High Selectivity Quantification of Protein Isoforms using MRM3 Workflow
Applications | 2016 | SCIEXInstrumentation
The cytochrome P450 enzyme family plays a central role in the metabolism of drugs and xenobiotics, often forming toxic metabolites. High sequence homology among over 200 isoforms complicates selective quantification of individual proteins. The MRM3 workflow enhances analytical specificity by monitoring secondary product ions, enabling accurate measurement of low abundance isoforms in complex biological matrices.
This study demonstrates the application of the MRM3 workflow on the QTRAP 4500 system to selectively quantify the CYP3A5 isoform in human liver microsomes. It compares conventional MRM with advanced MRM3 methods to evaluate improvements in specificity, sensitivity, and lower limits of quantitation.
Sample preparation employed synthetic light and heavy isotopically labeled tryptic peptides for CYP3A5 spiked into digested liver microsomes. Calibration curves ranged from 305 amol to 805 fmol on column. Chromatographic separation was conducted on a nanoLC Ultra system with cHiPLC trap elute mode using ChromXP C18 reversed phase chips and a linear gradient from 10 to 30 percent acetonitrile over 45 minutes at 1 microliter per minute. Mass spectrometry used the QTRAP 4500 system with unit resolution MRM in Q1 and Q3 and MRM3 via MSMSMS scans. Precursor ions were isolated in Q1, fragmented in Q2, and primary fragments trapped in the linear ion trap. A selected first generation ion was isolated and subjected to resonance excitation to generate second generation fragments for detection. Data processing was performed with MultiQuant software integrating primary and secondary product ion peaks.
The MRM3 workflow delivered superior selectivity compared to conventional MRM. For the CYP3A5 peptides DTINFLSK and SLGPVGFMK, two of three MRM transitions exhibited significant interferences at low femtomole levels, while MRM3 detection of secondary product ions produced clean, interference free signals. Calibration curves spanning 1.2 fmol to 805 fmol demonstrated excellent linearity, low coefficients of variance and high accuracy at the lower limit of quantitation of 1.2 fmol on column.
MRM3 workflows are poised to expand for multiplexed quantification of diverse protein isoforms and post‐translational modifications. Advances in ion trap sensitivity and scan speed will further lower detection limits. Automated assay design and data analysis will facilitate high‐throughput targeted proteomic screening in pharmaceutical, clinical, and industrial laboratories.
The combination of quadrupole isolation and high sensitivity linear ion trap scanning on the QTRAP 4500 enables an MRM3 workflow that surpasses conventional MRM in selectivity and sensitivity. This approach is well suited for accurate quantification of low abundance protein isoforms in complex samples and advances targeted proteomic applications in drug metabolism research.
LC/MS, LC/MS/MS, LC/QTRAP
IndustriesProteomics
ManufacturerSCIEX
Summary
Importance of the topic
The cytochrome P450 enzyme family plays a central role in the metabolism of drugs and xenobiotics, often forming toxic metabolites. High sequence homology among over 200 isoforms complicates selective quantification of individual proteins. The MRM3 workflow enhances analytical specificity by monitoring secondary product ions, enabling accurate measurement of low abundance isoforms in complex biological matrices.
Objectives and Study Overview
This study demonstrates the application of the MRM3 workflow on the QTRAP 4500 system to selectively quantify the CYP3A5 isoform in human liver microsomes. It compares conventional MRM with advanced MRM3 methods to evaluate improvements in specificity, sensitivity, and lower limits of quantitation.
Methodology and Instrumentation
Sample preparation employed synthetic light and heavy isotopically labeled tryptic peptides for CYP3A5 spiked into digested liver microsomes. Calibration curves ranged from 305 amol to 805 fmol on column. Chromatographic separation was conducted on a nanoLC Ultra system with cHiPLC trap elute mode using ChromXP C18 reversed phase chips and a linear gradient from 10 to 30 percent acetonitrile over 45 minutes at 1 microliter per minute. Mass spectrometry used the QTRAP 4500 system with unit resolution MRM in Q1 and Q3 and MRM3 via MSMSMS scans. Precursor ions were isolated in Q1, fragmented in Q2, and primary fragments trapped in the linear ion trap. A selected first generation ion was isolated and subjected to resonance excitation to generate second generation fragments for detection. Data processing was performed with MultiQuant software integrating primary and secondary product ion peaks.
- QTRAP 4500 system
- nanoLC Ultra with cHiPLC trap elute
- ChromXP C18 reversed phase chips
- Linear Accelerator trap technology
- MultiQuant software
Main Results and Discussion
The MRM3 workflow delivered superior selectivity compared to conventional MRM. For the CYP3A5 peptides DTINFLSK and SLGPVGFMK, two of three MRM transitions exhibited significant interferences at low femtomole levels, while MRM3 detection of secondary product ions produced clean, interference free signals. Calibration curves spanning 1.2 fmol to 805 fmol demonstrated excellent linearity, low coefficients of variance and high accuracy at the lower limit of quantitation of 1.2 fmol on column.
Benefits and Practical Applications
- Enhanced specificity for targeted protein isoform quantification in complex proteomic matrices
- Lower limits of quantitation through monitoring of secondary fragmentation products
- Robust performance in the presence of background interferences
- Valuable for drug metabolism and pharmacokinetic studies in clinical and research settings
Future Trends and Opportunities
MRM3 workflows are poised to expand for multiplexed quantification of diverse protein isoforms and post‐translational modifications. Advances in ion trap sensitivity and scan speed will further lower detection limits. Automated assay design and data analysis will facilitate high‐throughput targeted proteomic screening in pharmaceutical, clinical, and industrial laboratories.
Conclusion
The combination of quadrupole isolation and high sensitivity linear ion trap scanning on the QTRAP 4500 enables an MRM3 workflow that surpasses conventional MRM in selectivity and sensitivity. This approach is well suited for accurate quantification of low abundance protein isoforms in complex samples and advances targeted proteomic applications in drug metabolism research.
References
- MRM3 Quantitation for Highest Specificity in Complex Matrices SCIEX Technical Note RUO-MKT-02-2739-A
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Overview of LC-MS Quantitative Solutions for Biotherapeutic Analysis
2019|SCIEX|Others
Overview of LC-MS Quantitative Solutions for Biotherapeutic Analysis Featuring SCIEX Instrumentation and Technologies Lei Xiong, Elliott Jones SCIEX, Redwood City, California, USA Figure 1. Summary of SCIEX Quantitative Solutions and Instrumentations to Solve Analytical Challenges in Biotherapeutic Industry. Darkest green…
Key words
sciex, sciexmrmhr, mrmhrworkflow, workflowprotein, proteinmicrolc, microlcquantification, quantificationions, ionsmrm, mrmqtrap, qtrapswath, swathquantitation, quantitationhrams, hramsspectrometers, spectrometersmobility, mobilityselexion
Multiple Mass Spectrometric Strategies for High Selectivity Quantification of Proteins and Peptides
2015|SCIEX|Applications
Multiple Mass Spectrometric Strategies for High Selectivity Quantification of Proteins and Peptides Solve the Most Challenging Quantitative Problems with QTRAP® 5500 System Sahana Mollah, Christie Hunter SCIEX, USA There has been an exponential increase in the number of ‘potential’ protein…
Key words
bnp, bnpmrm, mrmqtrap, qtrapdms, dmsdifferential, differentialmobility, mobilitysim, simselectivity, selectivityselexion, selexionpeptide, peptidestrategies, strategiesquantitation, quantitationmonitoring, monitoringquantitative, quantitativebased
The LC-MS/MS workhorse - SCIEX 4500 Series mass spectrometers
2019|SCIEX|Brochures and specifications
The LC-MS/MS workhorse SCIEX 4500 Series mass spectrometers The Power of Precision The SCIEX 4500 Series of mass spectrometers takes the legendary API 4000™ LC-MS/MS system and intelligently re-engineers it to set a new benchmark for reliable quantitation. It offers…
Key words
your, yourlinac, linacsciex, sciexqtrap, qtrapacqurate, acquratequrved, qurvedion, ionlab, labpulse, pulsefast, fastmore, morecounting, countingcontainment, containmentqjet, qjetaccompany
Škola MS: SCIEX QTRAP
2023|SCIEX|Presentations
SCIEX QTRAP František Laštovička - AMEDIS, 11.9.2023 Škola MS 2023, Milovy 1 © 2020 DH Tech. Dev. Pte. Ltd. SCIEX mass spectrometry portfolio: 2023 SCIEX Triple Quad and QTRAP systems Triple Quad 3500 system QTRAP/TQ 4500 system SCIEX 5500+ system…
Key words
ems, emsqtrap, qtrapepi, epimrm, mrmscreening, screeningscan, scanunknown, unknownlibrary, libraryconfirmation, confirmationsciex, sciexenhanced, enhancedgeneral, generaltarget, targettrap, trapxic