CRM Tellurium (Te) - Aqueous calibration solution ASTASOL® AN90591C
Others | 2023 | ANALYTIKAInstrumentation
The availability of well-characterized aqueous calibration solutions is essential for ensuring the accuracy and comparability of trace element determinations in environmental, clinical and industrial laboratories. A certified reference material (CRM) for tellurium supports method validation, instrument calibration and quality control across a range of atomic and molecular spectrometric techniques.
This certificate describes the preparation, certification and intended use of ASTASOL® AN90591C, an aqueous calibration solution containing tellurium at nominal concentration of 1000 mg/L. The document outlines the matrix composition, production details, metrological traceability and validation procedures securing its role as a primary calibration standard.
The solution was prepared gravimetrically from high-purity elemental tellurium (99.999%) dissolved in 20% hydrochloric acid (v/v), using sub-boil distilled acid and ultrapure water (≥18 MΩ·cm, 0.22 μm filtered). Density at 20 °C was determined to convert concentration between mass fraction and volume basis. Characterization employed:
The certified value for tellurium is 1000 ±2 mg/L (expanded uncertainty, k=2) at 20 °C, corresponding to 960 ±3 mg/kg by mass fraction. Trace impurities in the solution matrix are all below detection thresholds (<0.1 mg/L for most elements), ensuring minimal spectral interferences. Homogeneity tests confirmed uniform distribution of analyte across bottles, and stability studies guarantee integrity over the five-year shelf life under recommended storage conditions.
The CRM is intended for calibration and validation of analytical methods including AAS, AES, ICP-OES, ICP-MS, molecular absorption spectrometry and selected electroanalytical techniques. Its high purity, SI-traceability and detailed uncertainty budget make it suitable for routine QA/QC, method development and interlaboratory comparisons.
Advances in spectrometric instrumentation demand CRMs with lower uncertainty and broader elemental coverage. Future directions include multi-element solutions, ultra-low concentration standards for emerging contaminants and CRMs tailored for hyphenated techniques. Integration with digital certificate formats and blockchain-based traceability may further enhance reliability and data integrity.
ASTASOL® AN90591C provides a robust, well-characterized calibration standard for tellurium determination in acidic aqueous media. Its certified value, validated uncertainty and guaranteed homogeneity and stability support high-confidence analytical results in research and industrial laboratories.
Standards and chemicals
IndustriesManufacturerANALYTIKA
Summary
Significance of the Topic
The availability of well-characterized aqueous calibration solutions is essential for ensuring the accuracy and comparability of trace element determinations in environmental, clinical and industrial laboratories. A certified reference material (CRM) for tellurium supports method validation, instrument calibration and quality control across a range of atomic and molecular spectrometric techniques.
Objectives and Study Overview
This certificate describes the preparation, certification and intended use of ASTASOL® AN90591C, an aqueous calibration solution containing tellurium at nominal concentration of 1000 mg/L. The document outlines the matrix composition, production details, metrological traceability and validation procedures securing its role as a primary calibration standard.
Methodology and Instrumentation
The solution was prepared gravimetrically from high-purity elemental tellurium (99.999%) dissolved in 20% hydrochloric acid (v/v), using sub-boil distilled acid and ultrapure water (≥18 MΩ·cm, 0.22 μm filtered). Density at 20 °C was determined to convert concentration between mass fraction and volume basis. Characterization employed:
- ICP-OES for primary concentration measurement
- Comparison against SRM NIST 3156 for independent verification
- AAS, ICP-OES and ICP-MS to quantify trace elemental impurities (all below detection limits for key elements).
Main Results and Discussion
The certified value for tellurium is 1000 ±2 mg/L (expanded uncertainty, k=2) at 20 °C, corresponding to 960 ±3 mg/kg by mass fraction. Trace impurities in the solution matrix are all below detection thresholds (<0.1 mg/L for most elements), ensuring minimal spectral interferences. Homogeneity tests confirmed uniform distribution of analyte across bottles, and stability studies guarantee integrity over the five-year shelf life under recommended storage conditions.
Benefits and Practical Applications
The CRM is intended for calibration and validation of analytical methods including AAS, AES, ICP-OES, ICP-MS, molecular absorption spectrometry and selected electroanalytical techniques. Its high purity, SI-traceability and detailed uncertainty budget make it suitable for routine QA/QC, method development and interlaboratory comparisons.
Future Trends and Applications
Advances in spectrometric instrumentation demand CRMs with lower uncertainty and broader elemental coverage. Future directions include multi-element solutions, ultra-low concentration standards for emerging contaminants and CRMs tailored for hyphenated techniques. Integration with digital certificate formats and blockchain-based traceability may further enhance reliability and data integrity.
Conclusion
ASTASOL® AN90591C provides a robust, well-characterized calibration standard for tellurium determination in acidic aqueous media. Its certified value, validated uncertainty and guaranteed homogeneity and stability support high-confidence analytical results in research and industrial laboratories.
References
- ISO Guide 31: Reference materials – Contents of certificates and labels
- ČSN EN ISO 17034:2017 – General requirements for the competence of reference material producers
- NIST SRM 3156 – Tellurium Standard Reference Material
- ČSN EN ISO/IEC 17025:2018 – General requirements for the competence of testing and calibration laboratories
- ČSN EN ISO 9001:2016 – Quality management systems
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
CRM Tellurium (Te) - Aqueous calibration solution ASTASOL® AN905910C
2023|ANALYTIKA|Others
ANALYTIKA®, spol. s r.o. Khodlova 1297/47, CZ-193 00 Praha 9 – Horní Počernice www.analytika.net Analytika, spol. s r.o., Oddělení RM, výrobce RM č. 7501, akreditovaný ČIA podle ČSN EN ISO 17034:2017 Analytika, spol. s r.o., Department of RM, Reference Materials…
Key words
tellurium, telluriumcrm, crmaqueous, aqueousreference, referencecalibration, calibrationbellow, bellowastasol, astasoldeclared, declaredexpiration, expirationanalysing, analysingsolutions, solutionshomogeneous, homogeneousstoring, storingshelf, shelfguarantees
CRM Tellurium (Te) - Aqueous calibration solution ASTASOL® AN8059C
2022|ANALYTIKA|Others
ANALYTIKA®, spol. s r.o. Khodlova 1297/47, CZ-193 00 Praha 9 – Horní Počernice www.analytika.net Analytika, spol. s r.o., Oddělení RM, výrobce RM č. 7501, akreditovaný ČIA podle ČSN EN ISO 17034:2017 Analytika, spol. s r.o., Department of RM, Reference Materials…
Key words
crm, crmaqueous, aqueouselectroanalytical, electroanalyticaltellurium, telluriumcalibration, calibrationbellow, bellowastasol, astasoldeclared, declaredexpiration, expirationanalysing, analysingreference, referencehomogeneous, homogeneousstoring, storingshelf, shelfsolutions
CRM Copper (Cu) - Aqueous calibration solution ASTASOL® AN8015N
2023|ANALYTIKA|Others
ANALYTIKA®, spol. s r.o. Khodlova 1297/47, CZ-193 00 Praha 9 – Horní Počernice www.analytika.net Analytika, spol. s r.o., Oddělení RM, výrobce RM č. 7501, akreditovaný ČIA podle ČSN EN ISO 17034:2017 Analytika, spol. s r.o., Department of RM, Reference Materials…
Key words
crm, crmaqueous, aqueouscalibration, calibrationbellow, bellowastasol, astasoldeclared, declaredexpiration, expirationreference, referenceanalysing, analysingstoring, storinghomogeneous, homogeneoussolutions, solutionsshelf, shelfguarantees, guaranteescopper
CRM Copper (Cu) - Aqueous calibration solution ASTASOL® AN8015C
2022|ANALYTIKA|Others
ANALYTIKA®, spol. s r.o. Khodlova 1297/47, CZ-193 00 Praha 9 – Horní Počernice www.analytika.net Analytika, spol. s r.o., Oddělení RM, výrobce RM č. 7501, akreditovaný ČIA podle ČSN EN ISO 17034:2017 Analytika, spol. s r.o., Department of RM, Reference Materials…
Key words
crm, crmaqueous, aqueouscalibration, calibrationbellow, bellowastasol, astasoldeclared, declaredexpiration, expirationreference, referenceanalysing, analysingstoring, storinghomogeneous, homogeneoussolutions, solutionsshelf, shelfguarantees, guaranteescopper