LCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Around the LC System in 45 Slides

Presentations | 2016 | Agilent TechnologiesInstrumentation
Consumables, HPLC, LC columns
Industries
Manufacturer
Agilent Technologies

Summary

Importance of the Topic


A smooth and reliable liquid chromatography workflow is essential in analytical chemistry. Proper instrument maintenance, sample preparation, and column selection directly impact data quality, reproducibility, and operating costs. By anticipating common sources of pressure instabilities and peak distortions, laboratories can maximize uptime, extend column lifetime, and improve analytical sensitivity for complex matrices.

Objectives and Overview


  • Identify common causes of LC pressure and retention problems.
  • Outline preventive maintenance and best practices for each LC module.
  • Present sample preparation strategies to reduce matrix effects.
  • Demonstrate column and connection technologies to ensure robust chromatography.

Used Instrumentation


  • Agilent 1200/1290 Series LC with quaternary pump, degasser, autosampler, and UV/DAD detector.
  • Columns: StableBond SB-C18/C8, Eclipse Plus, Poroshell 120 EC-C18/SB-C18.
  • Inline filters (0.2–0.3 μm) and Fast Guards (600–1200 bar).
  • Sample prep tools: Captiva ND/Lipids filtration plates and cartridges, Bond Elut SPE cartridges, QuEChERS/dSPE EMR-Lipid kits.

Methodology and Instrumentation


Key maintenance and cleanup steps address each module:


  • Degasser: flush channels with water and organic solvent; avoid long exposure to aqueous buffers to prevent microbial growth and bubbles.
  • Pump: schedule replacement of inlet/outlet valves, pistons, seals, purge-valve frits; perform seal wear-in procedures to maintain stable pressure and noise-free baselines.
  • Autosampler: maintain needle, rotor seal, metering seals, and needle seat; implement needle-wash routines to prevent carryover.
  • Detector: replace lamps and clean flow cells; ensure flow cells contain organic content to inhibit crystallization or microbial growth.
  • Connections: use InfinityLab Quick Connect and Quick Turn fittings to avoid dead volume and ensure reproducible, high-pressure connections.
  • Sample Preparation: employ filtration to remove particulates; Captiva ND and ND Lipids for protein and lipid removal; SPE (Bond Elut) for targeted cleanup; QuEChERS with EMR-Lipid for rapid pesticide and matrix removal.
  • Column Selection: choose pH-stable phases appropriate for anticipated mobile phase pH; monitor secondary interactions and silica degradation at extreme pH.

Main Results and Discussion


  • Routine replacement of pump and autosampler components stabilized system pressure and reduced retention-time variability.
  • Filtration and targeted sorbent cleanup prevented column inlet blockages, leading to extended column performance even with protein- or lipid-rich samples.
  • EMR-Lipid in QuEChERS workflows achieved cleaner extracts, enhanced signal-to-noise, and protected both LC columns and MS sources.
  • Poroshell columns with robust frits demonstrated sustained efficiency over hundreds of injections of precipitated plasma without centrifugation.
  • pH control and selection of stable bonded phases minimized silica dissolution and secondary silanol interactions, preserving peak shape and retention time.

Benefits and Practical Applications


  • Minimized instrument downtime and service costs through preventive maintenance and enhanced sample cleanup.
  • Improved method sensitivity and reproducibility by reducing contamination, carryover, and matrix effects.
  • Extended column lifetime and consistent chromatography via proper phase selection and connection hardware.
  • Flexible workflows adaptable to pharmaceutical QC, environmental monitoring, food safety, and clinical analyses.

Future Trends and Opportunities


Advancements in automated and high-throughput sample preparation, novel sorbent materials for selective interference removal, and development of high-pH stable column chemistries will drive method innovation. Integration of real-time system monitoring and smart maintenance scheduling can further streamline workflows and ensure analytical robustness.

Conclusion


A proactive strategy combining scheduled maintenance, optimized sample preparation, and informed column and connection choices is key to achieving reliable, high-performance LC analyses. Implementing these best practices reduces troubleshooting time, lowers operational costs, and enhances data quality across a broad range of analytical applications.

References


No external literature references were provided in the source material.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
The Chromatography Checklist
The Chromatography Checklist
2018|Agilent Technologies|Presentations
The Chromatography Checklist Rita Steed Alexander Ucci May 23, 2018 1 May 22, 2018 Agilent Restricted Checklist ❑ Is your sample ready? ✓ Solubility: sample solvent and starting mobile phase conditions ✓ Filters ✓ Sample Clean-Up ❑ Supplies ✓ In-line…
Key words
restricted, restrictedagilent, agilentcolumn, columndwell, dwellyour, yourfilters, filtersseal, seallipid, lipidschedule, scheduleless, lesschecklist, checklistguards, guardsrole, rolevolume, volumephase
A Gram of Prevention - Simple Tips for Maintaining LC Instrument Performance
A Gram of Prevention Simple Tips for Maintaining LC Instrument Performance Mark Powell Applications Engineer January 21, 2014 Confidentiality Label January 21, 2014 1 Overview • Know the flow path of your instrument (1260/1200/1100) • What is in your solvent…
Key words
confidentiality, confidentialitylabel, labelcare, carepump, pumpvalve, valvefittings, fittingsyour, yourdetector, detectordegasser, degassermau, mauautosampler, autosamplerseal, sealbottle, bottlelamp, lampsolvent
Surviving Chromatography - Part 1:   Preventive Action
Surviving Chromatography - Part 1: Preventive Action
2019|Agilent Technologies|Presentations
Surviving Chromatography Part 1: Preventive Action Mark Powell Alexander Ucci 1 September 10, 2018 Agilent restricted Planning Ahead • Planning and good habits can help prevent problems that can take you down the wrong path and cause you to lose…
Key words
restricted, restrictedagilent, agilentporoshell, poroshellinfinitylab, infinitylabvendor, vendorseal, sealrotor, rotorlamps, lampscolumn, columnother, otherlipid, lipidemr, emrfilters, filtersseals, sealsless
Why Did That Happen?
Why Did That Happen?
2021|Agilent Technologies|Presentations
“"Why Did That Happen?"” Golnar Javadi Applications Engineer LC Columns and Consumables Technical Support March 31, 2021 "Why Did That Happen?" Common chromatography problems Changes in System Pressure • High pressure • Low pressure • Pressure fluctuations 1 Changes in…
Key words
happen, happendid, didwhy, whyporoshell, poroshellfilter, filteryes, yeselut, elutpressure, pressureinfinitylab, infinitylabchanges, changesspe, spesolvent, solventphase, phaseinline, inlinecaptiva
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike