Achieving Ultra-High-Speed Analysis with LC/MS/MS
Technical notes | 2012 | ShimadzuInstrumentation
The combination of ultra-high-pressure liquid chromatography and tandem mass spectrometry addresses the growing demand for rapid, high-resolution analysis in fields such as pharmaceutical development, environmental monitoring, and food safety. By pushing analysis times to fractions of those achieved with conventional LC/MS/MS setups while maintaining or improving separation performance and sensitivity, laboratories can increase throughput, reduce solvent consumption, and enhance data reliability.
This technical report examines the limitations of typical LC/MS/MS instruments when applying sub-2 µm particle columns for ultra-high-speed separations. It identifies key challenges—including separation loss, reduced data point acquisition, and increased cross talk—and introduces a next-generation system designed to overcome these barriers. The core goal is to demonstrate how Shimadzu’s LCMS-8030 triple quadrupole mass spectrometer paired with Nexera UHPLC achieves unprecedented analysis speeds without compromising analytical performance.
The study contrasts conventional LC/MS/MS setups with an optimized configuration featuring:
1. Separation Performance
The report demonstrates that, while sub-2 µm columns theoretically sustain efficiency at higher flow rates, conventional LC flow paths introduce dispersion that degrades resolution. The optimized UHPLC system preserves sharp peaks by minimizing extra-column volume.
2. Data Point Density
Conventional MS/MS duty cycles impose long dwell, pause, and polarity-switching times, yielding as few as 4–5 points per peak and distorted chromatograms. The LCMS-8030 reduces pause time to 1 ms and dwell time to 3 ms, ensuring over 10 points per peak for smooth, reproducible peak shapes.
3. Cross Talk
In standard collision cells, ion slowdown causes residual product ions to enter subsequent measurements, distorting results. The UFsweeper technology efficiently removes product ions, dramatically lowering cross talk and improving quantitative accuracy.
By integrating these enhancements, laboratories achieve:
Continued advancements in UHPLC and mass spectrometry are expected to focus on further reducing cycle times and improving multiplexed analyses. Emerging areas include real-time reaction monitoring, ultra-high-throughput screening for biomarker discovery, and on-site environmental contaminants testing. Integration with advanced data processing algorithms and machine-learning-based method optimization will further streamline workflows.
The deployment of the LCMS-8030 with Nexera UHPLC represents a significant leap in ultra-high-speed LC/MS/MS analysis. By addressing separation deterioration, data acquisition limitations, and cross talk, the system enables rapid, reliable quantitative and qualitative measurements. Its modular design also allows straightforward upgrading of existing LC platforms to achieve high-throughput performance.
LC/MS, LC/MS/MS, LC/QQQ
IndustriesManufacturerShimadzu
Summary
Significance of the Topic
The combination of ultra-high-pressure liquid chromatography and tandem mass spectrometry addresses the growing demand for rapid, high-resolution analysis in fields such as pharmaceutical development, environmental monitoring, and food safety. By pushing analysis times to fractions of those achieved with conventional LC/MS/MS setups while maintaining or improving separation performance and sensitivity, laboratories can increase throughput, reduce solvent consumption, and enhance data reliability.
Objectives and Overview of the Study
This technical report examines the limitations of typical LC/MS/MS instruments when applying sub-2 µm particle columns for ultra-high-speed separations. It identifies key challenges—including separation loss, reduced data point acquisition, and increased cross talk—and introduces a next-generation system designed to overcome these barriers. The core goal is to demonstrate how Shimadzu’s LCMS-8030 triple quadrupole mass spectrometer paired with Nexera UHPLC achieves unprecedented analysis speeds without compromising analytical performance.
Methodology and Instrumentation Used
The study contrasts conventional LC/MS/MS setups with an optimized configuration featuring:
- UHPLC pump capable of 0.0001–5 mL/min at up to 130 MPa with a 20 mL high-efficiency mixer for accurate ultra-fast gradients
- SIL-30AC autosampler with 10-second injection cycles and minimal carryover
- Shim-pack XR series sub-2 µm columns for rapid high-efficiency separations
- LCMS-8030 equipped with UFsweeper collision cell technology to accelerate product ions, reducing cross talk
- Polarity switching module achieving positive/negative transitions in 15 ms
- Scan engine delivering up to 15 000 u/sec with dynamic voltage optimization to preserve sensitivity across a broad mass range
Main Results and Discussion
1. Separation Performance
The report demonstrates that, while sub-2 µm columns theoretically sustain efficiency at higher flow rates, conventional LC flow paths introduce dispersion that degrades resolution. The optimized UHPLC system preserves sharp peaks by minimizing extra-column volume.
2. Data Point Density
Conventional MS/MS duty cycles impose long dwell, pause, and polarity-switching times, yielding as few as 4–5 points per peak and distorted chromatograms. The LCMS-8030 reduces pause time to 1 ms and dwell time to 3 ms, ensuring over 10 points per peak for smooth, reproducible peak shapes.
3. Cross Talk
In standard collision cells, ion slowdown causes residual product ions to enter subsequent measurements, distorting results. The UFsweeper technology efficiently removes product ions, dramatically lowering cross talk and improving quantitative accuracy.
Benefits and Practical Applications
By integrating these enhancements, laboratories achieve:
- Analysis time reductions of ten-fold or more compared to conventional LC/MS/MS
- Enhanced peak resolution and reproducibility even under rapid gradients
- Increased sensitivity and accuracy through minimized cross talk and optimized scan conditions
- Greater throughput in high-volume environments such as pharmacokinetic studies, clinical assays, and quality control workflows
Future Trends and Possibilities of Use
Continued advancements in UHPLC and mass spectrometry are expected to focus on further reducing cycle times and improving multiplexed analyses. Emerging areas include real-time reaction monitoring, ultra-high-throughput screening for biomarker discovery, and on-site environmental contaminants testing. Integration with advanced data processing algorithms and machine-learning-based method optimization will further streamline workflows.
Conclusion
The deployment of the LCMS-8030 with Nexera UHPLC represents a significant leap in ultra-high-speed LC/MS/MS analysis. By addressing separation deterioration, data acquisition limitations, and cross talk, the system enables rapid, reliable quantitative and qualitative measurements. Its modular design also allows straightforward upgrading of existing LC platforms to achieve high-throughput performance.
Used Instrumentation
- Shimadzu Nexera UHPLC with high-pressure pump and high-efficiency mixer
- SIL-30AC autosampler
- Shim-pack XR sub-2 µm columns (ODS, C8, Phenyl)
- Shimadzu LCMS-8030 triple quadrupole mass spectrometer with UFsweeper collision cell and rapid polarity switching
References
- Prominence UFLC Next Generation High-Throughput HPLC Technical Report
- Migration Points to Fast Analysis; Optimization of the System and Analysis Conditions Technical Report
- HPLC Techniques to Improve LCMS Performance and Productivity Technical Report
- LCMS-2020 - Reduces Limitations of HPLC Analysis and UFLC Analysis Technical Report
- Reduced Solvent Consumption and Operational Costs with Ultra High Speed Analysis Technical Report
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Shimadzu Liquid Chromatograph Mass Spectrometer LCMS-8060
2019|Shimadzu|Brochures and specifications
C146-E286F Liquid Chromatograph Mass Spectrometer LCMS-8060 CHANGES EVERYTHING Sensitivity Speed Solutions LCMS-8060 The LCMS-8060 pushes the limits of what we can see. Creating the future of LC/MS/MS analysis can mean making something work with higher sensitivity, go faster or cost…
Key words
mrm, mrmlabsolutions, labsolutionsmin, minmediators, mediatorsscan, scanchromatograph, chromatographsurvey, surveymass, massspectrometer, spectrometerblank, blankspiked, spikedpause, pauseinsight, insightivit, ivitultra
Shimadzu LC World Talk 2011 Issue 1
2011|Shimadzu|Others
C190-E149 2011 Issue 1 LC World Talk SHIMADZU'S NEWSLETTER FOR THE HPLC GLOBAL COMMUNITY Shimadzu Introduces the LCMS-8030, the World’s Fastest Triple Quadrupole Mass Spectrometer n n LCMS-8030 Ultra Fast Tandem Quadrupole Mass Spectrometer, The Next Generation of Mass Spectrometer…
Key words
melamine, melamineimpurities, impuritiesmicrodialysis, microdialysispineal, pinealultra, ultramelatonin, melatoninfast, fastmobile, mobileufsweeper, ufsweepercyanuric, cyanuricdimension, dimensioncolumn, columnmrm, mrmanalysis, analysisjetlag
Shimadzu Liquid Chromatograph Mass Spectrometer LCMS-8050
2019|Shimadzu|Brochures and specifications
C146-E232F Liquid Chromatograph Mass Spectrometer LCMS-8050 Speed and Sensitivity Beyond Comparison Continuing the evolution of Shimadzu's UF technology, Shimadzu introduces the LCMS-8050 triple quadrupole mass spectrometer, offering unparalleled measurement speeds and high-sensitivity performance. High-sensitivity quantitation delivered at high speed Multi-component…
Key words
ufswitching, ufswitchingswitching, switchingmrm, mrmufsensitivity, ufsensitivityufscanning, ufscanningspeed, speedhigh, highlabsolutions, labsolutionsufsweeper, ufsweepermpp, mppinsight, insightnexera, nexerapolarity, polarityeven, evenchromatograph
Evaluation of MS Scanning Speeds with UHPLC Peak Widths
|Shimadzu|Presentations
Evaluation of MS Scanning Speeds with UHPLC Peak Widths William Hedgepeth, Masatoshi Takahashi, Shimadzu Scientific Instruments Introduction The trend toward high-speed LC analysis with reduced column particle sizes continues to increase in popularity. The narrow peak widths of high-speed LC…
Key words
speed, speedscan, scandaimuron, daimuroncarpropamid, carpropamidnegative, negativepositive, positivescanning, scanningpeak, peakneg, negpos, posion, iontrycropyr, trycropyrbentazone, bentazonehigh, highproton