PEPTIDE MAPPING: BEST PRACTICES FOR GENERATING RELIABLE AND ROBUST LIQUID CHROMATOGRAPHY METHODS

Posters | 2018 | Waters | HPLC SymposiumInstrumentation
HPLC
Industries
Proteomics
Manufacturer
Waters

Summary

Significance of the Topic


Reliable peptide mapping by liquid chromatography is critical for accurate identification and quantitation of protein digests in biopharmaceutical and proteomic applications. Small variations in solvent composition and gradient delivery can lead to shifts in peptide retention times and selectivity, compromising data quality and comparability across injections or over extended study periods.

Objectives and Study Overview


This study aims to compare the performance of a high-pressure binary mixing pump versus a low-pressure quaternary mixing pump in delivering reproducible gradients for peptide analysis. Using a standardized enolase digest, the retention time stability was evaluated both within single runs and across 136 days to assess intra-day and inter-day robustness.

Methodology and Instrumentation


Mobile phases consisted of 0.1% trifluoroacetic acid (TFA) in water (A) and 0.1% TFA in acetonitrile (B). Separation was performed on a Peptide BEH C18 column (1.7 µm, 2.1 × 100 mm) at 65 °C with a linear gradient from 1% to 50% B over 85 minutes at 0.2 mL/min. Detection was at 214 nm. Key instrumentation used:
  • Waters ACQUITY UPLC I-Class PLUS system
  • High-pressure binary mixing pump
  • Low-pressure quaternary mixing pump for comparison
  • ACQUITY APC reservoir caps to prevent solvent evaporation

Main Results and Discussion


Comparison of eight replicate injections showed that the high-pressure binary pump reduced retention time standard deviation by approximately threefold relative to the low-pressure quaternary pump. Over 136 days, relative retention time (RRT) measurements for selected peaks exhibited minimal drift, with standard deviations between 0.000 and 0.002 minutes. Overlay chromatograms confirmed consistent elution profiles at multiple timepoints, demonstrating the binary system’s capacity to deliver long, shallow gradients with high precision.

Benefits and Practical Applications of the Method


Enhanced gradient reproducibility supports accurate peptide identification and quantitation, vital for method validation, quality control in biopharmaceutical production, and comparative proteomics. The demonstrated stability reduces the need for frequent recalibration, increasing throughput and confidence in long-term studies.

Future Trends and Potential Applications


Advancements may include integration of real-time gradient monitoring and closed-loop feedback control to further minimize variability. Coupling with high-resolution mass spectrometry and automated sample preparation can streamline peptide mapping workflows. Machine learning-driven method optimization may also enable rapid development of robust separations tailored to complex biotherapeutic molecules.

Conclusion


The high-pressure binary mixing ACQUITY UPLC I-Class PLUS system demonstrated superior retention time precision both within runs and over extended periods compared to a low-pressure quaternary mixing system. This capability is especially important for methods employing long, shallow gradients where small compositional changes can markedly affect peptide separation.

References


  1. Simeone J, Hong P. Performance of the ACQUITY UPLC I-Class PLUS System for Methods which Employ Long, Shallow Gradients. Waters Application Note 720006290EN. May 2018.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Performance of the ACQUITY UPLC I-Class PLUS System  for Methods which Employ Long, Shallow Gradients
[ APPLICATION NOTE ] Performance of the ACQUITY UPLC I-Class PLUS System for Methods which Employ Long, Shallow Gradients Jennifer Simeone, Paula Hong, and Patricia R. McConville Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION Repeatable retention time delivery One…
Key words
enolase, enolasemassprep, massprepuplc, uplcacquity, acquityplus, plusinjections, injectionsclass, classsystem, systemreproducible, reproduciblenote, noteapplication, applicationminutes, minutesperformance, performancedigestion, digestionover
Performance Comparison of Binary Ultra High Performance  Liquid Chromatography Systems from Multiple Vendors for  Methods Which Utilize Long, Shallow Gradients
[ APPLICATION NOTE ] Performance Comparison of Binary Ultra High Performance Liquid Chromatography Systems from Multiple Vendors for Methods Which Utilize Long, Shallow Gradients Jennifer Simeone, Paula Hong, and Patricia R. McConville Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION…
Key words
vendor, vendorbinary, binaryuplc, uplcacquity, acquityplus, plusminutes, minutesclass, classenolase, enolaseperformance, performancesystems, systemsmassprep, massprepsystem, systemultra, ultrawaters, waterscomparison
Demonstrating the Applicability of the ACQUITY™ Premier Binary System for Long Shallow Gradient Peptide Mapping Analysis
Application Note Demonstrating the Applicability of the ACQUITY™ Premier Binary System for Long Shallow Gradient Peptide Mapping Analysis Corey E. Reed, Jennifer Simeone, Paula Hong Waters Corporation Abstract Reversed-Phase Liquid Chromatography (RPLC) is a key analytical tool in peptide mapping…
Key words
shallow, shallowbinary, binarypremier, premierpeptide, peptideacquity, acquitymapping, mappingapplicability, applicabilitydemonstrating, demonstratinglong, longgradient, gradientsystem, systemanalysis, analysiscompetitor, competitordeliver, deliverbiocompatible
EAS: Evaluation of Pump Performance for Long Shallow Gradient Peptide Mapping Analysis
Evaluation of Pump Performance for Long Shallow Gradient Peptide Mapping Analysis Andrew Steere, Norris Wong, Paula Hong Waters Corporation, Milford, MA, 01757 RESULTS AND DISCUSSION INTRODUCTION Peptide mapping is a critical tool in biopharmaceutical analysis for characterization and impurity testing.…
Key words
arc, arcpremier, premierquaternary, quaternarybinary, binarymapping, mappingqsm, qsmbsm, bsmsystems, systemsdeviation, deviationpeptide, peptidecshtm, cshtmxselecttm, xselecttmacross, acrossminutes, minutesnosie
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike