High Capacity Electrostatic Ion Trap Mass Spectrometer and its Signal Processing

Posters |  | ShimadzuInstrumentation
LC/MS, LC/IT
Industries
Manufacturer
Shimadzu

Summary

Importance of the Topic


Understanding the behavior of ions in high-capacity electrostatic ion traps is fundamental for modern mass spectrometry, enabling improved dynamic range, sensitivity, and mass accuracy in trace analysis and complex sample characterization.

Objectives and Study Overview


This study aims to design a planar electrostatic ion trap (PEIT) with enhanced trapping capacity and energy focusing, to develop advanced signal processing methods using wavelet transform combined with FFT, and to demonstrate harmonic elimination via multiple image charge pickups.

Methodology and Instrumentation


The trap geometry consists of two concentric planar ring electrode arrays generating a rotationally symmetrical electrostatic field. Ions are pre-cooled in a linear ion guide and injected around the trap periphery. Field simulations and ion trajectory optimizations were performed using SIMION and AXSIM to achieve isochronous motion in radial, axial, and tangential directions. For signal processing:
  • Image charge signals are acquired from multiple radial pickup electrodes.
  • Wavelet transforms decompose mixed harmonic signals into scale coefficients corresponding to mass ranges.
  • Fourier transforms applied to selected scales yield simplified frequency spectra for precise mass assignment.
  • Linear combinations of signals from five pickups eliminate undesired harmonics, preserving the fundamental component.


Main Results and Discussion


Simulation results show baseline resolution of ions at m/z 609 and 609.12 within a 12 ms transient and a full width at half maximum (FWHM) resolution exceeding 15,000. Wavelet-based spectra exhibit reduced peak complexity compared to traditional FFT, facilitating subsequent deconvolution. Harmonic elimination via weighted pickup combination effectively suppresses 2nd to 5th harmonics, improving signal clarity without loss of mass accuracy.

Benefits and Practical Applications


  • Increased trapping capacity yields higher signal intensity and dynamic range.
  • Enhanced mass resolution enables differentiation of near-isobaric species in short acquisition times.
  • Modular pickup electrode design allows real-time harmonic filtering, beneficial for complex mixture analysis.
  • Potential integration into QA/QC, proteomics, and environmental monitoring workflows.


Future Trends and Applications


Advances may include real-time adaptive wavelet filtering, coupling PEIT with chromatographic separations, and leveraging machine learning for spectra deconvolution. Experimental validation and miniaturization of planar traps could extend their use in field-deployable mass spectrometers.

Conclusion


The proposed planar electrostatic ion trap design, combined with wavelet-FFT signal processing and multi-pickup harmonic elimination, demonstrates significant improvements in capacity, resolution, and spectral clarity in simulation. These developments promise robust performance for high-throughput and high-precision mass analysis.

References


  • D. Zajfman, Y. Rudich, et al. International Journal of Mass Spectrometry, 2003, 229, 55.
  • Golikov, Soloyev, Sudakov, Kumashiro. WO2009001909A21.
  • A. Verentchikov. A Path from Multi-reflecting ToF to Electrostatic Trap, 10th European FTMS Workshop, 2012.
  • E.N. Nikolaev, M.V. Gorshkov. International Journal of Mass Spectrometry and Ion Processes, 1985, 64, 115.
  • Q. Sun, C. Gu, L. Ding. Journal of Mass Spectrometry, 2011, 46, 417–424.
  • J.B. Greenwood, et al. Review of Scientific Instruments, 2011, 82, 043103.
  • L. Ding, R. Badheka. An Electrostatic Ion Trap with Planar Rotational Field Structure, 10th European FTMS Workshop, 2012.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
CDMS mode of FTMS Orbital Frequency Analyzer
CDMS mode of FTMS Orbital Frequency Analyzer Aleksandr Rusinov1; Li Ding2; Sergey Smirnov1; Patrick Knight1; Roch Andrzejewski1; Hiroaki Waki3 1Shimadzu Research Laboratory, Manchester, United Kingdom; 2Ningbo University, Ningbo, China; 3Shimadzu Corp., Kyoto, Japan CDMS OFA data processing 1. Orbital Frequency…
Key words
cdms, cdmsofa, ofacharge, chargeelimination, eliminationftms, ftmshistogram, histogramions, ionsmass, massfrequency, frequencyion, ionharmonics, harmonicsscoring, scoringhigh, highaldolase, aldolaseinaccuracy
Orbitrap Mass Spectrometry: Ultrahigh Resolution for Every Lab
Orbitrap Mass Spectrometry: Ultrahigh Resolution for Every Lab
2012|Thermo Fisher Scientific|Presentations
Symposium: New Alternatives in HighResolution Mass Spectrometry Orbitrap Mass Spectrometry: Ultrahigh Resolution for Every Lab Alexander Makarov March 14, 2012 What is Orbitrap™ analyzer? Orbitrap analyzer = = Orbital trapping + Image current detection + Electrodynamic squeezing + External pulsed…
Key words
orbitrap, orbitrapabundance, abundancerelative, relativehcd, hcdions, ionstrap, trapanalyzer, analyzerltq, ltqvelos, velospreamplifier, preamplifiermass, masshigh, hightic, ticmakarov, makarovfrequency
IMSC: Orbitrap Mass Spectrometry: from Dream to Mainstream
IMSC: Orbitrap Mass Spectrometry: from Dream to Mainstream
2009|Thermo Fisher Scientific|Presentations
Lunch talk at 18th IMSC Bremen Orbitrap Mass Spectrometry: from Dream to Mainstream Orbitrap is… …awarded… Dr. Alexander Makarov Thermo Fisher Scientific …produced… …commercialized… September 2, 2009 …developed… Irkutsk …in Bremen! Long time ago… A.Pekaln “If you really want to…
Key words
orbitrap, orbitrapions, ionsoatof, oatoftrap, trapion, ionmass, massoscillations, oscillationstof, toffrequency, frequencyvoltage, voltageltq, ltqfield, fieldresolving, resolvingicr, icrpartof
A GPU ACCELERATED CDMS TRAP TRAJECTORY SIMULATOR AND OPTIMISER
A GPU ACCELERATED CDMS TRAP TRAJECTORY SIMULATOR AND OPTIMISER David Langridge, Keith Richardson, Jeffery Brown, Kevin Giles Waters Corporation, Wilmslow, UK INTRODUCTION Charge detection mass spectrometry (CDMS) is an increasingly popular method for the analysis of large and heterogeneous ions.…
Key words
trajectory, trajectoryelit, elitoptimiser, optimisertube, tubevoltages, voltagestrap, trapcdms, cdmssimulator, simulatorgeometries, geometriesoptimisation, optimisationgpu, gpuelectrode, electroderesolution, resolutionion, iongeometry
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike