Performance Characteristics of the Agilent 1260 Infinity II Bio-Inert LC

Technical notes | 2017 | Agilent TechnologiesInstrumentation
HPLC
Industries
Manufacturer
Agilent Technologies

Summary

Importance of Topic


Biochromatographic analysis often requires resistance to harsh solvents, extreme pH, high salt concentrations, and efficient, reproducible performance. The inert flow path design minimizes sample contact with metal surfaces, reducing adsorption and degradation of biological molecules. Reliable retention time precision, carryover control, and compatibility with both conventional HPLC and UHPLC are essential for applications such as protein characterization, pharmaceutical impurity profiling, and QA/QC in complex matrices.

Goals and Study Overview


This technical overview evaluates the performance characteristics of the Agilent 1260 Infinity II Bio-inert LC system. Key objectives include comparing retention time precision under various gradient and isocratic conditions, assessing injection volume linearity and carryover, verifying detector linearity, and comparing bio-inert versus standard heat exchangers in the multicolumn thermostat for temperature-sensitive separations.

Methodology


Analytical tests were conducted using a variety of standard mixtures and real-world samples under conventional and UHPLC conditions. Retention time precision was measured for both fast (<1 min) and conventional gradients on 4.6 mm and 2.1 mm columns. Injection performance was evaluated over volumes from 0.5 to 100 µL, including area precision and carryover checks using chlorhexidine. Detector linearity studies employed caffeine standards covering four orders of magnitude. Temperature sensitivity experiments used sulfonamide drugs to compare standard and bio-inert heat exchangers.

Used Instrumentation


  • 1260 Infinity II Bio-inert Pump (G5654A)
  • 1260 Infinity II Bio-inert Multisampler with Cooler (G5668A)
  • 1260 Infinity II Multicolumn Thermostat with Bio-inert Heat Exchanger (G7116A)
  • 1260 Infinity II DAD Wide Wavelength Range WR with Bio-inert Flow Cell (G7115A)

Main Results and Discussion


  • Retention Time Precision: Conventional gradients showed RSD <0.04% (n=7), fast gradients <0.12% RSD. Isocratic runs yielded <0.06% RSD.
  • Pump Mixing: Step-gradient overlay demonstrated mixing performance equivalent to the previous Bio-inert Quaternary LC system.
  • Low-Pressure SEC: Bovine serum albumin separation at 21 bar achieved <0.05% RT RSD, highlighting stability for protein size-exclusion.
  • Injection Performance: Area RSD <0.14% for 5 µL injections; no carryover detected after 1,000 ng chlorhexidine, using optimized draw speeds and exterior needle wash.
  • Injection Linearity: Constant caffeine mass injection across 0.78–100 µL yielded <1.45% area RSD.
  • Detector Linearity: DAD WR linear across 1.5–1,600 ng caffeine with R²=0.99995 and response factor stability within ±5% over a broad absorbance range.
  • Thermostat Comparison: Bio-inert heat exchanger delivered temperature precision and retention order comparable to standard stainless steel exchanger for sulfa drugs across 10–60 °C.

Benefits and Practical Applications


The 1260 Infinity II Bio-inert LC offers robust performance for high-salt, extreme pH, and denaturing solvent conditions without sample adsorption. Its compatibility with standard HPLC methods and UHPLC yields higher throughput or resolution on sub-2 µm columns. Low carryover and precise injection support quantitative bioanalysis, while multi-column capacity with individual temperature control enhances method flexibility.

Future Trends and Opportunities


Advances in bio-inert flow paths will address emerging challenges in biopharmaceutical analysis, including intact protein quality control, host cell protein quantitation, and complex biotherapeutic mixtures. Integration with automated column management and advanced detection modalities (MS, FLD) will expand high-throughput screening and real-time monitoring. Continuous improvements in miniaturized UHPLC and inline sample preparation modules will further streamline bioanalytical workflows.

Conclusion


The Agilent 1260 Infinity II Bio-inert LC system delivers high retention time precision, low carryover, excellent detector linearity, and stable temperature control under demanding biochromatographic conditions. Its bio-inert design and flexible configuration make it well suited for diverse applications in life-science research, pharmaceutical development, and quality control.

References


  1. Schneider, S. Performance characteristics of the 1260 Infinity Bio-inert Quaternary LC System. Agilent Technologies Technical Overview, 5990-9194EN, 2011.
  2. Agilent 1260 Infinity Bio-inert Quaternary LC: Features, Technical Details, Applications, and Specifications. Agilent Technologies Data Sheet, 5990-6129EN, 2012.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Seamless Method Transfer from an Agilent 1260 Infinity Bio-inert LC to an Agilent 1260 Infinity II Bio-inert LC
Seamless Method Transfer from an Agilent 1260 Infinity Bio-inert LC to an Agilent 1260 Infinity II Bio-inert LC Charge Variant Analysis of Rituximab Innovator and Biosimilar Application Note Biologics & Biosimilars Author Abstract Sonja Schneider Monitoring the charge heterogeneity of…
Key words
biosimilar, biosimilarinnovator, innovatorbio, bioinert, inertrituximab, rituximabcharge, chargevariants, variantsadvisor, advisormau, mauvariant, variantmin, minbuffer, bufferretention, retentionsystem, systemtime
High-Performance Biopharma Analysis - Application compendium
Agilent InfinityLab Bio LC Solutions High-Performance Biopharma Analysis Application compendium 2 Overcome Analytical Challenges in Biopharma The growing complexity of biopharma analysis calls for innovation that delivers excellence across the biopharmaceutical workflow. The Agilent InfinityLab Bio LC Solutions comprise a…
Key words
retention, retentionmin, minbio, biogradient, gradienttime, timenistmab, nistmabresponse, responsemau, maupeptide, peptideagilent, agilentmapping, mappingcharge, chargersd, rsdmab, mabantibody
Reducing Cycle Time for Quantification of Human IgG Using the Agilent Bio-Monolith Protein A HPLC Column
Reducing Cycle Time for Quantification of Human IgG Using the Agilent Bio-Monolith Protein A HPLC Column Alternating Column Regeneration Using an Agilent 1200 Infinity Quick-Change Bio-inert 2-position/10-port Valve and the 1290 Infinity Flexible Cube Application Note Biotherapeutics & Biosimilars Authors…
Key words
bio, biomonolith, monolithigg, igginert, inertprotein, proteincube, cubepump, pumphplc, hplccolumn, columnhuman, humanflexible, flexibleautosampler, autosampleriggs, iggsaffinity, affinityinfinity
Charge Variant and Aggregation Analysis of Innovator and Biosimilars of Rituximab
Application Note Pharma & Biopharma Charge Variant and Aggregation Analysis of Innovator and Biosimilars of Rituximab Author Bian Yulan Global Solution Development Center, Global Laboratory Solution Sales (LSS) Marketing, Agilent Technologies Singapore (Sales) Pte. Ltd. Abstract Monoclonal antibodies are an…
Key words
innovator, innovatorvariants, variantsbiosimilars, biosimilarsrituximab, rituximabrsd, rsdhmws, hmwsarea, areabasic, basiclmws, lmwsmain, mainmau, maupeak, peakmab, mabacidic, acidiccharge
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike