Analysis of Clopyralid
Applications | 2023 | ShimadzuInstrumentation
Clopyralid is a selective auxin-type herbicide commonly used in agriculture and horticulture. Its persistence in soil and potential carryover into composts and fertilizers raises concerns about unintended phytotoxic effects on non‐target plants. Sensitive and reliable analytical methods are essential for monitoring trace levels of clopyralid to comply with regulatory limits and ensure product safety.
The primary goal of this study was to establish a rapid, robust LC-MS/MS protocol for the determination of clopyralid in fertilizer and compost matrices. Key objectives included optimizing chromatographic separation, minimizing matrix effects, and achieving high sensitivity and reproducibility suitable for routine quality control.
A reversed-phase gradient was employed using 0.1 % formic acid in water (mobile phase A) and methanol (mobile phase B). The gradient program increased from 5 % to 60 % B over 5 minutes, held at 95 % B, then returned to initial conditions. A flow-switching valve directed the eluent containing clopyralid (1.4–5.5 min) into the mass spectrometer to reduce background. The flow rate was set to 0.4 mL/min, column temperature to 40 °C, and injection volume to 5 µL.
The method provided sharp, symmetric peaks for clopyralid with retention times reproducibly between 1.4 and 5.5 minutes. The selective MRM transitions enabled clear discrimination from co‐eluting interferences. Flow switching effectively reduced matrix noise outside the target window. Although quantitative figures (LOD/LOQ) were not specified in the source, the protocol demonstrated sufficient sensitivity for trace‐level detection in complex fertilizer matrices.
This LC-MS/MS approach offers rapid analysis, straightforward sample introduction, and minimal carryover, making it well-suited for routine QA/QC labs. The selective MRM detection ensures low false positives, while the short run time increases sample throughput. Laboratories testing fertilizers, composts, or environmental samples can adopt this method for compliance monitoring.
Advances in high-resolution mass spectrometry (HRMS) may further enhance selectivity and enable multi-residue screening of herbicides in a single run. Automated sample preparation and online SPE coupling could increase throughput and reduce manual handling. Green chemistry initiatives may drive the adoption of alternative, eco-friendly mobile phases without compromising performance.
The reported LC-MS/MS method using a Shim-pack Scepter C18 column and Shimadzu LCMS-8500 achieves fast, sensitive, and reproducible determination of clopyralid in fertilizer matrices. Its robustness and selectivity support regulatory compliance and effective quality control in agricultural and environmental testing laboratories.
Consumables, LC columns, LC/MS, LC/MS/MS, LC/QQQ
IndustriesEnvironmental
ManufacturerShimadzu
Summary
Significance of the Topic
Clopyralid is a selective auxin-type herbicide commonly used in agriculture and horticulture. Its persistence in soil and potential carryover into composts and fertilizers raises concerns about unintended phytotoxic effects on non‐target plants. Sensitive and reliable analytical methods are essential for monitoring trace levels of clopyralid to comply with regulatory limits and ensure product safety.
Objectives and Overview of the Study
The primary goal of this study was to establish a rapid, robust LC-MS/MS protocol for the determination of clopyralid in fertilizer and compost matrices. Key objectives included optimizing chromatographic separation, minimizing matrix effects, and achieving high sensitivity and reproducibility suitable for routine quality control.
Methodology
A reversed-phase gradient was employed using 0.1 % formic acid in water (mobile phase A) and methanol (mobile phase B). The gradient program increased from 5 % to 60 % B over 5 minutes, held at 95 % B, then returned to initial conditions. A flow-switching valve directed the eluent containing clopyralid (1.4–5.5 min) into the mass spectrometer to reduce background. The flow rate was set to 0.4 mL/min, column temperature to 40 °C, and injection volume to 5 µL.
Used Instrumentation
- UHPLC System: Nexera™ X3
- Analytical Column: Shim-pack Scepter™ C18-120, 100 mm × 2.1 mm I.D., 1.9 µm
- Mass Spectrometer: LCMS-8500 triple quadrupole with ESI in positive mode
- Ionization Conditions: ESI+ at 0.5 kV, nebulizing gas 3 L/min, drying gas 10 L/min, interface temperature 400 °C
- MRM Transitions: m/z 191.90 → 146.10 (CE 21 V) and 191.90 → 110.10 (CE 34 V)
Main Results and Discussion
The method provided sharp, symmetric peaks for clopyralid with retention times reproducibly between 1.4 and 5.5 minutes. The selective MRM transitions enabled clear discrimination from co‐eluting interferences. Flow switching effectively reduced matrix noise outside the target window. Although quantitative figures (LOD/LOQ) were not specified in the source, the protocol demonstrated sufficient sensitivity for trace‐level detection in complex fertilizer matrices.
Benefits and Practical Applications
This LC-MS/MS approach offers rapid analysis, straightforward sample introduction, and minimal carryover, making it well-suited for routine QA/QC labs. The selective MRM detection ensures low false positives, while the short run time increases sample throughput. Laboratories testing fertilizers, composts, or environmental samples can adopt this method for compliance monitoring.
Future Trends and Opportunities
Advances in high-resolution mass spectrometry (HRMS) may further enhance selectivity and enable multi-residue screening of herbicides in a single run. Automated sample preparation and online SPE coupling could increase throughput and reduce manual handling. Green chemistry initiatives may drive the adoption of alternative, eco-friendly mobile phases without compromising performance.
Conclusion
The reported LC-MS/MS method using a Shim-pack Scepter C18 column and Shimadzu LCMS-8500 achieves fast, sensitive, and reproducible determination of clopyralid in fertilizer matrices. Its robustness and selectivity support regulatory compliance and effective quality control in agricultural and environmental testing laboratories.
Reference
- Shimadzu Application News an_01-00352. Shimadzu Corporation, First Edition August 2023.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Analysis of Clopyralid in Manure Compost and Sludge Fertilizer
2024|Shimadzu|Applications
ERAS-1000-0573 LC-MS Reversed-phase Shim-packTM Series Shim-pack ScepterTM C18-120 Analysis of Clopyralid in Manure Compost and Sludge Fertilizer 573 Keywords: fertilizers, herbicide Cattle manure compost Horse manure compost [HPLC conditions] (NexeraTM X3) Column Mobile phases Time program Flow rate Injection vol.…
Key words
manure, manurecompost, compostfertilizer, fertilizersludge, sludgecomposted, compostedclopyralid, clopyralidpig, pigcattle, cattlehorse, horsepacktm, packtmflow, flowshim, shimtransition, transitiongas, gasprobe
Analysis of Diarrhetic Shellfish Toxines
2023|Shimadzu|Applications
ERAS-1000-0531 Reversed-phase LC Shim-packTM Series Shim-pack ScepterTM Phenyl-120 Analysis of Diarrhetic Shellfish Toxines 531 Keywords: Food poisoning, Scallop (x1,000) 3 4.5 1: OA 803.30>255.10 (-) 2: DTX2 803.30>255.10 (-) 3: DTX1 817.30>255.10(-) 4.0 3.5 1 2 3.0 2.5 2.0 1.5…
Key words
toxines, toxinesdiarrhetic, diarrheticshellfish, shellfishphase, phasetemperature, temperaturepacktm, packtmmobile, mobileshim, shimcid, cidmin, mingas, gasprobe, probeelectrospray, electrosprayheating, heatingdrying
Analysis of Veterinary Drugs
2023|Shimadzu|Applications
ERAS-1000-0440 LC-MS Reversed-phase Shim-packTM Series Shim-pack ScepterTM C18-120 [Metal free column] Analysis of Veterinary Drugs 440 Keywords: non-target analysis, antibacterial drugs, antimicrobial drugs Sarafloxacin Sulfabenzamide Sulfabromomethazine Na Sulfacetamide Sulfachlorpyridazine Sulfadiazine Sulfadimethoxine Sulfadimidine Sulfadoxine Sulfaethoxypyridazine Sulfamerazine Sulfamethoxazole Sulfamethoxypyridazine Sulfametoxydiazine Sulfamonomethoxine Sulfapyridine…
Key words
lumequine, lumequineoxolinicacid, oxolinicacidpiromidicacid, piromidicacidmiloxacin, miloxacinsulfametoxydiazine, sulfametoxydiazinesulfisozole, sulfisozolesulfatroxazole, sulfatroxazolesulfabromomethazine, sulfabromomethazinesulfachlorpyridazine, sulfachlorpyridazinesulfaethoxypyridazine, sulfaethoxypyridazinediaveridine, diaveridinesulfisomidine, sulfisomidineorbifloxacin, orbifloxacinormetoprim, ormetoprimsulfabenzamide
Analysis of Small Molecular Pharmaceuticals
2024|Shimadzu|Applications
ERAS-1000-0588 LC-MS Reversed-phase Shim-packTM Series Shim-pack ScepterTM C18-120 Analysis of Small Molecular Pharmaceuticals 588 Keywords: pharmaceuticals 1.Quinidine 2. Impurity of Quinidine 3.Lidocaine 4. Metoclopramide 5.Papaverine 6.Dibucaine 7.Amitriptyline [HPLC conditions] (NexeraTM X3) (Method Scouting System) Column Mobile phase A Mobile phase…
Key words
quinidine, quinidineqarray, qarrayphase, phasevoltage, voltagegas, gaspacktm, packtmflow, flowmobile, mobilenebulizing, nebulizingshim, shimmode, modepharmaceuticals, pharmaceuticalsheating, heatingdrying, dryingreversed