FlipLC Polar molecules in complex samples

Brochures and specifications |  | SIELCInstrumentation
HPLC
Industries
Other
Manufacturer
SIELC

Summary

Importance of the Topic


Analysis of small polar and charged compounds in complex matrices such as food, biological fluids, soil or water is often hampered by co-eluting interferences. Traditional sample cleanup methods and selective detectors (e.g., mass spectrometry) add time, cost and complexity. The FlipLC approach integrates online cleanup with chromatographic separation, delivering cleaner baselines, shorter runtimes and extended column lifetimes while preserving sensitivity and specificity.

Objectives and Study Overview


This whitepaper presents the design, operating principle and practical performance of a FlipLC system. Key goals include demonstrating how a dual-column configuration with timed valve switching removes matrix interferents, evaluating mixed-mode Primesep columns for orthogonal retention, and showcasing applications across diverse analytes in various sample types.

Methodology and Instrumentation


The FlipLC setup comprises an upstream isolation column and a downstream analytical column arranged in series, separated by a high-pressure switching valve. After sample injection onto the isolation column, valve actuation at a predefined time back-flushes late-eluting impurities to waste via the detector outlet, while target analytes proceed to the analytical column under normal forward flow. Primesep mixed-mode cation-exchange/reverse phase or anion-exchange/reverse phase columns exploit orthogonal retention based on charge and hydrophobicity. When paired with UV or MS detection, this configuration enables automated online cleanup without additional offline sample preparation.

Used Instrumentation


  • HPLC system with dual pump capability or standard pump plus secondary wash pump
  • High-pressure switching valve upstream of analytical column
  • Primesep isolation columns (e.g., Primesep 100, Primesep SB)
  • Primesep analytical columns (e.g., Primesep B, Primesep 200, Primesep SB)
  • UV detector (200–280 nm) or mass spectrometer for enhanced specificity
  • Mobile phases: gradients of acetonitrile/water with MS-compatible buffers (e.g., ammonium formate, H₂SO₄, H₃PO₄)

Main Results and Discussion


Application examples illustrate dramatic baseline improvement and sharper peaks across analytes including nitrates in chicken broth, ascorbic acid in fruit juices, metabisulfite in wine, neurotransmitters (DOPA, dopamine, epinephrine, norepinephrine, serotonin) in urine, tramadol and hydrocodone in urine, and serotonin in serum. Valve switch times between 0.5–1.5 min effectively exclude early or late matrix components. Limits of quantitation down to 10 ppb were achieved with UV detection; MS detection promises even lower detection limits and improved selectivity.

Benefits and Practical Applications


  • Automated, online sample cleanup reduces manual preparation steps and solvent use
  • Cleaner chromatograms shorten runtimes and improve peak resolution
  • Extended lifetime of expensive analytical columns by preventing irreversible contamination
  • Flexible method tuning via valve timing and column selection for diverse sample types
  • Compatible with existing HPLC-UV and HPLC-MS instrumentation for routine QA/QC and research workflows

Future Trends and Applications


Integration of FlipLC with high-resolution MS and tandem MS will expand its utility for trace-level quantitation of emerging contaminants and biomarkers. Further miniaturization and automated valve control can enable high-throughput platforms. Custom mixed-mode chemistries may be developed to target new classes of analytes, including highly polar ionic species and labile metabolites.

Conclusion


The FlipLC dual-column switching methodology offers a versatile, cost-effective solution for analyzing charged and polar compounds in complex matrices. By combining orthogonal retention mechanisms and timed valve actuation, it achieves high specificity, reduced interference and streamlined workflows using standard HPLC equipment.

References


No external literature references were provided in the source document.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
FlipLCTM. Measuring Charged Analytes in Complex Mixtures. Part IV. Metabisulfite
​ ​SIELC Technologies, Inc. Wheeling, IL, USA www.sielc.com​ Ph. ​847-229-2629 FlipLC​TM​. Measuring Charged Analytes in Complex Mixtures. Part IV. Metabisulfite. We recently reported a new method that was developed for the analysis of charged analytes in complex mixtures such as…
Key words
vine, vinerimesep, rimesepmetabisulfite, metabisulfitecharged, chargedflip, flipretains, retainsdrastically, drasticallysufficiently, sufficientlyisolated, isolatedcomplicated, complicatedrecently, recentlyconsists, consistsisolation, isolationsimplified, simplifiedconnected
FlipLCTM. A New Method for Measuring Charged Analytes in Complex Mixtures. Part I.
​ ​SIELC Technologies, Inc. Wheeling, IL, USA www.sielc.com​ Ph. ​847-229-262​9 FlipLC​TM​. A New Method for Measuring Charged Analytes in Complex Mixtures. Part I. Measuring by HPLC a particular component of complex samples such as natural products, body fluids, environmental probes…
Key words
grocery, grocerystore, storewatermelon, watermelongarden, gardenpellegrino, pellegrinobologna, bolognafrom, fromcharged, chargedmelon, melonsausages, sausagesproposing, proposingactuated, actuatedlettuce, lettucesan, sanapple
FlipLCTM. A New Method for Measuring Charged Analytes in Complex Mixtures. Part III.
​ ​SIELC Technologies, Inc. Wheeling, IL, USA www.sielc.com​ Ph. ​847-229-2629 FlipLC​TM​. A New Method for Measuring Charged Analytes in Complex Mixtures. Part III. We recently reported a new method that was developed for the analysis of charged analytes in complex…
Key words
plant, plantcornelian, cornelianrimesep, rimeseprosehips, rosehipsblackcurrant, blackcurranttangerine, tangerinehorseradish, horseradishcharged, chargedsource, sourcecabbage, cabbagecherry, cherryflip, flipgrapefruit, grapefruitstrawberry, strawberrylemon
The Do’s and Don’ts of Method Development in Mixed-Mode Chromatography Part 1. Reversed-Phase Ion-Exchange Columns
August 2009 Newsletter The Do’s and Don’ts of Method Development in Mixed-Mode Chromatography Part 1. Reversed-Phase Ion-Exchange Columns SIELC Technologies, Inc., Prospect Heights, IL 60070 If a reversed-phase chromatography column does not work for your separation, it is most likely…
Key words
primesep, primesepstrength, strengthcolumn, columndon’ts, don’tselsd, elsdequilibration, equilibrationtry, trymin, minbuffer, bufferparaben, parabenproblem, problemcompounds, compoundsphase, phasemixed, mixedmobile
Other projects
GCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike