Accurate Molecular Weight Determination of Polystyrene-Tetrahydrofuran Solutions using the Arc™ HPLC System with a Strong Solvent Compatibility Kit and Refractive Index (RI) Detector
Applications | 2024 | WatersInstrumentation
Accurate molecular weight determination of polymers is critical for predicting their physical properties, ensuring consistent quality in production and guiding formulation decisions. Gel permeation chromatography (GPC) coupled with refractive index detection provides a robust approach for characterizing molecular weight distributions, which influence mechanical strength, elasticity and processability.
This application note demonstrates the use of the Arc HPLC System equipped with a strong solvent compatibility kit and a refractive index (RI) detector to:
Sample Preparation:
Chromatographic System:
Columns in Series:
Data Processing:
Calibration:
Sample Analysis:
Inter-Day Performance:
Emerging directions in polymer GPC include multi-detector setups (e.g., light scattering, viscometry) for absolute molecular weight determination, integration of automated sample handling workflows, and application of machine learning algorithms to enhance calibration and data interpretation. Expansion to polar and water-soluble polymers using compatible solvent kits can broaden method applicability.
The Arc HPLC System with a strong solvent compatibility kit, combined with RI detection and Empower GPC software, delivers accurate, reproducible molecular weight measurements for polystyrene in THF. The approach supports rigorous polymer characterization and quality control in research and industrial laboratories.
GPC/SEC
IndustriesEnergy & Chemicals
ManufacturerWaters
Summary
Significance of the Topic
Accurate molecular weight determination of polymers is critical for predicting their physical properties, ensuring consistent quality in production and guiding formulation decisions. Gel permeation chromatography (GPC) coupled with refractive index detection provides a robust approach for characterizing molecular weight distributions, which influence mechanical strength, elasticity and processability.
Objectives and Study Overview
This application note demonstrates the use of the Arc HPLC System equipped with a strong solvent compatibility kit and a refractive index (RI) detector to:
- Generate reliable calibration curves using narrow polystyrene standards dissolved in tetrahydrofuran (THF).
- Determine molecular weight parameters (Mp, Mw, Mn) and polydispersity index (PDI) for an unknown polystyrene sample.
- Assess method repeatability and inter-day performance using Empower 3 Software with the GPC option.
Methodology and Instrumentation
Sample Preparation:
- Narrow polystyrene standards and an unknown sample prepared at 1.5 mg/mL in HPLC-grade THF.
- Dissolution for several hours prior to analysis.
Chromatographic System:
- Arc HPLC System with quaternary solvent manager, flow-through-needle and strong solvent compatibility kit.
- Isocratic elution with THF at 1.0 mL/min, column temperature 35 °C, sample temperature 15 °C.
- Injection volume: 50 µL.
- Detector: Refractive index, sampling rate 10 pts/sec, flow cell at 35 °C.
Columns in Series:
- Styragel HR 4 (10 000 Å; MW range 5 000–600 000 Da)
- Styragel HR 2 (500 Å; MW range 500–20 000 Da)
- Styragel HR 1 (100 Å; MW range 100–5 000 Da)
Data Processing:
- Empower 3 FR5 SR5 with GPC option to calculate retention time vs. log(Mp) calibration and sample molecular weight distribution.
Main Results and Discussion
Calibration:
- Third-order polynomial fit of narrow standards gave R² > 0.9997, indicating excellent linearity in log(Mp) vs. retention time.
Sample Analysis:
- Six replicate injections yielded RSDs of 0.20–0.67% for Mp, Mw and Mn.
- Accuracy compared to target values (34 500, 34 000, 33 200 Da) was 100.6% (Mp), 100.0% (Mw) and 98.9% (Mn).
- Empower GPC distribution plots provided clear visualization of molecular weight profiles.
Inter-Day Performance:
- Repeated analyses over multiple days confirmed stable calibration curves and consistent molecular weight results, underscoring method robustness.
Benefits and Practical Applications
- High-resolution separation across a broad molecular weight range using a combined column set.
- Strong solvent compatibility enables direct analysis of polymer solutions prepared in THF.
- Empower GPC software streamlines calibration, calculation and reporting of polymer molecular weight distributions.
Future Trends and Opportunities
Emerging directions in polymer GPC include multi-detector setups (e.g., light scattering, viscometry) for absolute molecular weight determination, integration of automated sample handling workflows, and application of machine learning algorithms to enhance calibration and data interpretation. Expansion to polar and water-soluble polymers using compatible solvent kits can broaden method applicability.
Conclusion
The Arc HPLC System with a strong solvent compatibility kit, combined with RI detection and Empower GPC software, delivers accurate, reproducible molecular weight measurements for polystyrene in THF. The approach supports rigorous polymer characterization and quality control in research and industrial laboratories.
References
- Knol WC, Pirok BW, Peters RAH. Detection Challenges in Quantitative Polymer Analysis by Liquid Chromatography. Journal of Separation Science, 2020.
- Walsh D, Schinski DA, Schneider RA, Guironnet D. General route to design polymer molecular weight distributions through flow chemistry. Nature Communications, 2020.
- Waters Corporation. Arc HPLC QSM-FTN Strong Solvent Compatibility Kit Product Solutions (715009279), 2023.
- Waters Corporation. Empower GPC Software Getting Started Guide (71500031303), 2023.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Analysis of Polylactic-co-Glycolic Acid (PLGA) by Gel Permeation Chromatography (GPC) using the Arc™ HPLC System with a Refractive Index (RI) Detector
2024|Waters|Applications
Application Note Analysis of Polylactic-co-Glycolic Acid (PLGA) by Gel Permeation Chromatography (GPC) using the Arc™ HPLC System with a Refractive Index (RI) Detector Margaret Maziarz Waters Corporation Abstract Polylactic-co-glycolic acid (PLGA) is a biodegradable polymer widely used in medical applications…
Key words
plga, plgapolylactic, polylacticglycolic, glycolicgpc, gpcarc, arcpermeation, permeationrefractive, refractivegel, gelindex, indexhplc, hplcacid, aciddetector, detectorchromatography, chromatographysystem, systemusing
Reproducibility and Reliability of GPC Analysis of Polystyrene on the Arc™ HPLC System
2024|Waters|Applications
Application Note Reproducibility and Reliability of GPC Analysis of Polystyrene on the Arc™ HPLC System Kaveh Amini, Paula Hong Waters Corporation Abstract In this application note, we demonstrate the ability to adapt the Arc HPLC System for GPC analyses and…
Key words
polystyrene, polystyrenegpc, gpcarc, arcreliability, reliabilityhplc, hplcreproducibility, reproducibilitysystem, systemanalysis, analysisprivacy, privacychc, chcstrong, strongreproducible, reproduciblesample, samplereliable, reliablestabilized
Performance Verification of a Proposed USP Monograph for Sorbitan Monooleate Using a Gel Permeation Chromatography (GPC) Method With Refractive Index (RI) Detection
2024|Waters|Applications
Application Note Performance Verification of a Proposed USP Monograph for Sorbitan Monooleate Using a Gel Permeation Chromatography (GPC) Method With Refractive Index (RI) Detection Margaret Maziarz Waters Corporation Abstract The United States Pharmacopeia (USP) proposed revisions to the monograph for…
Key words
monooleate, monooleatesorbitan, sorbitanusp, uspmonograph, monographproposed, proposedgpc, gpcrefractive, refractivepermeation, permeationgel, gelindex, indexverification, verificationperformance, performancechromatography, chromatographydetection, detectionmethod
Verifying Performance and Compliance of a Gel Permeation Chromatography (GPC) Method With the Requirements in a Proposed USP Monograph for Sorbitan Monostearate
2025|Waters|Applications
Application Note Verifying Performance and Compliance of a Gel Permeation Chromatography (GPC) Method With the Requirements in a Proposed USP Monograph for Sorbitan Monostearate Margaret Maziarz Waters Corporation Abstract The proposed United States Pharmacopeia (USP) monograph for sorbitan monostearate utilizes…
Key words
sorbitan, sorbitanmonostearate, monostearatemonograph, monographusp, uspproposed, proposedgpc, gpcverifying, verifyingpermeation, permeationgel, gelcompliance, compliancerequirements, requirementsperformance, performancemethod, methodchromatography, chromatographyimpurities