Impurities test for Cefradine (EP- method)

Applications |  | Thermo Fisher ScientificInstrumentation
HPLC
Industries
Pharma & Biopharma
Manufacturer
Thermo Fisher Scientific

Summary

Significance of the Topic


The determination of related substances in Cefradine is essential to ensure drug safety, efficacy, and compliance with pharmacopeial specifications. Accurate impurity profiling supports quality control in pharmaceutical production and helps protect patient health.

Objectives and Study Overview


This study presents a European Pharmacopoeia–compliant HPLC method for the identification and quantification of impurities in Cefradine. The workflow covers the preparation of multiple reference and test solutions, chromatographic separation under gradient conditions, and validation of system suitability criteria.

Used Instrumentation


  • High-performance liquid chromatograph: UltiMate 3000 LC
  • Column: Hypersil Gold, 4.6 × 150 mm, 5 μm particles
  • Detector: UV set at 220 nm

Methodology


Sample preparation involved dissolving known amounts of Cefradine, Cefalexin, and impurity standards in the phosphate buffer mobile phase, followed by appropriate dilutions. Mobile phase A consisted of a 2.72 g/L potassium dihydrogen phosphate solution adjusted to pH 3, and mobile phase B was methanol. A gradient elution program was applied over 25 minutes, with a flow rate of 1.0 mL/min, column temperature maintained at 30 °C, and an injection volume of 25 μL.

Main Results and Discussion


System suitability tests demonstrated robust performance:
  • Resolution between Cefalexin and Cefradine: 7.32 (minimum requirement 4.0)
  • Retention time of Cefradine: 15.06 minutes (target about 15 minutes)
  • Relative retention times (RRT) for seven impurities fell within acceptable EP ranges, showing clear separation and reproducibility.

Chromatograms confirmed distinct peak shapes, baseline separation of all impurities, and stable retention times throughout the gradient run.

Benefits and Practical Applications


This validated method offers:
  • High resolution of Cefradine and its related substances
  • Reproducible retention times for routine quality control
  • Compatibility with existing HPLC systems in pharmaceutical laboratories

Its straightforward sample preparation and gradient program make it suitable for high-throughput impurity testing in manufacturing and QC environments.

Future Trends and Potential Applications


Advances may include coupling this method with mass spectrometry for improved impurity identification, implementation of green solvents to reduce environmental impact, and automation of data processing for increased throughput. Miniaturized and high-resolution columns could further shorten analysis time and solvent consumption.

Conclusion


The presented EP method provides reliable and efficient impurity profiling of Cefradine, meeting all system suitability requirements and ensuring pharmaceutical quality. Its adoption in QC laboratories can enhance the safety and compliance of Cefradine drug products.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Impurities test for Ampicillin Sodium (EP-8.0 method)
Impurities test for Ampicillin Sodium (EP-8.0 method)
|Thermo Fisher Scientific|Applications
Impurities test for Ampicillin Sodium (EP-8.0 method): SAMPLE PREPARATION: Reference solution (a): Dissolve 27.0 mg of Ampicillin in Mobile phase A and dilute to 50ml with mobile phase A. Reference solution (b): Dissolve 2mg of cefradine CRS in mobile phase…
Key words
ampicillin, ampicillincefradine, cefradinerrt, rrtdimer, dimerreference, referencemobile, mobilephase, phasecriteria, criteriaobtained, obtainedparameters, parametersresolution, resolutionsolution, solutionmin, minresults, resultstime
Assay Analysis for Cefalexin (EP- method)
Assay Analysis for Cefalexin (EP- method)
|Thermo Fisher Scientific|Applications
Assay Analysis for Cefalexin (EP- method): SAMPLE PREPARATION: Reference solution-a: Dissolve 50 mg of Cefalexin monohydrate in water and dilute to 100 ml with water. Reference solution-b: Dissolve 10 mg of Cefradine in 20 ml of reference solution (a) and…
Key words
cefradine, cefradinecefalexin, cefalexinreference, referencesolution, solutionsuitability, suitabilitycriteria, criteriaobtained, obtainedparameters, parametersresolution, resolutionbetween, betweensystem, systemresults
Impurities test for Cefalexin (EP-8.0 method)
Impurities test for Cefalexin (EP-8.0 method)
|Thermo Fisher Scientific|Applications
Impurities test for Cefalexin (EP-8.0 method): SAMPLE PREPARATION: Test solution: Dissolve 50 mg of the substance to be examined in Mobile phase A and dilute to 50.0 ml with mobile phase A Reference solution (a): Dissolve 10 mg of D-phenylglycine…
Key words
cent, centimp, impphase, phasemobile, mobilecefotaxime, cefotaximecefalexin, cefalexinper, perresolution, resolutiondilute, dilutebetween, betweenusp, uspcriteria, criteriamix, mixobtained, obtainedparameters
Impurities test for Hydrochlorothiazide (EP- method)
Impurities test for Hydrochlorothiazide (EP- method)
|Thermo Fisher Scientific|Applications
Impurities test for Hydrochlorothiazide (EP- method): SAMPLE PREPARATION: Reference solution-a: Dissolve 3mg of Impurity-A and 3mg of Hydrochlorothiazide in 5ml of mixture of equal volumes of acetonitrile and methanol if necessary dilute to 20ml with phosphate buffer solution pH 3.2.…
Key words
hydrochlorothiazide, hydrochlorothiazidemobile, mobilephase, phaseusp, uspimpurity, impuritycriteria, criteriareference, referenceobtained, obtainedparameters, parametersresolution, resolutionbetween, betweensolution, solutionmin, minresults, resultstime
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike