Agilent Ultivo Hardware - Site Preparation Checklist

Manuals | 2017 | Agilent TechnologiesInstrumentation
LC/MS, LC/MS/MS, LC/QQQ
Industries
Manufacturer
Agilent Technologies

Summary

Importance of Site Preparation


Proper laboratory setup is critical for ensuring consistent performance and longevity of high-end analytical instruments. Inadequate site conditions can lead to installation delays, suboptimal data quality, and increased maintenance costs. A systematic checklist helps laboratories fulfill spatial, utility, and environmental prerequisites before system delivery.

Study Objectives and Overview


This checklist guides customers in readying their facilities for the installation of an Agilent Ultivo LC/TQ mass spectrometer. It:
  • Outlines bench space and weight requirements
  • Defines environmental, gas supply, and power specifications
  • Specifies exhaust venting and stacking configurations
  • Assigns user responsibilities to avoid installation delays

Methodology and Instrumentation


Site requirements were categorized into key areas:
  • Laboratory space dimensions and load capacity
  • Ambient temperature, humidity, and altitude limits
  • Electrical supply standards and grounding
  • Nitrogen gas purity, pressure, and flow demands
  • Exhaust venting rates and tubing specifications
  • Recommended configuration options for module stacking
Instrumentation:
  • Agilent Ultivo LC/TQ (model G6465A)
  • Agilent Jet Stream, Electrospray (G1948B), and APCI sources (G1947B)
  • MS40+ foreline pump

Key Findings and Discussion


  • Bench footprint: the LC/TQ system requires a stable surface supporting up to 59 kg, plus 30 cm clearance on three sides.
  • Environmental window: 15–35 °C, humidity < 85% RH, non-condensing atmosphere, maximum altitude for full performance.
  • Power: single-phase or 208-Wye supply at 200–240 VAC, 50/60 Hz, with < 0.5 V rms ground-neutral potential.
  • Nitrogen: ≥95% purity from a generator or ≥99.5% from an LN₂ dewars, 5.5–6.8 bar, up to 30 L/min demand.
  • Exhaust: dual negative-pressure lines or single line with staged outlets, continuous venting up to 30 L/min.
  • Stacking: recommended single or dual LC/TQ configurations to optimize bench use and waste handling.

Benefits and Practical Applications


Correct site preparation delivers:
  • Smoother installation and commissioning process
  • Enhanced analytical precision and repeatability
  • Reduced downtime and maintenance interventions
  • Improved compliance with safety and regulatory standards

Future Trends and Opportunities


Emerging developments in lab equipment integration point to:
  • Real-time environmental monitoring with IoT sensors
  • Automated pre-installation diagnostics and self-configuration protocols
  • Digital twin simulations for optimal lab layout planning
  • Advanced predictive maintenance driven by machine learning

Conclusion


A detailed site preparation checklist is essential for the reliable operation of sophisticated LC/MS systems. By addressing spatial, environmental, utility, and safety requirements in advance, laboratories can ensure seamless instrument delivery, maintain data integrity, and optimize overall laboratory productivity.

Reference


  • Agilent Technologies. Site Preparation Checklist for Ultivo LC/TQ, Document Part Number G6465-90001, Rev. 1.2, 2017.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Agilent 6400 Series Triple Quad LC/MS - Site Preparation Checklist
Agilent 6400 Series Triple Quad LC/MS Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation…
Key words
checklist, checklistsite, sitenitrogen, nitrogenhydrocarbon, hydrocarbonspecial, specialground, groundnotes, notesfree, freetechnologies, technologiesagilent, agilentmodel, modelline, linepreparation, preparationgenerator, generatorinstrument
Agilent G6475A and 6495D LC/TQ Site Preparation Checklist
Agilent CrossLab Start Up Services Agilent G6475A and 6495D LC/TQ Site Preparation Checklist Thank you for purchasing an instrument from Agilent Technologies. CrossLab Start Up is focused on helping customers shorten the time it takes to start realizing the full…
Key words
special, specialnitrogen, nitrogenground, groundrequirements, requirementsline, linehydrocarbon, hydrocarbonservice, servicesource, sourcecustomer, customersite, sitenotes, notesfree, freeinstrument, instrumentexhaust, exhaustengineer
Agilent 6400 Series LC/TQ - Site Preparation Checklist
Agilent 6400 Series LC/TQ Site Preparation Checklist Thank you for purchasing an Agilent 6400 LC/TQ. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is…
Key words
checklist, checklistsite, sitenitrogen, nitrogenspecial, specialhydrocarbon, hydrocarbonnotes, notessource, sourcefree, freeexhaust, exhaustsupplies, suppliescustomer, customerpreparation, preparationforeline, forelineinstrument, instrumentmaximum
Agilent 6400 Series LC/TQ - Site Preparation Checklist
Agilent 6400 Series LC/TQ Site Preparation Checklist Thank you for purchasing an Agilent 6400 LC/TQ. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is…
Key words
checklist, checklistsite, sitenitrogen, nitrogenspecial, specialhydrocarbon, hydrocarbonnotes, notessource, sourcefree, freeexhaust, exhaustsupplies, suppliescustomer, customerpreparation, preparationforeline, forelineinstrument, instrumentmaximum
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike