Improved Linearity and Quantification Using the SPD-M40 Photodiode Array Detector - Analytical Intelligence Part 4 -

Technical notes | 2019 | ShimadzuInstrumentation
HPLC
Industries
Manufacturer
Shimadzu

Summary

Importance of the Topic


The accurate measurement of absorbance in liquid chromatography is fundamental for quantitative and qualitative analysis in pharmaceuticals, environmental monitoring and general analytical chemistry. Stray light and electronic noise are two key factors that limit the linearity and sensitivity of UV-VIS and photodiode array (PDA) detectors. Improving these parameters expands the dynamic range of measurement and enhances the reliability of trace-level quantification.

Objectives and Study Overview


This report examines the impact of stray light and noise on detector linearity and quantification accuracy. It introduces a novel SPD-M40 photodiode array detector designed to minimize stray light and reduce noise. The study evaluates the detector’s linearity up to 2.5 absorbance units (AU) and its capability to quantify analytes across a wide concentration span.

Methodology and Instrumentation


The principle of PDA detection involves irradiating the flow cell with white light and spectrally resolving transmitted light. Stray light from reflection, scattering, grating imperfections and lamp noise was characterized by measuring its intensity relative to incident light. Detector noise was assessed at different time constants. A standard ketoprofen solution spanning 0.5 to 800 mg/L was analyzed under optimized chromatographic conditions to generate calibration data.
  • Light source: Deuterium/ Tungsten lamp
  • Detector: SPD-M40 photodiode array
  • Optical elements: high‐performance grating and clean flow cell to limit stray light
  • Electronic layout: optimized for low noise at both fast (25 ms) and slow (2 s) time constants


Main Results and Discussion


Stray light was reduced to one third of levels seen in conventional PDA detectors, extending linearity beyond 2.5 AU. Modeling showed that stray light intensities above 0.1% of incident light significantly degrade linearity beyond 2 AU. The SPD-M40 consistently achieved linear response up to 2.5 AU with error under 5%. Noise measurements demonstrated a marked reduction at both short and long time constants, improving detection limits in the low absorbance range.
  • Linearity coefficient for ketoprofen calibration: R2 = 0.9998 over 0.5–800 mg/L
  • Reproducibility: area percent RSD ≤ 1% for impurities at 0.1% level
  • Dynamic range: simultaneous quantification of major components at high mg/L levels and impurities at sub-mAU absorbance


Benefits and Practical Applications


The improvements in stray light suppression and noise reduction allow analysts to:
  • Quantify low-concentration impurities alongside high-concentration active ingredients in a single run
  • Achieve reliable results over an extended absorbance range without detector saturation
  • Enhance throughput by reducing the need for sample dilution or multiple injections


Future Trends and Opportunities


Advances may include integration of real-time stray light correction algorithms, further miniaturization of optical components for portable analyzers and combination with machine learning methods to predict and compensate for detector drift. Expansion into UV-VIS hybrid detectors and enhanced software tools will continue to drive sensitivity and linearity improvements.

Conclusion


The SPD-M40 photodiode array detector effectively minimizes stray light and electronic noise, achieving linearity beyond 2.5 AU and enabling accurate quantification across a broad concentration range. These enhancements support robust analytical workflows in pharmaceutical impurity profiling and quality control applications.

References


  • Application News No L538 Impurity Analysis in Pharmaceutical Products with the Advanced Photodiode Array Detector SPD-M40

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Improvement of pharmaceutical impurity analysis by reducing the effects of stray light and room temperature fluctuations on PDA detector
Improvement of pharmaceutical impurity analysis by reducing the effects of stray light and room temperature fluctuations on PDA detector Takato Uchikata1, Satoshi Akita1, Hidetoshi Terada1, Masato Watanabe1, Keiko Matsumoto1; 1Shimadzu Corporation, Kyoto, Japan 1. Introduction A photodiode array detector (PDA)…
Key words
pda, pdafluctuations, fluctuationsimpurity, impuritystray, strayketoprofen, ketoprofenbaseline, baselineabsorbance, absorbancetemperaturecontrolled, temperaturecontrolledtemperature, temperaturepharmaceutical, pharmaceuticallight, lightindispensable, indispensabledetection, detectionnon, nonphotodiode
Ultra High Performance Liquid Chromatograph Shimadzu Nexera series - Technical Reports / Applications
C190-E256 Ultra High Performance Liquid Chromatograph Nexera series Technical Reports / Applications The Nexera Series: Reinvent Your Workflow Just as the internet and smartphones have changed our lifestyles, and robots with AI capabilities continue to make our lives more convenient,…
Key words
acid, acidmau, mauanalysis, analysismobile, mobilemin, mincolumn, columnphase, phaseshim, shimnexera, nexerainjection, injectionpack, packorganic, organicatp, atpacids, acidspantothenate
Impurity Analysis in Pharmaceutical Products with the Advanced Photodiode Array Detector SPD-M40
LAAN-A-LC-E311 Application News No. L538 High Performance Liquid Chromatography Impurity Analysis in Pharmaceutical Products with the Advanced Photodiode Array Detector SPD-M40 Impurities in pharmaceuticals must be appropriately controlled to ensure product safety. The ICH guidelines on impurities in pharmaceutical substances…
Key words
ketoprofen, ketoprofenimpurity, impuritymau, mauimpurities, impuritiespeak, peakpharmaceutical, pharmaceuticalarea, areamin, minheight, heightnews, newsmain, mainpurity, purityvelox, veloxstray, strayzoomed
Impurity Analysis in Pharmaceutical Products with the Advanced Photodiode Array Detector SPD-M40
LAAN-A-LC-E311 Application News No. L538 High Performance Liquid Chromatography Impurity Analysis in Pharmaceutical Products with the Advanced Photodiode Array Detector SPD-M40 Impurities in pharmaceuticals must be appropriately controlled to ensure product safety. The ICH guidelines on impurities in pharmaceutical substances…
Key words
ketoprofen, ketoprofenimpurity, impuritymau, mauimpurities, impuritiespeak, peakpharmaceutical, pharmaceuticalarea, areamin, minheight, heightnews, newspurity, puritymain, mainvelox, veloxstray, strayzoomed
Other projects
GCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike