Multiple Mass Spectrometric Strategies for High Selectivity Quantification of Proteins and Peptides
Applications | 2015 | SCIEXInstrumentation
The rapid identification of protein biomarkers has created an urgent need for quantification methods capable of confirming candidate utility in large validation cohorts. High selectivity is crucial to distinguish low-abundance targets from complex biological backgrounds and to support accelerated workflows.
This work examines three mass spectrometric strategies on the QTRAP 5500 System for selective peptide measurement, using Brain Natriuretic Peptide as a model analyte. Comparative evaluation of Multiple Reaction Monitoring, enhanced MRM with MS3 fragmentation, and single ion monitoring with differential mobility separation aims to address challenging analytical scenarios.
Sample Preparation included precipitation of plasma spiked with BNP at varying ratios of acetonitrile to identify optimal recovery. HPLC separation utilized a C18 column under a linear gradient of acetonitrile in formic acid. Quantification methods were developed for MRM, MS3-based MRM workflow, and SIM scans with or without differential mobility separation.
A plasma precipitation ratio of 1 to 1 achieved the highest recovery of BNP. Both MRM and MS3 modes yielded a lower limit of quantitation of 128 picograms per on-column injection, corresponding to an effective 512 pg/mL in plasma. SIM alone suffered from matrix interference, but integration of differential mobility separation reduced chemical noise and improved the LLOQ by approximately 25-fold. The advanced workflows demonstrated clear enhancements in signal to noise without extensive sample cleanup.
Further optimization of sample preparation and chromatography can push sensitivity limits even lower. Expanding these workflows to multiplexed assays and applying alternative chemical modifiers in mobility separation will broaden applicability. Integration with data-driven acquisition and real-time decision support may accelerate biomarker validation and clinical translation.
The QTRAP 5500 System enables highly selective quantification of peptides through complementary MS strategies. MRM, MS3-based workflows, and SIM combined with differential mobility separation each address distinct analytical challenges, delivering low detection limits and robust performance in complex biological samples.
LC/MS, LC/MS/MS, LC/QTRAP
IndustriesProteomics
ManufacturerSCIEX
Summary
Significance of the Topic
The rapid identification of protein biomarkers has created an urgent need for quantification methods capable of confirming candidate utility in large validation cohorts. High selectivity is crucial to distinguish low-abundance targets from complex biological backgrounds and to support accelerated workflows.
Objectives and Study Overview
This work examines three mass spectrometric strategies on the QTRAP 5500 System for selective peptide measurement, using Brain Natriuretic Peptide as a model analyte. Comparative evaluation of Multiple Reaction Monitoring, enhanced MRM with MS3 fragmentation, and single ion monitoring with differential mobility separation aims to address challenging analytical scenarios.
Methodology
Sample Preparation included precipitation of plasma spiked with BNP at varying ratios of acetonitrile to identify optimal recovery. HPLC separation utilized a C18 column under a linear gradient of acetonitrile in formic acid. Quantification methods were developed for MRM, MS3-based MRM workflow, and SIM scans with or without differential mobility separation.
Instrumentation Used
- SCIEX QTRAP 5500 System with Turbo V ion source
- SelexION differential mobility device
- Shimadzu HPLC with Thermo Aquasil C18 column
Key Results and Discussion
A plasma precipitation ratio of 1 to 1 achieved the highest recovery of BNP. Both MRM and MS3 modes yielded a lower limit of quantitation of 128 picograms per on-column injection, corresponding to an effective 512 pg/mL in plasma. SIM alone suffered from matrix interference, but integration of differential mobility separation reduced chemical noise and improved the LLOQ by approximately 25-fold. The advanced workflows demonstrated clear enhancements in signal to noise without extensive sample cleanup.
Benefits and Practical Applications
- High selectivity workflows reduce false positives and improve accuracy in complex matrices
- MS3 mode provides an orthogonal fragmentation-based filter for challenging targets
- Differential mobility separation offers rapid interference removal without altering chromatography
- Versatile platform supports a range of peptides and assay throughput requirements
Future Trends and Opportunities
Further optimization of sample preparation and chromatography can push sensitivity limits even lower. Expanding these workflows to multiplexed assays and applying alternative chemical modifiers in mobility separation will broaden applicability. Integration with data-driven acquisition and real-time decision support may accelerate biomarker validation and clinical translation.
Conclusion
The QTRAP 5500 System enables highly selective quantification of peptides through complementary MS strategies. MRM, MS3-based workflows, and SIM combined with differential mobility separation each address distinct analytical challenges, delivering low detection limits and robust performance in complex biological samples.
References
- MRM Quantitation for Highest Specificity in Complex Matrices. SCIEX Technical Note RUO-MKT-02-2739-A
- SCIEX SelexION Technology A New Solution to Selectivity Challenges in Quantitative Bioanalysis RUO-MKT-02-3251-A
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Overview of LC-MS Quantitative Solutions for Biotherapeutic Analysis
2019|SCIEX|Others
Overview of LC-MS Quantitative Solutions for Biotherapeutic Analysis Featuring SCIEX Instrumentation and Technologies Lei Xiong, Elliott Jones SCIEX, Redwood City, California, USA Figure 1. Summary of SCIEX Quantitative Solutions and Instrumentations to Solve Analytical Challenges in Biotherapeutic Industry. Darkest green…
Key words
sciex, sciexmrmhr, mrmhrworkflow, workflowprotein, proteinmicrolc, microlcquantification, quantificationmrm, mrmions, ionsswath, swathqtrap, qtrapquantitation, quantitationhrams, hramsspectrometers, spectrometersmobility, mobilityselexion
SelexION® Technology: The Solution to Selectivity Challenges in Quantitative Analysis
2015|SCIEX|Technical notes
SelexION® Technology: The Solution to Selectivity Challenges in Quantitative Analysis Differential Mobility Separations Enhanced with Chemical Modifiers: A Higher Dimension in Selectivity High selectivity is a key component of successful quantitative analysis. The ever-increasing sensitivity and throughput requirements of MS…
Key words
selexion, selexiondifferential, differentialmobility, mobilitydms, dmsday, dayplanar, planarcov, covmrm, mrmtechnology, technologybioanalytical, bioanalyticalintra, intrahigh, highcell, cellfield, fieldkey
SCIEX Differential Mobility Spectrometry Applications - See How SelexION® DMS Technology Helps Overcome Your Biggest Analytical Challenges
2016|SCIEX|Brochures and specifications
Peptide and Protein Quantitation Lipidomics Differential Mobility Spectrometry Applications Small Molecule Bioanalysis and Metabolism Forensics Food and Environmental Testing See How SelexION® DMS Technology Helps Overcome Your Biggest Analytical Challenges A New Dimension in Selectivity As analytical challenges become more…
Key words
dms, dmsselexion, selexionmobility, mobilitydifferential, differentiallipid, lipidexplore, exploretechnology, technologyquantitation, quantitationpeptides, peptidesselectivity, selectivityhow, howlipidomics, lipidomicsbioanalysis, bioanalysisfood, foodmetabolism
Quantitation of Limaprost, an Analogue of PGE1 in Human Plasma
2017|SCIEX|Applications
Quantitation of Limaprost, an Analogue of PGE1 in Human Plasma A highly simplified and sensitive assay using the QTRAP® 6500+ LC-MS/MS system and the SelexION®+ differential mobility separation technology device 1 Chao Zhang; 1Dandan Si; 2Xiangjun Meng; 2Lei Yin; 2Jingkai…
Key words
selexion, selexionlimaprost, limaprostdms, dmsmobility, mobilitydevice, devicetechnology, technologysubpg, subpgantiplatelet, antiplateletdifferential, differentialvasodilatory, vasodilatoryseparation, separationlloq, lloqinterferences, interferencesendogenous, endogenousions