Polystyrene analysis using MALDI-TOF MS
Applications | 2020 | BrukerInstrumentation
Matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) offers rapid and detailed polymer analysis by combining molecular weight distribution measurement with chemical composition insights. This capability is critical for quality control, safety compliance, and innovation in polymer manufacturing and research.
This work evaluates the performance of an autoflex maX MALDI-TOF MS system coupled with PolyTools software to characterize three polystyrene standards over a mass range from 2 kDa to 200 kDa. The primary goals were to:
The analytical workflow comprised:
The combination of a high-resolution MALDI-TOF MS system and PolyTools software delivers comprehensive polystyrene characterization across a broad mass range. The methodology provides accurate molecular weight metrics, monomer verification, and end-group identification, supporting both research and industrial quality control.
MALDI, LC/TOF, LC/MS, LC/MS/MS
IndustriesMaterials Testing
ManufacturerBruker
Summary
Significance of the Topic
Matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) offers rapid and detailed polymer analysis by combining molecular weight distribution measurement with chemical composition insights. This capability is critical for quality control, safety compliance, and innovation in polymer manufacturing and research.
Objectives and Study Overview
This work evaluates the performance of an autoflex maX MALDI-TOF MS system coupled with PolyTools software to characterize three polystyrene standards over a mass range from 2 kDa to 200 kDa. The primary goals were to:
- Determine average molecular weights (Mn, Mw) and dispersity (ĐM).
- Identify the styrene monomer unit and sample end-group chemistries.
- Assess spectral acquisition in both reflector and linear modes.
Applied Methodology and Instrumentation
The analytical workflow comprised:
- Sample Preparation
- Dissolution of polystyrene standards in THF (10 g/L).
- Use of DCTB matrix (20 g/L) and silver trifluoroacetate dopant (0.1 M).
- Mixing ratios optimized for each sample mass range and manual spotting (0.5 μL) onto a steel target plate.
- Data Acquisition and Processing
- Acquisition on a Bruker autoflex maX: reflector mode for low-mass and linear mode for higher-mass samples.
- Positive-ion mode calibration on adjacent standards using COMPASS for flex 1.4.
- Automated peak detection, monomer and end-group assignment, and molecular weight calculations performed in PolyTools 1.0 SR 1.
Main Results and Discussion
- PS1 (∼2 kDa):
- Reflector mode spectra resolved oligomers from 1 to 3.3 kDa with 104.0624 Da repeat units.
- Isotopic patterns confirmed silver cationization.
- PolyTools results: Mn = 2137.7 Da, Mw = 2212.3 Da, ĐM = 1.035, DP ≈ 20.5; end groups: hydrogen and butyl.
- PS2 (∼19 kDa):
- Linear mode provided clear oligomer separation despite loss of isotopic detail.
- Average mass spacing of 104.19 Da matched styrene units; residual mass error < 0.3 Da.
- Monomer and end-group assignments aligned with expected composition.
- PS3 (∼210 kDa):
- High-mass range prevented individual oligomer resolution, yet PolyTools accurately computed average molecular weights.
- Detection of multiply charged species and polymer clusters illustrated complex ion formation at high masses.
Benefits and Practical Applications
- Unified determination of molecular weight distribution and chemical structure expedites polymer profiling.
- Automated software reduces manual interpretation errors and speeds up analysis.
- Applicable to quality assurance, batch comparison, copolymer studies, and impurity screening.
Future Trends and Potential Applications
- Integration of MALDI-TOF/TOF (MS/MS) for detailed end-group and sequence analysis.
- Coupling with chromatographic separations for advanced copolymer characterization.
- Development of cloud-based spectral libraries and real-time data processing tools.
- Standardization of protocols for regulatory and industrial acceptance.
Conclusion
The combination of a high-resolution MALDI-TOF MS system and PolyTools software delivers comprehensive polystyrene characterization across a broad mass range. The methodology provides accurate molecular weight metrics, monomer verification, and end-group identification, supporting both research and industrial quality control.
References
- ISO 10927:2018 – Determination of molecular mass and distribution of polymer species by MALDI-TOF MS.
- Knop et al. (2010). Systematic MALDI-TOF CID Investigation on Different Substituted mPEG 2000. Macromol. Chem. Phys. 211, 677–684.
- Weidner et al. (2007). Copolymer characterization by LC-MALDI-TOF MS. Rapid Commun. Mass Spectrom. 21, 2750–2758.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
MALDI-CID Study of Poly(methyl methacrylate)
2019|Bruker|Applications
MALDI-CID Study of Poly(methyl methacrylate) Matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) was chosen for an in-detail analysis of poly(methyl methacrylate) (PMMA) in order to determine the possible fragmentation mechanism with the help of collision-induced dissociation. Every parent peak…
Key words
region, regionmaldi, maldipmma, pmmascission, scissionfragmentation, fragmentationmolar, molarpoly, polyparent, parenttrimer, trimertof, tofdimer, dimerexplained, explainedmethacrylate, methacrylatemonomer, monomerdistribution
Surface Layer MALDI-MSI of Synthetic Materials on a Molecular Level
2020|Bruker|Applications
Surface Layer MALDI-MSI of Synthetic Materials on a Molecular Level A surface-layer matrix-assisted laser desorption ionization mass spectrometry imaging technique (SL-MALDI-MSI) is developed to study the chemical composition of multicomponent polymer surfaces. Abstract Using this technique, four different types of…
Key words
pmma, pmmamaldi, maldisurface, surfacecationizing, cationizingdefects, defectsfilm, filmmsi, msisublimed, sublimedtof, tofnatfa, natfaultraflex, ultraflexlaser, laserimages, imagesbruker, brukercasting
The enhanced ultrafleXtreme - Industry Leading Performance for Polymer and Chemical Research
2020|Bruker|Brochures and specifications
The enhanced ultrafle treme Industry Leading Performance for Polymer and Chemical Research Innovation with Integrity MALDI-TOF/TOF MS Simple, Straightforward, and Specific Rapid Determination of Chemical Features of Polymers Determination of: • Sum of end groups including single end group analysis…
Key words
maldi, malditof, toftlc, tlcbruker, brukerpolymer, polymermpeg, mpegpolytools, polytoolsimaging, imagingindustry, industrylaser, laserionization, ionizationdeeper, deeperglycerol, glyceroldistributions, distributionscharacterization
Automated High-Throughput Oligo QC: From Automated Sample Preparation to MALDI Spectra Assessment
2020|Bruker|Applications
Automated High-Throughput Oligo QC: From Automated Sample Preparation to MALDI Spectra Assessment Synthetic oligonucleotides find a variety of applications in molecular biology. Primers for instance are used for DNA sequencing and amplification. Mass spectrometry is a well-established technology for quality…
Key words
maldi, maldicta, ctagac, gacprimers, primersanchorchip, anchorchipspectra, spectramass, masscgg, cggpreparation, preparationoligo, oligoagg, agggct, gctcag, cagtaa, taabruker