The Role of End-Capping in RP

We, 9.7.2025
| Original article from: Phenomenex
New end-capping technologies can significantly improve the analysis of compounds that were once considered challenging.
<p><strong>Phenomenex: </strong>The Role of End-Capping in RP</p>

Phenomenex: The Role of End-Capping in RP

Silica used to pack HPLC columns is a polymer of silicon and oxygen with Si-OH groups (silanols) at the surface. To these silanols, functional groups are bonded using a silane reagent to functionalize the base media.

When HPLC columns are functionalized, it is impossible to ensure 100% phase coverage and react all available silanol groups (free hydroxyls on the surface of the silica) with the bonded phase. Sterically, as you begin to bond a phase you increase the hindrance to bonding making subsequent bonding stages more difficult than the previous ones.

Without end-capping, a column would be left with significant numbers of accessible, highly active residual silanols which have the potential to alter the selectivity of the stationary phase. These residual groups have the potential to form strong ionic and polar interactions (often called “secondary interactions”) with polar and basic groups on the compounds introduced to them. So, reducing the number of free silanols after bonding the stationary phase through end-capping is imperative.

Historically, end-capping was done with a small hydrophobic group, similar in nature to a trimethyl silane. In recent years several manufacturers have also introduced other groups to add a different selectivity element to the column.

Phenomenex: The Role of End-Capping in RP.Phenomenex: The Role of End-Capping in RP.

If the reagent used for end-capping contains charged functionality, for example, this will impart ion exchange functionality to the surface and give the column ion exchange properties together with the reversed phase interactions from the C18 ligands. Likewise, a polar moiety will give polar characteristics to the surface.

Phenomenex: The Role of End-Capping in RP.Phenomenex: The Role of End-Capping in RP.

Now that we have understood the role of end-capping, let’s review the main benefits of silica end-capping depending on the chemical groups used.

Phenomenex: The Role of End-Capping in RP.Phenomenex: The Role of End-Capping in RP.

In summary, new end-capping technologies can significantly improve the analysis of compounds that were once considered challenging. For example, nowadays, great peak shape and shorter run times can be achieved for the analysis of basic drugs using PS-C18 columns. The Polar C18 chemistry has been proven to best retain compounds such as metabolites, and XB-C18 chemistry has shown excellent chromatography for peptide analysis, with enhanced retention of small polar peptides.

Phenomenex
LinkedIn Logo
 

Related content

Simultaneous Quantitation and Discovery analysis: Combining targeted and untargeted metabolomics on Orbitrap mass spectrometers

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Metabolomics

Analysis of Per- and Polyfluoroalkyl Substances (PFAS) in Wastewater

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Agilent Technologies
Industries
Environmental

Non-Targeted Screening of Biosolids with the Xevo™ MRT Mass Spectrometer Reveals New Isoforms of PFAS

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/TOF, LC/HRMS
Manufacturer
Waters
Industries
Environmental

High Resolution Characterization of Lipid Nanoparticles Using the Xevo™ Charge Detection Mass Spectrometry (CDMS) Instrument - Single Particle Mass Analysis of Intact LNP-mRNA Formulations

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/HRMS, Particle characterization, LC/IT
Manufacturer
Waters
Industries
Lipidomics

Out-of-the-box workflow for PFAS quantitation using a fullscan high-resolution approach with the Orbitrap Exploris EFOX Mass Detector

Applications
| 2025 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Environmental
 

Related articles

Overcoming Peak Tailing of Basic Analytes in Silica Type A Stationary Phases in RP
Article | Academy

Overcoming Peak Tailing of Basic Analytes in Silica Type A Stationary Phases in RP

Struggling with peak tailing of basic compounds in RP-LC using type A silica columns? Discover how low-pH buffers and silanol suppressors can improve peak shape—plus why type B silica is a better choice.
Phenomenex
tag
share
more
Basic analytes and the benefits of modified silica supports in reversed-phase liquid chromatography
Article | Academy

Basic analytes and the benefits of modified silica supports in reversed-phase liquid chromatography

C18 columns with residual positive charge improve peak shape of basic analytes at low pH by reducing silanol interactions and enabling 100% aqueous mobile phases.
Phenomenex
tag
share
more
The danger of mobile phase mismatch
Article | Academy

The danger of mobile phase mismatch

After SPE, samples may not be ready for direct LC injection. Solvent mismatch with the mobile phase can distort peak shapes—proper solvent choice is essential.
Phenomenex
tag
share
more
The Role of End-Capping in Reversed-Phase
Article | Academy

The Role of End-Capping in Reversed-Phase

Learn how silica end-capping improves HPLC column performance. Discover how C18, PS-C18, Polar C18, and XB-C18 chemistries enhance retention, peak shape, and selectivity for various analytes.
Phenomenex
tag
share
more
 

Related content

Simultaneous Quantitation and Discovery analysis: Combining targeted and untargeted metabolomics on Orbitrap mass spectrometers

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Metabolomics

Analysis of Per- and Polyfluoroalkyl Substances (PFAS) in Wastewater

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Agilent Technologies
Industries
Environmental

Non-Targeted Screening of Biosolids with the Xevo™ MRT Mass Spectrometer Reveals New Isoforms of PFAS

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/TOF, LC/HRMS
Manufacturer
Waters
Industries
Environmental

High Resolution Characterization of Lipid Nanoparticles Using the Xevo™ Charge Detection Mass Spectrometry (CDMS) Instrument - Single Particle Mass Analysis of Intact LNP-mRNA Formulations

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/HRMS, Particle characterization, LC/IT
Manufacturer
Waters
Industries
Lipidomics

Out-of-the-box workflow for PFAS quantitation using a fullscan high-resolution approach with the Orbitrap Exploris EFOX Mass Detector

Applications
| 2025 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Environmental
 

Related articles

Overcoming Peak Tailing of Basic Analytes in Silica Type A Stationary Phases in RP
Article | Academy

Overcoming Peak Tailing of Basic Analytes in Silica Type A Stationary Phases in RP

Struggling with peak tailing of basic compounds in RP-LC using type A silica columns? Discover how low-pH buffers and silanol suppressors can improve peak shape—plus why type B silica is a better choice.
Phenomenex
tag
share
more
Basic analytes and the benefits of modified silica supports in reversed-phase liquid chromatography
Article | Academy

Basic analytes and the benefits of modified silica supports in reversed-phase liquid chromatography

C18 columns with residual positive charge improve peak shape of basic analytes at low pH by reducing silanol interactions and enabling 100% aqueous mobile phases.
Phenomenex
tag
share
more
The danger of mobile phase mismatch
Article | Academy

The danger of mobile phase mismatch

After SPE, samples may not be ready for direct LC injection. Solvent mismatch with the mobile phase can distort peak shapes—proper solvent choice is essential.
Phenomenex
tag
share
more
The Role of End-Capping in Reversed-Phase
Article | Academy

The Role of End-Capping in Reversed-Phase

Learn how silica end-capping improves HPLC column performance. Discover how C18, PS-C18, Polar C18, and XB-C18 chemistries enhance retention, peak shape, and selectivity for various analytes.
Phenomenex
tag
share
more
 

Related content

Simultaneous Quantitation and Discovery analysis: Combining targeted and untargeted metabolomics on Orbitrap mass spectrometers

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Metabolomics

Analysis of Per- and Polyfluoroalkyl Substances (PFAS) in Wastewater

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Agilent Technologies
Industries
Environmental

Non-Targeted Screening of Biosolids with the Xevo™ MRT Mass Spectrometer Reveals New Isoforms of PFAS

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/TOF, LC/HRMS
Manufacturer
Waters
Industries
Environmental

High Resolution Characterization of Lipid Nanoparticles Using the Xevo™ Charge Detection Mass Spectrometry (CDMS) Instrument - Single Particle Mass Analysis of Intact LNP-mRNA Formulations

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/HRMS, Particle characterization, LC/IT
Manufacturer
Waters
Industries
Lipidomics

Out-of-the-box workflow for PFAS quantitation using a fullscan high-resolution approach with the Orbitrap Exploris EFOX Mass Detector

Applications
| 2025 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Environmental
 

Related articles

Overcoming Peak Tailing of Basic Analytes in Silica Type A Stationary Phases in RP
Article | Academy

Overcoming Peak Tailing of Basic Analytes in Silica Type A Stationary Phases in RP

Struggling with peak tailing of basic compounds in RP-LC using type A silica columns? Discover how low-pH buffers and silanol suppressors can improve peak shape—plus why type B silica is a better choice.
Phenomenex
tag
share
more
Basic analytes and the benefits of modified silica supports in reversed-phase liquid chromatography
Article | Academy

Basic analytes and the benefits of modified silica supports in reversed-phase liquid chromatography

C18 columns with residual positive charge improve peak shape of basic analytes at low pH by reducing silanol interactions and enabling 100% aqueous mobile phases.
Phenomenex
tag
share
more
The danger of mobile phase mismatch
Article | Academy

The danger of mobile phase mismatch

After SPE, samples may not be ready for direct LC injection. Solvent mismatch with the mobile phase can distort peak shapes—proper solvent choice is essential.
Phenomenex
tag
share
more
The Role of End-Capping in Reversed-Phase
Article | Academy

The Role of End-Capping in Reversed-Phase

Learn how silica end-capping improves HPLC column performance. Discover how C18, PS-C18, Polar C18, and XB-C18 chemistries enhance retention, peak shape, and selectivity for various analytes.
Phenomenex
tag
share
more
 

Related content

Simultaneous Quantitation and Discovery analysis: Combining targeted and untargeted metabolomics on Orbitrap mass spectrometers

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Metabolomics

Analysis of Per- and Polyfluoroalkyl Substances (PFAS) in Wastewater

Applications
| 2026 | Agilent Technologies
Instrumentation
LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Agilent Technologies
Industries
Environmental

Non-Targeted Screening of Biosolids with the Xevo™ MRT Mass Spectrometer Reveals New Isoforms of PFAS

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/TOF, LC/HRMS
Manufacturer
Waters
Industries
Environmental

High Resolution Characterization of Lipid Nanoparticles Using the Xevo™ Charge Detection Mass Spectrometry (CDMS) Instrument - Single Particle Mass Analysis of Intact LNP-mRNA Formulations

Applications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/HRMS, Particle characterization, LC/IT
Manufacturer
Waters
Industries
Lipidomics

Out-of-the-box workflow for PFAS quantitation using a fullscan high-resolution approach with the Orbitrap Exploris EFOX Mass Detector

Applications
| 2025 | Thermo Fisher Scientific
Instrumentation
LC/MS, LC/MS/MS, LC/Orbitrap, LC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Environmental
 

Related articles

Overcoming Peak Tailing of Basic Analytes in Silica Type A Stationary Phases in RP
Article | Academy

Overcoming Peak Tailing of Basic Analytes in Silica Type A Stationary Phases in RP

Struggling with peak tailing of basic compounds in RP-LC using type A silica columns? Discover how low-pH buffers and silanol suppressors can improve peak shape—plus why type B silica is a better choice.
Phenomenex
tag
share
more
Basic analytes and the benefits of modified silica supports in reversed-phase liquid chromatography
Article | Academy

Basic analytes and the benefits of modified silica supports in reversed-phase liquid chromatography

C18 columns with residual positive charge improve peak shape of basic analytes at low pH by reducing silanol interactions and enabling 100% aqueous mobile phases.
Phenomenex
tag
share
more
The danger of mobile phase mismatch
Article | Academy

The danger of mobile phase mismatch

After SPE, samples may not be ready for direct LC injection. Solvent mismatch with the mobile phase can distort peak shapes—proper solvent choice is essential.
Phenomenex
tag
share
more
The Role of End-Capping in Reversed-Phase
Article | Academy

The Role of End-Capping in Reversed-Phase

Learn how silica end-capping improves HPLC column performance. Discover how C18, PS-C18, Polar C18, and XB-C18 chemistries enhance retention, peak shape, and selectivity for various analytes.
Phenomenex
tag
share
more
Other projects
GCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike