Size Exclusion Chromatography of Gum Arabic Using Acclaim SEC-1000

Applications | 2013 | Thermo Fisher ScientificInstrumentation
LC columns, Consumables, GPC/SEC
Industries
Food & Agriculture
Manufacturer
Thermo Fisher Scientific

Summary

Significance of the Topic


Cetylpyridinium salts complexed with gum Arabic serve as mild antibacterial agents in oral care, pharmaceutical, and industrial applications. Understanding their molecular weight distribution and composition is crucial for ensuring product consistency, efficacy, and safety.

Objectives and Study Overview


This work aims to establish a robust size exclusion chromatography (SEC) method, coupled with charged aerosol and conductivity detection, to characterize gum Arabic polymers and to separate and quantify associated cetylpyridinium species.

Methodology and Instrumentation


  • Chromatograph: Thermo Scientific Dionex ICS-3000 system.
  • Column: Acclaim™ SEC-1000 (4.6 × 300 mm, 3 µm particles).
  • Mobile Phase: 100 mM ammonium acetate buffer (pH 5.0) with acetonitrile, formic acid, and water in gradient program.
  • Flow Rate: 0.50 mL/min; Column Temperature: 25 °C; Injection Volume: 5 µL.
  • Detection: Charged aerosol detector (Corona Ultra CAD), UV at 220 nm with blank subtraction, and suppressed conductivity detection (Dionex™ CSRS™ 300 suppressor with external water, 1.0 mL/min, 8 mA).

Main Results and Discussion


  • SEC-CAD analysis achieved clear resolution between high-molecular-weight gum Arabic fractions and low-molecular-weight ions (sodium and cetylpyridinium).
  • Peak symmetry and baseline stability were optimized using a volatile buffer and CAD, enhancing quantification accuracy.
  • Conductivity detection provided complementary confirmation of ionic species identity.
  • Method reproducibility was demonstrated through consistent retention times and response over multiple injections.

Benefits and Practical Applications


  • The combined SEC-CAD and conductivity approach extends analysis capabilities to non-UV-active biopolymers and ionic surfactants in complex matrices.
  • Applicable to routine quality control of mouthwash, food additives, cough drops, and pharmaceutical formulations containing gum Arabic and cetylpyridinium.
  • Use of volatile mobile phases simplifies coupling to mass spectrometry for further structural insights.

Future Trends and Possibilities


  • Emerging column technologies (e.g., sub-2 µm particles, mixed-mode media) will improve resolution of complex polymer–surfactant systems.
  • Integration with high-resolution mass spectrometry may provide detailed compositional and structural data.
  • Automation and advanced data analytics could enable real-time process monitoring and enhanced quality control in industrial settings.

Conclusion


The developed SEC method with Acclaim™ SEC-1000 and charged aerosol plus conductivity detection offers a sensitive, reproducible platform for characterizing gum Arabic and cetylpyridinium complexes. Its straightforward implementation and compatibility with volatile buffers support broad use in research and industrial laboratories.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Size Exclusion Chromatography of Polyquaternium-10 Using Acclaim SEC-1000
Cetylpyridinium in Mouthwash Using Size Exclusion Chromatography of Polyquaternium-10 Acclaim Surfactant withSEC-1000 Conductivity Detection Using Plus Acclaim ™ Column: Thermo Acclaim SEC-1000 Column: Scientific Acclaim™ Surfactant Plus Dimension: 3.0 µm, 4.63.0 x 300 mmmm Dimensions: × 150 ™ ™ Mobile…
Key words
acclaim, acclaimsurfactant, surfactantconductivity, conductivitycetylpyridinium, cetylpyridiniumammonium, ammoniumhighseparates, highseparatesinsalt, insaltmayplus, mayplusquaternary, quaternarymmmm, mmmmantistatic, antistatichydroxyethyl, hydroxyethylattraction, attractionpyridinium, pyridiniummouthwash
Polyacrylic Acid Using Acclaim SEC-1000 with Charged Aerosol versus UV Detection
Cetylpyridinium in Mouthwash Using Polyacrylic Acid Using Acclaim SEC-1000 Acclaim Plus with Conductivity Detection withSurfactant Charged Aerosol versus UV Detection ™ Column: Thermo Thermo Scientific™ Column: Scientific Acclaim™ Surfactant Plus Acclaim™ SEC-1000 Dimensions: 3.0 µm, 3.0 × 150 mm Dimensions:Thermo…
Key words
acclaim, acclaimaerosol, aerosolcharged, chargedconductivity, conductivitypolyacrylic, polyacrylicblank, blankdionex, dionexsubtraction, subtractioncorona, coronamau, mauaresurfactant, aresurfactantcetylpyridiniumbelow, cetylpyridiniumbelowcoughhas, coughhashasmild, hasmildnmdiluted
Poly-2-methyl-2-oxazoline using Acclaim C30
Poly-2-methyl-2-oxazoline using Acclaim C30
2013|Thermo Fisher Scientific|Applications
Cetylpyridinium in Mouthwash Using C30 Poly-2-methyl-2-oxazoline using Acclaim Acclaim Surfactant Plus with Conductivity Detection 14300 pA ™ ™ Column: Thermo Scientific™ Acclaim™ C30Plus Column: Thermo Scientific Acclaim Surfactant Dimensions: 3.0 µm, 3 µm, 150mm mm Dimensions: 3.03××150 ™ ™ LC…
Key words
acclaim, acclaimsurfactant, surfactantcetylpyridinium, cetylpyridiniumscientific, scientificinitiated, initiatedpolymerization, polymerizationconductivity, conductivitythermo, thermoandweight, andweightareofmild, areofmildgoodnarrow, goodnarrowmeotsminutes, meotsminuteschci, chcidionexdionex, dionexdionexacclaimand
Quaternary Surfactants in a Cosmetic Formulation Using Acclaim Surfactant Plus
Cetylpyridinium Quaternary SurfactantsininMouthwash a CosmeticUsing Formulation Acclaim Surfactant Plus with Conductivity Using Acclaim Surfactant Plus Detection 300 120 pA mAU 6 1 2 3 4 2 5 0 6 1 2 1 3 5 µS 1 0 2 2 ™…
Key words
surfactant, surfactantacclaim, acclaimscientific, scientificconductivity, conductivityplus, plusquaternary, quaternarysubtraction, subtractionsurfactants, surfactantsthermo, thermoandactive, andactiveareinterference, areinterferencecosmeticusing, cosmeticusingcurrentcorona, currentcoronaplusmm, plusmmsilicacetylpyridinium
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike