Agilent 1290 Infinity II ELSD – Pre-Installation Site Preparation Checklist

Manuals | 2014 | Agilent TechnologiesInstrumentation
HPLC
Industries
Manufacturer
Agilent Technologies

Summary

Importance of the Topic


Proper site preparation for a sensitive analytical instrument like the Agilent 1290 Infinity II ELSD ensures accurate data, reliable operation, and extended instrument lifetime. By meeting specific space, environmental, power, and supply requirements, laboratories can prevent installation delays, reduce maintenance issues, and maintain safety standards.

Objectives and Overview


This document outlines the pre-installation requirements and checklist for the Agilent 1290 Infinity II ELSD (Evaporative Light Scattering Detector). It aims to guide customers in preparing their laboratory environment to facilitate a smooth installation and optimal instrument performance.

Methodology


The methodology involves evaluating the following key areas before instrument delivery:
  • Bench Space and Weight Capacity: Verify sufficient lateral and rear clearance (minimum 25 cm on each side, 15 cm at the rear) and confirm bench load capacity based on system configuration (cooled vs non-cooled modules).
  • Environmental Conditions: Ensure ambient temperature remains between 10–35 °C, with humidity levels of 10–80% non-condensing. Minimize temperature fluctuations (below 2 °C/hour) and avoid direct sunlight or drafts.
  • Power Supply: Provide compatible AC power (100–240 V, 50/60 Hz ±5%), with at least one dedicated outlet per instrument plus any computing system. Maximum consumption per unit is 150 W.
  • Operating Supplies: Prepare HPLC-grade solvents (water and acetonitrile), a caffeine standard, and restriction capillary tubing in recommended quantities for system setup and validation.
  • Exhaust and Gas Management: Install an extraction line (13 mm ID PVC or larger) to a fume hood for solvent vapor removal. Supply filtered nitrogen (98% purity) at 4.1–6.9 bar and 0.9–3.25 SLM for optimal detector operation.
  • Software and Drivers: Install OpenLAB CDS ChemStation or EZChrom (version A.01.05 or later) with ELSD driver 1.3 for data acquisition.

Used Instrumentation


  • Agilent 1290 Infinity II ELSD G7102A (cooled and non-cooled configurations)
  • OpenLAB CDS ChemStation and EZChrom Edition software with ELSD drivers

Key Considerations and Discussion


  • Space and Weight: Laboratory benches must support up to 13 kg per module and allow for adequate airflow and connectivity access.
  • Temperature Stability: Frequent or large temperature shifts can cause signal drift; controlled climate increases baseline stability and extends component life.
  • Solvent and Vapor Control: Proper exhaust routing and vapor sensors protect electronic components and ensure compliance with safety protocols.
  • Installation Readiness: Failure to meet requirements may lead Agilent to reschedule the installation, potentially delaying laboratory workflows.

Benefits and Practical Applications


Implementing these preparation steps:
  • Reduces installation time and costs by preventing last-minute site modifications
  • Ensures dependable ELSD performance for trace-level detection of non-UV-active compounds
  • Enhances laboratory safety by managing solvent vapors and electrical loads
  • Facilitates immediate user training and qualification upon instrument commissioning

Future Trends and Potential Applications


Advances in detector miniaturization and integrated diagnostics are expected to simplify installation requirements. Remote monitoring and automated environmental controls will further improve data stability and maintenance workflows. Expanded compatibility with software platforms and cloud-based data management may streamline instrument qualification and compliance.

Conclusion


A well-prepared laboratory environment is essential for the reliable operation of the Agilent 1290 Infinity II ELSD. By adhering to the specifications for space, environment, power, and supplies, users can maximize instrument uptime, data quality, and safety.

Reference


Agilent Technologies, Agilent 1290 Infinity II ELSD – G7102A Pre-Installation Site Preparation Checklist, July 2014.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Agilent InfinityLab LC Series - Site Preparation Checklist
Site Preparation Checklist Agilent InfinityLab LC Series Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site…
Key words
checklist, checklistsite, sitepreparation, preparationcabinet, cabinetcompartment, compartmentspecial, specialrequirements, requirementscondensing, condensinginstrument, instrumentsampler, samplerinfinitylab, infinitylabpump, pumpdescription, descriptionmodule, modulesolvent
Agilent InfinityLab LC Series - Site Preparation Checklist
Site Preparation Checklist Agilent InfinityLab LC Series Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site…
Key words
checklist, checklistsite, sitepreparation, preparationcabinet, cabinetcompartment, compartmentspecial, specialrequirements, requirementscondensing, condensinginstrument, instrumentsampler, samplerinfinitylab, infinitylabpump, pumpmodule, moduledescription, descriptionsolvent
Agilent 1290 Infinity LC/1260 Infinity LC  - Site Preparation Checklist
1290 Infinity LC/1260 Infinity LC Site Preparation Checklist 1290Preparation Site Infinity LC/1260 Checklist Infinity LC Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or…
Key words
checklist, checklistsite, sitetechnologies, technologiesagilent, agilentpreparation, preparationfrpsduwphqw, frpsduwphqwxwrvdpsohu, xwrvdpsohuhwhfwru, hwhfwruroxpq, roxpqhjdvvhu, hjdvvhurequirements, requirementsdescription, descriptionrrohu, rrohucustomer, customerconsumption
1220 Infinity/Infinity II LC - Site Preparation Checklist
1220 Infinity/Infinity II LC Site Preparation Checklist 1220Preparation Site Infinity/Infinity Checklist II LC Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of…
Key words
site, sitechecklist, checklistoperating, operatingtechnologies, technologiescustomer, customerweight, weightagilent, agilentpreparation, preparationambient, ambientconsumption, consumptionheight, heightinstallation, installationinfinity, infinityhumidity, humiditypower
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike