Poly-2-methyl-2-oxazoline using Acclaim C30

Applications | 2013 | Thermo Fisher ScientificInstrumentation
HPLC, Consumables, LC columns
Industries
Energy & Chemicals
Manufacturer
Thermo Fisher Scientific

Summary

Significance of the topic


Characterizing cationic polymers with quaternary ammonium end groups, such as cetylpyridinium-functionalized poly(2-methyl-2-oxazoline), is essential for ensuring consistent antimicrobial performance in personal care and pharmaceutical formulations.
Precise separation and detection of these components support quality assurance and regulatory compliance in products like mouthwash and cough drops.

Objectives and study overview


This study aimed to develop and validate an HPLC-based method for the characterization of cetylpyridinium and poly(2-methyl-2-oxazoline) using reversed-phase chromatography with conductivity and aerosol detection.
The work highlights column selection, mobile phase design, and detection mode optimization for interference-free analysis.

Methodology and instrumentation


Polymerization was initiated with methyl p-toluenesulfonate at a molar ratio of 1:10 (initiator to monomer) to yield poly(2-methyl-2-oxazoline) with a toluenesulfonate end group.
Two stationary phases were evaluated: Acclaim Surfactant Plus for suppressed conductivity detection and Acclaim C30 for strong hydrophobic retention.
Mobile phases comprised acetonitrile, water with formic acid, and 0.1 M ammonium acetate buffer (pH 5.2).
A gradient elution with an inverted, offset profile was applied to ensure uniform detector response.
Separations were performed at 25 °C with a 0.5 mL/min flow rate.
Detection was achieved using suppressed conductivity (Dionex CSRS 300) and Corona Ultra Charged Aerosol Detector with post-column makeup flow.

Main results and discussion


Both columns provided sharp peaks and good resolution of the target analytes.
The Surfactant Plus column efficiently separated toluenesulfonate from cetylpyridinium, yielding reproducible conductivity signals.
The C30 column demonstrated strong retention of the hydrophobic polymer, facilitating accurate molecular weight distribution profiling.
The charged aerosol detector exhibited consistent mass response across the gradient after the addition of the makeup flow, enabling quantitation of polymer fractions.

Benefits and practical applications


  • High sensitivity and selectivity for quaternary ammonium species.
  • Robust separation of monomeric and polymeric components.
  • Compatibility with highly aqueous mobile phases.
  • Mass-sensitive detection for non-UV-active polymers.

Future trends and applications


  • Integration with mass spectrometry for structural elucidation.
  • Real-time reaction monitoring in flow systems.
  • Expansion to other functionalized polymers and complex matrices.
  • Automated high-throughput screening for quality control.

Conclusion


The combined use of specialized stationary phases and complementary detection techniques allows comprehensive analysis of cetylpyridinium-functionalized poly(2-methyl-2-oxazoline).
This approach delivers reliable separation and quantitation, supporting the development and quality assessment of antimicrobial polymer formulations.

Instrumentation used


  • Thermo Scientific Dionex ICS-3000 IC system
  • Thermo Scientific Acclaim Surfactant Plus column (3 µm, 3.03 × 150 mm)
  • Thermo Scientific Acclaim C30 column (3 µm, 3.03 × 150 mm)
  • Suppressed conductivity detector (Dionex CSRS 300, 2 mm)
  • Thermo Scientific Dionex Corona Ultra Charged Aerosol Detector

Reference


No external references were cited in this summary.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Polyunsaturated Lipids in a Nutritional Supplement Using Accucore C30
Cetylpyridinium Polyunsaturated Lipids in in Mouthwash a NutritionalUsing Supplement Acclaim Surfactant Plus with Conductivity Detection Using Accucore C30 ™ ™ Column: Thermo Scientific™ Accucore™ C30 Column: Thermo Scientific Acclaim Surfactant Plus Dimensions: 3.0 µm, 2.7 3.0 µm,×4.6 100 mm Dimensions:…
Key words
accucore, accucorescientific, scientificcetylpyridinium, cetylpyridiniumthermo, thermosubtraction, subtractionconductivity, conductivitysupplement, supplementsurfactant, surfactantacclaim, acclaimacetateacid, acetateacidblendsurfactant, blendsurfactantflaxseedplus, flaxseedplusnutritionalusing, nutritionalusingoilseparates, oilseparatespyridiniumfatty
Electrolytes in Sports Beverages Using Acclaim Trinity P2
Electrolytes in Sports Beverages Using Acclaim Trinity P2
2013|Thermo Fisher Scientific|Applications
Cetylpyridinium in Mouthwash Using Trinity P2 Electrolytes in Sports Beverages Using Acclaim Acclaim Surfactant Plus with Conductivity Detection 1 300 45 pA 4 mAU 2 7 0 3 32 5 UV 220 nm 6 1 9 10 11 2 µS…
Key words
acclaim, acclaimdionex, dionexsurfactant, surfactantscientific, scientificconductivity, conductivitycetylpyridinium, cetylpyridiniumsports, sportselectrolytes, electrolytessubtraction, subtractionplus, plusexternalfunction, externalfunctioninafter, inafterthegood, thegoodthermodata, thermodatabeverages
Polyacrylic Acid Using Acclaim SEC-1000 with Charged Aerosol versus UV Detection
Cetylpyridinium in Mouthwash Using Polyacrylic Acid Using Acclaim SEC-1000 Acclaim Plus with Conductivity Detection withSurfactant Charged Aerosol versus UV Detection ™ Column: Thermo Thermo Scientific™ Column: Scientific Acclaim™ Surfactant Plus Acclaim™ SEC-1000 Dimensions: 3.0 µm, 3.0 × 150 mm Dimensions:Thermo…
Key words
acclaim, acclaimaerosol, aerosolcharged, chargedconductivity, conductivitypolyacrylic, polyacrylicblank, blankdionex, dionexsubtraction, subtractioncorona, coronamau, mauaresurfactant, aresurfactantcetylpyridiniumbelow, cetylpyridiniumbelowcoughhas, coughhashasmild, hasmildnmdiluted
Cationic Impurities in Colesevelam Analyzed by MPIC Using IonPac NS2
Cetylpyridinium in Colesevelam Mouthwash Using Cationic Impurities in Analyzed Acclaim Surfactant Plus withIonPac Conductivity by MPIC Using NS2 Detection 300 3.0 mAU S total ID-1 2 UV 220 nm 0 ID-2 ID-5 ID-4 1 3 ID-7 ID-3 2 µS ID-6…
Key words
colesevelam, colesevelammpic, mpicsurfactant, surfactantacclaim, acclaimconductivity, conductivitycetylpyridinium, cetylpyridiniumscientific, scientificdionex, dionexplus, plusacidssuch, acidssuchandgood, andgoodandimpurities, andimpuritiesiscough, iscoughsequesteringproducts, sequesteringproductstoagents
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike