LCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

HPLC Separation Robustness and Ruggedness - Assessing the Effects of Experimental Variables During Method Development

Presentations | 2015 | Agilent TechnologiesInstrumentation
HPLC
Industries
Manufacturer
Agilent Technologies

Summary

Importance of the Topic


Analytical methods in HPLC must be reliable under small variations in conditions to ensure routine performance in quality control, method transfer and regulatory compliance. Robust and rugged methods reduce rework, save resources and support reproducible separations across laboratories.

Objectives and Study Overview


This study examines the influence of deliberate changes in chromatographic parameters on method performance during development. It aims to define robustness (resistance to small operational variations) and ruggedness (reproducibility across labs/operators) and to provide practical guidelines for assessing these attributes in HPLC separations.

Methodology and Instrumentation


The methodology involves systematically varying parameters such as column type and lot, mobile phase composition (buffer pH, concentration, organic modifier), sample conditions (injection volume, solvent strength), instrument settings (temperature, detector cell volume), and gradient factors (dwell volume, gradient steepness). Resolution (Rs), selectivity (α), peak shape and efficiency were monitored to quantify effects.

Instrumentation Used


  • Agilent 1290 Infinity HPLC system
  • Columns: ZORBAX StableBond C18, Poroshell 120 EC-C18, Eclipse XDB-C8
  • UV detector (e.g., 254 nm, 250 nm)
  • Temperature-controlled column compartment
  • Standard solvent filtration and pH meter

Main Results and Discussion


  • Column Stability: StableBond and Poroshell columns demonstrated extended lifetimes under acidic and basic conditions at elevated temperature, with minimal loss in efficiency.
  • Column Lot Reproducibility: Testing three lots revealed minor retention and selectivity variations; slight adjustments in organic content improved lot-to-lot consistency.
  • Buffer Effects: Buffered mobile phases enhanced retention and peak symmetry for ionizable analytes; pH shifts as small as 0.1 units significantly altered resolution for acidic and basic compounds.
  • Sample Conditions: Injection volumes and solvent strength influenced peak shape and Rs; optimal volumes minimized band broadening.
  • Temperature Control: Variations of ±5 °C altered selectivity and Rs; stable temperature regulation is critical.
  • Detector Cell Volume: Larger flow cells increased band broadening, reducing plate count and resolution.
  • Gradient Factors: Dwell volume differences between instruments affected retention profiles; compensation via initial isocratic hold or injection delay and specifying dwell volume in methods improves transferability. Gradient steepness changes within ±20% affected resolution for late-eluting analytes.

Benefits and Practical Applications


Robust and rugged HPLC methods ensure reproducible performance in routine testing and across laboratories, facilitate regulatory compliance (ICH, FDA, USP), and minimize downtime due to method failures. Incorporating robustness studies into method development streamlines validation and transfer, reducing cost and resource expenditure.

Future Trends and Potential Applications


Advances in column technology (e.g., superficially porous particles for high-pH stability), improved software for dwell volume and gradient simulations, and automated robustness screening are expected. Integration of quality-by-design principles and chemometric tools will further optimize method resilience and expedite method transfer in regulated environments.

Conclusion


Systematic assessment of critical HPLC parameters during method development builds resilience against operational variability, ensuring accurate and precise separations. Careful selection of columns, buffers and instrument settings, combined with clear method documentation, underpins robust and rugged methods that meet regulatory and practical demands.

Reference


  • Kirkland JJ, Henderson JW. Journal of Chromatographic Science. 1994;32:473–480.
  • Snyder LR, Kirkland JJ, Glajch JL. Practical HPLC Method Development. 2nd ed.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Conquer Method Variability - Evaluate Variables During Method Development
Conquer Method Variability Evaluate Variables During Method Development Rita Steed LC Columns Application Engineer December 11, 2018 1 December 11, 2018 Conquer Method Variability Agilent Restricted Common Separation Goals and Method Performance Criteria Good System Suitability Parameters • • •…
Key words
conquer, conquerrestricted, restrictedvariability, variabilityagilent, agilentmethod, methodruggedness, ruggednessdwell, dwellvariables, variablesgradient, gradientvolume, volumeresolution, resolutionbuffer, buffercolumn, columnsteepness, steepnessrobustness
Getting Off to a Good Start - The makings of a reliable method
Getting Off to a Good Start The makings of a reliable method Rita Steed LC Columns Application Engineer January 16, 2020 The Importance of Developing a Reliable Method* Scenario: You’re trying to reproduce an experiment in a journal article but…
Key words
getting, gettingrestricted, restrictedoff, offstart, startgood, goodporoshell, poroshellagilent, agilentinfinitylab, infinitylabphase, phasemobile, mobilevolume, volumeime, imeruggedness, ruggednessmau, maudwell
Eliminating Unwanted Variability in LC Methods: More Uptime, More Sample Throughput
Eliminating Unwanted Variability in LC Methods: More Uptime, More Sample Throughput HPLC Separation Robustness and Ruggedness Group/Presentation Title Agilent Restricted June 20, 2017Month ##, 200X Page 1 Different Compounds Interact Differently with the Column and Mobile Phase Changes in Gradient…
Key words
gagilent, gagilentcountry, countrypermitted, permittedrestricted, restrictedcannabis, cannabisintended, intendedstate, statelaboratories, laboratoriessafety, safetytesting, testingwhere, whereunder, underagilent, agilentsolutions, solutionsquality
New To HPLC - Avoiding Beginner Pitfalls
New To HPLC - Avoiding Beginner Pitfalls
2013|Agilent Technologies|Presentations
New To HPLC Avoiding Beginner Pitfalls Rita Steed LC Columns Application Engineer April 24, 2013 Outline  Instrument • Connections • Performance • Detector  Column • Characteristics • Lifetime  Method Conditions • Mobile phase • pH • Temperature…
Key words
beginner, beginnerpitfalls, pitfallsavoiding, avoidingrestricted, restrictedagilent, agilentcolumn, columnmobile, mobilemin, minphase, phasemau, maufittings, fittingsvolume, volumepitfall, pitfallyour, yourred
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike