G6495D LC/TQ instruments. It provides a structured checklist covering space allocation, environmental controls, utilities, consumables, and user responsibilities to facilitate a seamless start-up service.

Methodology and Instrumentation


The site readiness criteria were organized into functional areas: space and weight support, environmental parameters, exhaust and venting, electrical requirements, and gas supply.
  • Space and Weight: Bench footprint, clearance, and load capacity.
  • Environmental: Temperature, humidity, altitude, and vibration.
  • Exhaust: Separate vents, tubing, and flow rates.
  • Power: Voltage configuration, grounding, and outlet count.
  • Gas Supply: Purity, pressure, fittings, and flow for nitrogen supplies.

Used Instrumentation


  • Agilent G6475A LC/TQ system
  • Agilent G6495D LC/TQ system
  • MS40+ and MS120 foreline pumps
  • Agilent Jet Stream, APCI, APPI, MMI, ESI, NSI ion sources

Key Findings and Discussion


Bench Space and Weight:
  • G6475A: 84×77.3×47.5 cm, 117 kg; G6495D: 91.5×77.3×47.5 cm, 123 kg.
  • Minimum 30 cm clearance on sides and 1 m around for service access.

Environmental Conditions:
  • 15–35 °C, <85% RH at 35 °C, non‐condensing, indoor use, altitude up to 3,300–3,700 m.

Exhaust Venting:
  • Two independent negative‐pressure vents; ≥30 L/min max for G6475A, ≥50 L/min max for G6495D; 0.5″ ID tubing.

Power Requirements:
  • Single‐phase 200–240 VAC or 208‐Y; isolated, noise‐free ground; up to 3,800 VA; two dedicated outlets.

Gas Supplies:
  • Main N₂ ≥99.5% purity, 5.5–6.8 bar, 30–50 L/min.
  • Collision N₂ ≥99.999% purity, 1–2 bar, ≥0.001 L/min.

Benefits and Practical Applications


Implementing these site preparation steps ensures timely installation, minimizes downtime and service visits, and maximizes instrument uptime and data quality in laboratories handling high‐throughput quantitative analyses.

Future Trends and Opportunities


Emerging practices include integration of smart sensors for real‐time environment monitoring, remote commissioning via digital platforms, predictive maintenance, and automated site readiness validation tools to further accelerate instrument deployment.

Conclusion


A comprehensive site preparation protocol for Agilent G6475A and G6495D LC/TQ systems is essential for optimal performance and long‐term reliability. Adherence to detailed specifications in space, environmental control, utilities, and consumables enables efficient installation and operation.

Reference


Agilent Technologies. G6475A and G6495D LC/TQ Site Preparation Checklist. 2023.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Agilent 6400 Series LC/TQ - Site Preparation Checklist
Agilent 6400 Series LC/TQ Site Preparation Checklist Thank you for purchasing an Agilent 6400 LC/TQ. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is…
Key words
checklist, checklistsite, sitenitrogen, nitrogenspecial, specialhydrocarbon, hydrocarbonnotes, notessource, sourcefree, freeexhaust, exhaustsupplies, suppliescustomer, customerpreparation, preparationforeline, forelineinstrument, instrumentmaximum
Agilent 6400 Series LC/TQ - Site Preparation Checklist
Agilent 6400 Series LC/TQ Site Preparation Checklist Thank you for purchasing an Agilent 6400 LC/TQ. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is…
Key words
checklist, checklistsite, sitenitrogen, nitrogenspecial, specialhydrocarbon, hydrocarbonnotes, notessource, sourcefree, freeexhaust, exhaustsupplies, suppliescustomer, customerpreparation, preparationforeline, forelineinstrument, instrumentmaximum
Agilent Revident LC/QTOF Site Preparation Checklist
Agilent CrossLab Start Up Services Agilent Revident LC/QTOF Site Preparation Checklist Thank you for purchasing an instrument from Agilent Technologies. CrossLab Start Up is focused on helping customers shorten the time it takes to start realizing the full value of…
Key words
revident, revidentsite, sitechecklist, checklisttof, tofagilent, agilentcontino, continopreparation, preparationseries, seriesnathan, nathanservice, servicespecial, specialforeline, forelineqtof, qtofrequirements, requirementsrevision
Agilent InfinityLab Pro iQ Series Site Preparation Checklist
Agilent CrossLab Start Up Services Agilent InfinityLab Pro iQ Series Site Preparation Checklist Thank you for purchasing an instrument from Agilent Technologies. CrossLab Start Up is focused on helping customers shorten the time it takes to start realizing the full…
Key words
checklist, checklistsite, siteinfinitylab, infinitylabpro, proagilent, agilentseries, seriespreparation, preparationajs, ajsspecial, specialsource, sourceground, groundline, linecustomer, customerservice, serviceinstrument
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike