Xevo TQ Absolute XR: Maximum Robustness and Sensitivity For High-Throughput Bioanalysis

Applications | 2025 | WatersInstrumentation
LC/MS, LC/MS/MS, LC/QQQ
Industries
Clinical Research
Manufacturer
Waters

Summary

Importance of the Topic


The reliable quantification of drug candidates, their metabolites and biomarkers in biological fluids is a cornerstone of modern pharmaceutical research and development. High-throughput bioanalysis demands instrumentation that maintains both sensitivity and robustness over large sample sets. The introduction of advanced ion guide technology promises to address common challenges such as matrix-induced contamination, downtime for cleaning and variability in signal response.

Objectives and Study Overview


This study aimed to evaluate the long-term performance of the StepWave XR ion guide integrated into the Xevo TQ Absolute XR mass spectrometer. A continuous sequence of 10,000 LC-MS/MS injections of protein-precipitated rat plasma was conducted over three weeks without unscheduled interruptions. Peak area responses for a panel of pharmaceutical analytes were monitored to assess signal stability and instrument uptime.

Methodology


Sample Preparation:
  • Rat plasma (100 µL) precipitated with cold acetonitrile (200 µL), vortexed and chilled at –20 °C for 1 hour.
  • Centrifugation at 14,000 g for 5 minutes; supernatant diluted 1:1 with water containing an analyte mix (5 ng/mL each).

Chromatography:
  • ACQUITY Premier UPLC with an HSS T3 2.1 × 50 mm, 1.7 µm column at 60 °C.
  • Gradient elution from 5% to 95% organic in 1 minute at 600 µL/min, with rapid re-equilibration.

Mass Spectrometry:
  • Xevo TQ Absolute XR with positive/negative ESI switching and multiple reaction monitoring.
  • MRM transitions optimized for 12 analytes and a stable-isotope-labeled internal standard.

Instrumentation


  • LC System: ACQUITY Premier UPLC with Binary Solvent Manager and Flow-Through Needle.
  • Mass Spectrometer: Xevo TQ Absolute XR featuring the StepWave XR slotted bandpass ion guide.
  • Software: waters_connect for Quantitation for real-time performance monitoring.

Main Results and Discussion


Over 10,000 consecutive injections:
  • Coefficient of variation (%CV) for individual analyte peak areas ranged from 13.1% to 20.4% with no observable signal decline.
  • Internal standard-corrected peak area ratios for gefitinib and its O-desmethyl metabolite exhibited %CVs of 2.3% and 2.7%, respectively.
  • No trend of sensitivity loss was detected, demonstrating effective mitigation of high-mass matrix ions by the StepWave XR guide and protection of the Q1 quadrupole.

Routine source cleaning was simplified by tool-free removal of the sampling cone without breaking vacuum, ensuring minimal downtime.

Benefits and Practical Applications


  • Exceptional uptime over large batches reduces the need for frequent maintenance interruptions.
  • Consistent sensitivity supports quantification at low pg/mL levels, critical for pharmacokinetic studies and biomarker assays.
  • Rapid, tool-free source cleaning expedites maintenance and enhances laboratory throughput.
  • Integrated software monitoring allows automated flagging of performance deviations using internal standards.

Future Trends and Potential Applications


Ongoing developments are likely to focus on further integration of remote diagnostics and predictive maintenance using AI algorithms. Emerging ion guide designs may offer even finer control of matrix suppression. The combination of high-throughput LC-MS/MS with advanced data analytics will expand applications into areas such as metabolomics, microdosing studies and real-time therapeutic monitoring.

Conclusion


The StepWave XR ion guide in the Xevo TQ Absolute XR mass spectrometer demonstrated outstanding robustness and sensitivity across 10,000 high-throughput bioanalysis injections. The system maintained stable signal responses without unscheduled downtime, simplified maintenance procedures and provided real-time performance tracking. This platform is well suited to demanding pharmaceutical and clinical bioanalytical workflows.

Reference


1. Ackermann BL, Berna MJ, Murphy AT. Recent advances in use of LC/MS/MS for quantitative high-throughput bioanalytical support of drug discovery. Curr Top Med Chem. 2002;2(1):53–66.
2. Lappin G, Wagner CC, Langer O, van de Merbel N. New ultrasensitive detection technologies and techniques for use in microdosing studies. Bioanalysis. 2009;1(2):357–66.
3. Thakur A, Tan Z, Kameyama T, El-Khateeb E, Nagpal S, Malone S, Jamwal R, Nwabufo CK. Bioanalytical strategies in drug discovery and development. Drug Metab Rev. 2021;53(3):434–458.
4. Miller VP. SPE-MS analysis of absorption, distribution, metabolism and excretion assays: a tool to increase throughput and streamline workflow. Bioanalysis. 2012;4(9):1111–21.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Application of a Slotted Bandpass Ion Guide to Increase Robustness in Tandem Quadrupole LC-MS/MS Bioanalysis
Application of a Slotted Bandpass Ion Guide to Increase Robustness in Tandem Quadrupole LC-MS/MS Bioanalysis Abstract ID number: 322533 1Noelle Elliott, 1Peter Hancock, 2Robert Plumb 1Waters Corporation, Stamford Ave, Wilmslow, Cheshire, UK, 2Waters Corporation, 34 Maple St, Milford, MA, USA.…
Key words
gefitinib, gefitinibinjections, injectionsarea, areapropafenone, propafenonenefazodone, nefazodoneresponse, responsepeak, peaknifedipine, nifedipineguide, guidenumber, numberresolving, resolvingdextromethorphan, dextromethorphanion, ionrat, ratphenacitin
A Streamlined Workflow for Quantitative Bioanalysis using waters_connect for Quantitation Software: A Case Study Using Gefitinib
Application Note A Streamlined Workflow for Quantitative Bioanalysis using waters_connect for Quantitation Software: A Case Study Using Gefitinib Robert Plumb Waters Corporation, United States Published on November 20, 2025 Abstract Preclinical and early human drug metabolism and pharmacokinetics (DMPK) studies…
Key words
gefitinib, gefitinibstudy, studypharmacokinetics, pharmacokineticsusing, usingplasma, plasmaxevo, xevodrug, drugabsolute, absolutesubcutaneous, subcutaneousmrm, mrmpreclinical, preclinicalpremier, premiermetabolites, metabolitesacquity, acquityrat
Robustness in Regulated Bioanalysis: A 30,000 Injection Study of Naltrexone in Human Plasma using the Xevo™ TQ Absolute XR Mass Spectrometer
Application Note Robustness in Regulated Bioanalysis: A 30,000 Injection Study of Naltrexone in Human Plasma using the Xevo™ TQ Absolute XR Mass Spectrometer Henry Foddy1, David Farrell2, Sally Hannam2, Peter Hancock1, Nikunj Tanna3, Robert S. Plumb3 1 Waters Corporation, Wilmslow,…
Key words
absolute, absolutespectrometer, spectrometermass, massxevo, xevopremier, premierprivacy, privacyrobustness, robustnessacquity, acquityacquitytm, acquitytmpreserves, preservesacross, acrossaccelerates, acceleratespharmacokinetic, pharmacokineticconditions, conditionsquadrupole
Improvements in Robustness of Signal Response in Multiresidue Pesticide Analysis with Xevo TQ Absolute XR Mass Spectrometer
Application Note Improvements in Robustness of Signal Response in Multiresidue Pesticide Analysis with Xevo TQ Absolute XR Mass Spectrometer David Gould, Peter Hancock, Stuart Adams Waters Corporation, United States Published on June 02, 2025 Contact Sales Abstract Typical LC-MS/MS pesticide…
Key words
multiresidue, multiresiduexevo, xevoabsolute, absoluteimprovements, improvementspesticide, pesticiderobustness, robustnessspectrometer, spectrometersignal, signalresponse, responsemass, massanalysis, analysissolvent, solventappendix, appendixprivacy, privacypremier
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike