Reproducibility and Reliability of GPC Analysis of Polystyrene on the Arc™ HPLC System
Applications | 2024 | WatersInstrumentation
Gel Permeation Chromatography (GPC) is essential for determining polymer molecular weight distributions, a critical parameter in materials science and quality control. Reliable and reproducible retention times ensure accurate calculation of molecular weight averages (Mn, Mw, Mp) and polydispersity. A system capable of withstanding strong organic solvents like THF without performance degradation is vital for routine polymer analysis in research and industrial laboratories.
This study assessed the reproducibility, reliability and durability of the Waters Arc HPLC System equipped with a strong solvent compatibility kit for GPC analysis of polystyrene. Over five consecutive days, narrow polystyrene standards and a broad polystyrene sample were analyzed daily with replicate injections. Key goals included verifying retention time precision, calibration curve stability and consistency of calculated molecular weight values.
All analyses were performed under isocratic conditions using stabilized tetrahydrofuran as mobile phase at 1.0 mL/min and 40 °C column temperature. A four-column GPC bank (Styragel HR 3, HR 4, HR 4E and HR 5E) covered a wide MW range (approximately 50 Da to 4 M Da). Detection was by refractive index (10 pts/s, 40 °C). Samples and ReadyCal polystyrene standards (Mp ranges 370–2 520 000 Da) were dissolved in THF and injected at 50 µL. Data acquisition and processing used Empower™ 3.8.0 CDS with the GPC option.
Calibration curves generated each day showed a third-order fit between log molecular weight and retention time with R² consistently ≥ 0.9997. Intra-day retention time precision for key standards (Mp ≈ 277 000 Da and 133 000 Da) exhibited RSDs below 0.32%. Inter-day variability remained within 0.01 min. The broad polystyrene sample displayed Mn, Mw and Mp values within 10% of expected values and polydispersity within 5%, with retention time RSDs < 0.02% across five days. Overlay of molecular weight distribution plots confirmed day-to-day consistency.
Advances may include coupling GPC with multi-angle light scattering or mass spectrometry for absolute molecular weight determination. Miniaturized and high-throughput GPC formats, automated sample handling and AI-driven data interpretation promise faster, more comprehensive polymer characterization. Developing solvent-resistant materials and modules will further extend system lifetime and application scope.
The Arc HPLC System fitted with a strong solvent compatibility kit demonstrated excellent reproducibility and reliability in polystyrene GPC analysis over five days. It maintained stable retention times, consistent calibration performance and accurate molecular weight determinations, confirming its suitability for routine GPC workflows.
GPC/SEC
IndustriesEnergy & Chemicals
ManufacturerWaters
Summary
Significance of the Topic
Gel Permeation Chromatography (GPC) is essential for determining polymer molecular weight distributions, a critical parameter in materials science and quality control. Reliable and reproducible retention times ensure accurate calculation of molecular weight averages (Mn, Mw, Mp) and polydispersity. A system capable of withstanding strong organic solvents like THF without performance degradation is vital for routine polymer analysis in research and industrial laboratories.
Study Objectives and Overview
This study assessed the reproducibility, reliability and durability of the Waters Arc HPLC System equipped with a strong solvent compatibility kit for GPC analysis of polystyrene. Over five consecutive days, narrow polystyrene standards and a broad polystyrene sample were analyzed daily with replicate injections. Key goals included verifying retention time precision, calibration curve stability and consistency of calculated molecular weight values.
Methodology and Instrumentation
All analyses were performed under isocratic conditions using stabilized tetrahydrofuran as mobile phase at 1.0 mL/min and 40 °C column temperature. A four-column GPC bank (Styragel HR 3, HR 4, HR 4E and HR 5E) covered a wide MW range (approximately 50 Da to 4 M Da). Detection was by refractive index (10 pts/s, 40 °C). Samples and ReadyCal polystyrene standards (Mp ranges 370–2 520 000 Da) were dissolved in THF and injected at 50 µL. Data acquisition and processing used Empower™ 3.8.0 CDS with the GPC option.
Results and Discussion
Calibration curves generated each day showed a third-order fit between log molecular weight and retention time with R² consistently ≥ 0.9997. Intra-day retention time precision for key standards (Mp ≈ 277 000 Da and 133 000 Da) exhibited RSDs below 0.32%. Inter-day variability remained within 0.01 min. The broad polystyrene sample displayed Mn, Mw and Mp values within 10% of expected values and polydispersity within 5%, with retention time RSDs < 0.02% across five days. Overlay of molecular weight distribution plots confirmed day-to-day consistency.
Benefits and Practical Applications
- The Arc HPLC System with strong solvent compatibility kit delivers robust performance in THF for extended GPC use.
- High flow-rate precision ensures stable retention times and reproducible calibration curves.
- Integration with Empower CDS GPC option streamlines molecular weight calculations and reporting.
- Suitable for routine polymer quality control, materials research and industrial analytics.
Future Trends and Opportunities
Advances may include coupling GPC with multi-angle light scattering or mass spectrometry for absolute molecular weight determination. Miniaturized and high-throughput GPC formats, automated sample handling and AI-driven data interpretation promise faster, more comprehensive polymer characterization. Developing solvent-resistant materials and modules will further extend system lifetime and application scope.
Conclusion
The Arc HPLC System fitted with a strong solvent compatibility kit demonstrated excellent reproducibility and reliability in polystyrene GPC analysis over five days. It maintained stable retention times, consistent calibration performance and accurate molecular weight determinations, confirming its suitability for routine GPC workflows.
References
- Empower GPC Software Getting Started Guide, Waters Corporation User Guide, 71500031303
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Accurate Molecular Weight Determination of Polystyrene-Tetrahydrofuran Solutions using the Arc™ HPLC System with a Strong Solvent Compatibility Kit and Refractive Index (RI) Detector
2024|Waters|Applications
Application Note Accurate Molecular Weight Determination of Polystyrene-Tetrahydrofuran Solutions using the Arc™ HPLC System with a Strong Solvent Compatibility Kit and Refractive Index (RI) Detector Margaret Maziarz Waters Corporation Abstract Achieving accurate and reproducible molecular weight measurements is paramount to…
Key words
polystyrene, polystyrenearc, arctetrahydrofuran, tetrahydrofuranrefractive, refractivecompatibility, compatibilitymolecular, molecularstrong, strongweight, weightindex, indexkit, kitaccurate, accuratehplc, hplcsolutions, solutionsdetermination, determinationsolvent
Analysis of Polylactic-co-Glycolic Acid (PLGA) by Gel Permeation Chromatography (GPC) using the Arc™ HPLC System with a Refractive Index (RI) Detector
2024|Waters|Applications
Application Note Analysis of Polylactic-co-Glycolic Acid (PLGA) by Gel Permeation Chromatography (GPC) using the Arc™ HPLC System with a Refractive Index (RI) Detector Margaret Maziarz Waters Corporation Abstract Polylactic-co-glycolic acid (PLGA) is a biodegradable polymer widely used in medical applications…
Key words
plga, plgapolylactic, polylacticglycolic, glycolicgpc, gpcarc, arcpermeation, permeationrefractive, refractivegel, gelindex, indexhplc, hplcacid, aciddetector, detectorchromatography, chromatographysystem, systemusing
Considerations for GPC Analysis Using Lithium Bromide Modified Dimethylformamide as Mobile Phase
2025|Waters|Posters
CONSIDERATIONS FOR GPC ANALYSIS USING LITHIUM BROMIDE MODIFIED DIMETHYLFORMAMIDE AS MOBILE PHASE Authors: Kaveh Amini, Elom Pedanou, Paula Hong Affiliations: Waters Corporation, Milford, MA, USA INTRODUCTION While GPC is commonly performed with DMF as a mobile phase, this mobile phase…
Key words
cont’d, cont’dgpc, gpcrsd, rsdarea, areastyragel, styrageldimethylformamide, dimethylformamidepdi, pdipeak, peakobtained, obtainedbromide, bromideempower, empowerretention, retentionlithium, lithiumviscosity, viscosityconsiderations
Performance Verification of a Proposed USP Monograph for Sorbitan Monooleate Using a Gel Permeation Chromatography (GPC) Method With Refractive Index (RI) Detection
2024|Waters|Applications
Application Note Performance Verification of a Proposed USP Monograph for Sorbitan Monooleate Using a Gel Permeation Chromatography (GPC) Method With Refractive Index (RI) Detection Margaret Maziarz Waters Corporation Abstract The United States Pharmacopeia (USP) proposed revisions to the monograph for…
Key words
monooleate, monooleatesorbitan, sorbitanusp, uspmonograph, monographproposed, proposedgpc, gpcrefractive, refractivepermeation, permeationgel, gelindex, indexverification, verificationperformance, performancechromatography, chromatographydetection, detectionmethod