Summary

Significance of the Topic


The reliability and sensitivity of tandem quadrupole LC-MS/MS bioanalysis are often challenged by matrix-derived high-mass ions that accumulate in the MS1 quadrupole, causing signal suppression and frequent downtime for maintenance. Introducing a slotted bandpass resolving ion guide enhances robustness by selectively transmitting target m/z ions while excluding undesired high-mass contaminants, thus maintaining consistent performance during large-scale studies.

Objectives and Study Overview


  • Present a novel slotted bandpass ion guide integrating radial DC and axial fields to filter high-mass ions upstream of the MS1 quadrupole.
  • Assess long-term stability and reproducibility during more than 10,000 injections of protein-precipitated rat plasma samples.
  • Demonstrate quantitative analysis of the anticancer drug gefitinib and its major metabolite in rat plasma and urine following subcutaneous dosing.

Methodology


Rat plasma (100 µL) was protein-precipitated with 200 µL cold acetonitrile containing internal standards. After vortexing, cold incubation, and centrifugation, the supernatant was diluted 1:1 with water. Two microliters of extract were injected onto an ACQUITY Premier UPLC system coupled to a Xevo TQ Absolute XR MS. Chromatography employed a 2.1×50 mm HSS T3 (1.7 µm) column at 60 °C with a 5–95 % B gradient over one minute at 600 µL/min (A: 0.1 % formic acid in water; B: 95:5 acetonitrile/water with 0.1 % formic acid). Analytes were monitored in positive/negative ESI MRM mode using optimized transitions.

Instrumentation Used


  • ACQUITY Premier UPLC System
  • Xevo TQ Absolute XR Tandem Quadrupole Mass Spectrometer
  • StepWave slotted bandpass resolving ion guide

Results and Discussion


During ten successive batches of 1,000 injections each (total 10,000), the system maintained uninterrupted acquisition without vacuum breaks or cone cleaning. The raw peak area coefficients of variation for key analytes (nefazodone, dextromethorphan, nifedipine, propafenone and others) ranged from 13.1 to 20.4 %, demonstrating high reproducibility. Peak area ratios for gefitinib and its O-desmethyl metabolite remained within 0.9–1.4 across the injection series. Calibration curves exhibited linearity over the expected concentration range, and pharmacokinetic profiling in male Wistar rats (10 mg/kg subcutaneously) yielded clear plasma and urine concentration–time profiles, confirming assay suitability for in vivo studies.

Benefits and Practical Applications


  • Enhanced uptime and throughput by eliminating frequent source cleaning and vacuum interruptions.
  • Consistent sensitivity and precision over extensive sample batches.
  • Streamlined workflow for high-throughput bioanalysis in drug discovery and development.

Future Trends and Opportunities


Expansion of slotted bandpass ion guide technology may include tunable mass filtering for broader analyte classes, integration with automated sample preparation, and application to metabolomics or proteomics. Further design refinements could improve selectivity and reduce chemical noise, opening paths for ultra-trace analyses in complex matrices.

Conclusion


The slotted bandpass resolving ion guide demonstrates a substantial advance in LC-MS/MS robustness, enabling over 10,000 injections of protein-precipitated plasma without loss of performance. This innovation reduces maintenance demands and supports reliable, high-throughput quantitative bioanalysis.

References


  • Ackermann BL, Berna MJ, Murphy AT. Recent advances in use of LC/MS/MS for quantitative high-throughput bioanalytical support of drug discovery. Curr Top Med Chem. 2002;2(1):53–66.
  • Lappin G, Wagner CC, Langer O, van de Merbel N. New ultrasensitive detection technologies and techniques for use in microdosing studies. Bioanalysis. 2009;1(2):357–366.
  • Thakur A et al. Bioanalytical strategies in drug discovery and development. Drug Metab Rev. 2021;53(3):434–458.
  • Miller VP. SPE-MS analysis of ADME assays: a tool to increase throughput and streamline workflow. Bioanalysis. 2012;4(9):1111–1121.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Xevo TQ Absolute XR: Maximum Robustness and Sensitivity For High-Throughput Bioanalysis
Application Note Xevo TQ Absolute XR: Maximum Robustness and Sensitivity For High-Throughput Bioanalysis Robert S. Plumb Waters Corporation, United States Published on June 02, 2025 Contact Sales Abstract The accurate LC-MS/MS quantification of candidate pharmaceuticals, their metabolites and biomarkers in…
Key words
xevo, xevoabsolute, absolutebioanalysis, bioanalysisslotted, slottedbandpass, bandpassspectrometer, spectrometerrobustness, robustnessmrm, mrmprivacy, privacymass, masspremier, premierquadrupole, quadrupolemonitoring, monitoringdependent, dependentbatch
A Streamlined Workflow for Quantitative Bioanalysis using waters_connect for Quantitation Software: A Case Study Using Gefitinib
Application Note A Streamlined Workflow for Quantitative Bioanalysis using waters_connect for Quantitation Software: A Case Study Using Gefitinib Robert Plumb Waters Corporation, United States Published on November 20, 2025 Abstract Preclinical and early human drug metabolism and pharmacokinetics (DMPK) studies…
Key words
gefitinib, gefitinibstudy, studypharmacokinetics, pharmacokineticsusing, usingplasma, plasmaxevo, xevoabsolute, absolutesubcutaneous, subcutaneousdrug, drugpremier, premierpreclinical, preclinicalmrm, mrmmetabolites, metabolitesrat, ratspectrometer
Versatile LC-MS/MS Solutions for the Evolving Bioanalysis Landscape
[ APPLICATION NOTEBOOK ] Versatile LC-MS/MS Solutions for the Evolving Bioanalysis Landscape Small Molecules Oligonucleotide Protein/Peptide Therapeutics Software Tools Targeted Imaging LC-MS/MS Solutions to Support Reliable, Sensitive and Efficient Workflows for the Bioanalysis of Drug Modalities Welcome to the Waters…
Key words
bioanalysis, bioanalysisimaging, imagingoligonucleotide, oligonucleotidetherapeutics, therapeuticsbioanalytical, bioanalyticalpeptide, peptidemolecules, moleculestargeted, targetedprotein, proteinhighlights, highlightstools, toolsread, readquantitative, quantitativequantification, quantificationsmall
Rapid, High Sensitivity Quantification of a Proteolysis Targeting Chimera, PROTAC, in Rat Plasma From a Microsampling DMPK Study Using UPLC-MS/MS
Application Note Rapid, High Sensitivity Quantification of a Proteolysis Targeting Chimera, PROTAC, in Rat Plasma From a Microsampling DMPK Study Using UPLC-MS/MS Robert S. Plumb Waters Corporation Abstract Proteolysis targeting chimeras are bifunctional molecules with two molecular entities connected by…
Key words
dmpk, dmpkmicrosampling, microsamplinguplc, uplcstudy, studyprotacs, protacsusing, usingpremier, premieracquity, acquityproteolysis, proteolysistargeting, targetingmolecules, moleculesprivacy, privacyability, abilityformate, formatensclc
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike