Laboratory gases and gas generators
Industries
Manufacturer
Peak Scientific

Summary

Význam tématu


Reliable, high-purity nitrogen generation directly in the laboratory eliminates dependence on gas cylinders and minimizes associated safety, storage and logistical challenges. On-demand nitrogen supply supports a wide range of applications in analytical chemistry, such as GC-MS, LC-MS, and inert atmospheres for sample preparation.

Cíle a přehled textu


This user manual provides a comprehensive guide to the Solaris XE nitrogen generator, covering unpacking, installation, safety guidelines, technical specifications, operation procedures and maintenance instructions. It is designed for end users who require a step-by-step reference for setup, operation and troubleshooting.

Použitá metodika a instrumentace


The Solaris XE uses pressure swing adsorption (PSA) technology to separate nitrogen from compressed air. Key components and requirements include:
  • Compressed air inlet: 5.5–10 bar, ISO 8573-1 Class 1.4.1 quality
  • Air consumption: up to 100 L/min at maximum nitrogen output
  • PSA adsorption module with activated carbon pre-filter
  • Output regulator and pressure gauge, adjustable up to 8 bar (116 psi)
  • Electrical supply: 100–230 VAC, 50/60 Hz, Class II insulation

Hlavní výsledky a diskuse


Performance characteristics and operational feedback demonstrate that Solaris XE:
  • Delivers up to 35 L/min of nitrogen at 95–99.5% purity
  • Achieves target purity within five minutes of start-up
  • Maintains stable output pressure under varying ambient conditions
  • Features color-coded LED indicators for system status and pressure alerts
  • Operates reliably in 5–35 °C ambient temperature and up to 2 000 m altitude

Přínosy a praktické využití metody


On-site PSA generation reduces downtime, eliminates cylinder handling risks and lowers operational costs. Applications include:
  • Mass spectrometry and chromatography support
  • Inert gas for sample storage and preparation
  • Quality control in pharmaceutical and food industries
  • Continuous operation in busy analytical laboratories

Budoucí trendy a možnosti využití


Future developments will focus on:
  • Remote monitoring and IoT connectivity for predictive maintenance
  • Energy-efficient compressor integration and heat recovery
  • Modular expansion for multi-gas generation systems
  • Enhanced automation interfaces for laboratory workflow software

Závěr


The Solaris XE nitrogen generator offers a compact, user-friendly and robust solution for continuous, high-purity nitrogen delivery. Its combination of performance, safety features and low maintenance makes it an ideal choice for modern analytical laboratories.

Reference


  • ISO 8573-1:2010 Compressed air purity classes
  • EN 61010-1:2010 Safety requirements for electrical equipment
  • EN 61326-1:2013 EMC requirements for laboratory equipment
  • FCC 47 CFR Part 15 Class B Unintentional radiator emissions limits
  • EU Directives 2014/35/EU (Low Voltage), 2014/30/EU (EMC), 2012/19/EU (WEEE), 2011/65/EU & 2015/863 (RoHS)

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Peak Solaris Air Compressor  - User Manual
Peak Solaris Air Compressor - User Manual
2019|Peak Scientific|Manuals
Solaris Air Compressor User Manual / Solaris Air Compressor User Manual Rev 5 RSID 1603 EN 07/11/19 Contents Change History How to use this Manual Introduction Warranties and Liabilities Safety Notices Symbols Safety Notice to Users Declaration of Conformity Environmental…
Key words
compressor, compressorair, airsolaris, solarispeak, peakcouttie, couttieliam, liamservice, servicewarning, warningcaution, cautiongenerator, generatorscientific, scientificprotected, protecteddeclaration, declarationpower, poweryour
Peak Solaris 10 - User Manual
Peak Solaris 10 - User Manual
2018|Peak Scientific|Manuals
Solaris 10 User Manual Solaris 10 User Manual Rev 2 RSID 1604 PN UM-0014 26/07/18 Contents Change History How to use this Manual Introduction Warranties and Liabilities Safety Notices Symbols Safety Notice to Users Declaration of Conformity Environmental Declaration Technical…
Key words
generator, generatorpeak, peakcaution, cautionwarning, warningprotected, protectedair, airunpacking, unpackingdeclaration, declarationyour, yourscientific, scientificminimum, minimumservice, servicemaximum, maximumany, anysupply
Genius XE (All Models) User Manual
Genius XE (All Models) User Manual
2020|Peak Scientific|Manuals
Genius XE (All Models) User Manual Copyright © Peak Scientific Instruments Ltd 2020 — GeniusXE User Manual Rev 9 RSID 2299 EN —18/02/2020 Register Product to Operate To begin operation you will need to register your generator. You can do…
Key words
screen, screengenerator, generatorservice, servicehome, homexxxx, xxxxwarning, warningcouttie, couttieliam, liamscreens, screensoverdue, overduepeak, peakpsi, psierror, erroreco, ecocaution
Peak Infinity XE 50X1 Series - User Manual
Peak Infinity XE 50X1 Series - User Manual
2019|Peak Scientific|Manuals
Infinity XE 50X1 Series User Manual Copyright © Peak Scientific Instruments Ltd 2019 — Infinity XE 50X1 Series User Manual Rev 2 RSID 2029 EN —27/11/2019 Contents Change History How to use this Manual Introduction Warranties and Liabilities Warranty &…
Key words
generator, generatorinlet, inletpeak, peakair, airwarning, warningpressure, pressureminimum, minimumrear, rearprotected, protectedelectrical, electricaloutlet, outletanalyser, analyserservice, servicedeclaration, declarationyour
Other projects
GCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike