Summary

Significance of the Topic


Automated fraction collection is a vital component of preparative liquid chromatography and bioseparation workflows. It enables precise separation and recovery of target analytes, improves reproducibility, and reduces manual intervention. Integrating a fraction collector with chromatography data systems ensures synchronized data acquisition and instrument control, streamlining method development and routine operation.

Objectives and Study Overview


This document presents the integration of the Teledyne Foxy Jr fraction collector into the Clarity chromatography station by DataApex. The goals are to describe hardware requirements, software configuration, fraction collection programming, real-time monitoring, and troubleshooting guidance. The result is a seamless control module that stores method parameters and fraction events within each chromatogram.

Methodology and Instrumentation


The integration uses RS-232 serial communication between the PC and the Foxy Jr unit. Key steps include:
  • Installing appropriate Clarity control licenses (LC Control for pumps and fraction collectors).
  • Connecting the fraction collector via DB9F-DB25F or DB9F-RJ11 serial cables.
  • Adding the device in Clarity’s System Configuration and assigning it to the LC instrument slot.
  • Auto-detecting communication settings in the DataApex UNI Setup (port, baud rate, rack type).

Used Instrumentation


The hardware and software components are:
  • Teledyne Foxy Jr fraction collector
  • PC with Clarity chromatography station software and LC Control module
  • Serial communication cables (SK04 or SK05)
  • Optionally, a MultiCOM adapter for multiple RS-232 devices

Main Results and Discussion


Once configured, the Foxy Jr collector operates under a user-defined fraction table. Key features include:
  • Definition of flow rate and tubing delay volumes to synchronize detection and collection.
  • Fraction table rows specifying start/stop times, detector signal conditions (level, slope, combined logic), vial volume limits, and data filtering.
  • Vial numbering control, reset options, and end-of-vial handling (waste or shutdown).
  • Real-time control via Device Monitor: manual vial advance, start/stop collecting, go-to-position, and resume method.
  • Visual display of fraction collection events in the Data Acquisition window, with color-coded delays and actual fraction intervals.
  • Troubleshooting via COM port logging in COMMDRV.INI for detailed message capture.

Benefits and Practical Applications


Integrating the Teledyne Foxy Jr with Clarity provides:
  • Automated, rule-based fraction collection synchronized with chromatographic events.
  • Embedded method parameters and fraction metadata within chromatogram files.
  • Reduced risk of overflow or sample loss through delay compensation and automated vial switching.
  • Improved method reproducibility and streamlined workflow for research, QA/QC, and preparative separations.

Future Trends and Potential Applications


Advances may include remote or cloud-based instrument management, enhanced scripting for complex fractionation schemes, integration with multi-omics workflows, and AI-driven dynamic fraction criteria. Expanded support for high-throughput formats and automated post-collection processing will further increase productivity.

Conclusion


The Teledyne Foxy Jr control module in Clarity offers comprehensive, automated fraction collection capabilities. By combining detailed method programming, real-time monitoring, and integrated data logging, this solution enhances operational efficiency and method reliability in preparative chromatography.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Clarity Control Module TELEDYNE FOXY ISCO 200
TELEDYNE FOXY ISCO 200 Clarity Control Module ENG Code/Rev.: M165/90B Date: 2024-11-13 DataApex Ltd. Phone: +420 251 013 400 Petrzilkova 2583/13 [email protected] 158 00 Prague 5 www.dataapex.com Czech Republic ® are trademarks of DataApex Ltd. Microsoft® and WindowsTM are Clarity®,…
Key words
fraction, fractionsignal, signalfulfilled, fulfilledvial, vialstart, startstop, stopcollector, collectordetector, detectordecreases, decreasesderivative, derivativefirst, firstcollection, collectionsetup, setuptab, tabstops
Clarity Control Module TELEDYNE FOXY R1 - R2
TELEDYNE FOXY R1 - R2 Clarity Control Module ENG Code/Rev.: M164/90B Date: 2024-11-13 DataApex Ltd. Phone: +420 251 013 400 Petrzilkova 2583/13 [email protected] 158 00 Prague 5 www.dataapex.com Czech Republic ® are trademarks of DataApex Ltd. Microsoft® and WindowsTM are…
Key words
fraction, fractionsignal, signalfulfilled, fulfilledvial, vialstart, startstop, stopdetector, detectordecreases, decreasesderivative, derivativefirst, firstsetup, setupcollection, collectiondialog, dialogstops, stopsclarity
Clarity Control Module FC BY UPCHURCH VALVE
FC BY UPCHURCH VALVE Clarity Control Module ENG Code/Rev.: M237/90D Date: 2024-11-13 DataApex Ltd. Phone: +420 251 013 400 Petrzilkova 2583/13 [email protected] 158 00 Prague 5 www.dataapex.com Czech Republic ® are trademarks of DataApex Ltd. Microsoft® and WindowsTM are Clarity®,…
Key words
upchurch, upchurchfraction, fractionvalve, valvesignal, signalfulfilled, fulfilledvial, vialstart, startstop, stopdetector, detectorcollector, collectorsetup, setupdecreases, decreasesderivative, derivativefirst, firstcollection
Clarity Control Module FC BY RHEODYNE VALVE
FC BY RHEODYNE VALVE Clarity Control Module ENG Code/Rev.: M238/90D Date: 2024-11-13 DataApex Ltd. Phone: +420 251 013 400 Petrzilkova 2583/13 [email protected] 158 00 Prague 5 www.dataapex.com Czech Republic ® are trademarks of DataApex Ltd. Microsoft® and WindowsTM are Clarity®,…