Demonstrating the Waters ACQUITY RDa Detector as an Ease of Use Solution for Routine Accurate Mass Measurements in Forced Degradation Studies

Applications | 2021 | WatersInstrumentation
LC/TOF, LC/HRMS, LC/MS
Industries
Pharma & Biopharma
Manufacturer
Waters

Summary

Significance of the Topic


Forced degradation testing is essential for developing stability-indicating methods in pharmaceutical analysis. It enables identification and characterization of impurities and degradation products that may form under stress conditions. High-resolution mass spectrometry (HRMS) offers precise mass data for confident compound identification but typically requires specialized expertise. The Waters ACQUITY RDa Detector simplifies accurate mass measurements, making advanced HRMS capabilities accessible to non-expert users and streamlining routine forced degradation workflows.

Objectives and Study Overview


This study demonstrates the application of the ACQUITY RDa Detector coupled with the ACQUITY UPLC I-Class PLUS system and waters_connect software for routine accurate mass analysis in forced degradation of simvastatin. Acidic, basic, and oxidative stress conditions were applied to reveal major degradants, and automated workflows were evaluated for data acquisition, processing, and reporting in a regulated environment.

Methodology and Instrumentation


A methanolic stock solution of simvastatin was subjected to three stress conditions: 0.5 M HCl at 60 °C, 5 mM NaOH at room temperature, and 10 % H₂O₂ at 60 °C. Aliquots were collected over time to monitor degradation progress. Chromatographic separation employed an ACQUITY UPLC I-Class PLUS system with a BEH C18 column (2.1 × 50 mm, 1.7 µm) at 45 °C, using a gradient from 25 % to 90 % acetonitrile (both solvents containing 0.1 % formic acid) over 8 min, flow rate 0.6 mL/min, injection volume 1 µL.
Mass spectrometric detection used the ACQUITY RDa Detector in ESI+ mode, full scan 100–2000 Da, capillary voltage 1.5 kV, cone voltage 20 V, and fragmentation cone ramp 120–170 V at 10 Hz scan rate. waters_connect platform with UNIFI software handled automated setup, calibration, acquisition, processing, fragment annotation, and report generation under SmartMS workflows.

Key Results and Discussion


Automated detector tuning and calibration yielded mass accuracy within ±3 ppm for simvastatin, its degradants, and fragment ions. Simvastatin acid was identified as the primary degradant under acidic and basic conditions, resulting from lactone ring cleavage and dihydroxy moiety formation. Under oxidative stress, a dihydrodiol derivative predominated. High and low energy spectra acquired simultaneously provided structural confirmation through fragment ion patterns. Summary plots illustrated rapid base-catalyzed degradation (complete conversion in under 1 min at room temperature) versus slower acidic and oxidative pathways (10–20 % degradation over 90 min and 5 h, respectively).

Benefits and Practical Applications


  • Routine, high-accuracy mass data for forced degradation studies without HRMS expertise.
  • Automated SmartMS workflows for end-to-end acquisition, processing, and compliant reporting.
  • Intuitive system health checks and no manual tuning or calibration required.
  • Comprehensive data integrity with full audit trails for regulated environments.
  • Enhanced confidence in impurity profiling and stability assessment in pharmaceutical development.

Future Trends and Opportunities


The integration of routine accurate mass detectors into standard QC/QA laboratories will expand, enabling broader adoption of stability-indicating mass spectrometric assays. Advances in software intelligence and machine learning for automated impurity identification will further reduce analysis time and expert intervention. Continued development of end-to-end platforms will support real-time reaction monitoring, high-throughput screening, and multi-attribute methods across diverse pharmaceutical modalities.

Conclusion


The Waters ACQUITY RDa Detector combined with ACQUITY UPLC I-Class PLUS and waters_connect/UNIFI software delivers a user-friendly, compliant workflow for accurate mass measurements in forced degradation studies. By automating setup, calibration, data acquisition, and processing, it enables chemists without HRMS expertise to obtain rich analytical information, facilitating informed decisions on drug stability and degradation profiling.

References


  1. Patolia VN. An Introduction To Forced Degradation Studies For Drug Substance & Drug Product. Pharmaceutical Online, Jan 9, 2020.
  2. Alden P, Jones MD, Lefebre P, Plumb R. Utilizing UPLC-MS for Conducting Forced Degradation Studies. Application Note.
  3. ICH Q1A(R2): Stability Testing of New Drug Substances and Products. International Conference on Harmonization.
  4. FDA Guidance for Industry: INDs for Phase II and III Studies – Chemistry, Manufacturing, and Controls Information. U.S. FDA.
  5. Laurent L, Mortishire-Smith RJ, Huisman M, Cuyckens F, Hartshorn MJ, Hill A. IsoCore: Automated Localization of Biotransformations by Mass Spectrometry Using Product Ion Scoring of Virtual Regioisomers. Rapid Communications in Mass Spectrometry, 2008;22(1):39–50.
  6. Kirk J, Mortishire-Smith R, Wrona M. A Novel Approach Using UPLC-TOF MSE and the UNIFI System to Facilitate Impurity Profiling of Pharmaceuticals. Application Note 720005604EN, Waters Corporation, Feb 2016.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Demonstrating the Waters ACQUITY RDa Detector as an Ease of Use Solution for Routine Accurate Mass Measurements in Forced Degradation Studies
Application Note Demonstrating the Waters ACQUITY RDa Detector as an Ease of Use Solution for Routine Accurate Mass Measurements in Forced Degradation Studies Chris Henry, Paul D. Rainville Waters Corporation Abstract Forced degradation is a critical analytical study for the…
Key words
simvastatin, simvastatinunifi, unifidegradation, degradationrda, rdaacquity, acquityforced, forceddegradants, degradantsautomatically, automaticallycleavage, cleavageidentify, identifyhrms, hrmstimepoints, timepointsvisualised, visualisedequips, equipsbiotransformations
Targeted and Non-Targeted Identification and Characterization of the Forced Degradation Products of Glipizide Using the ACQUITY RDa Detector and UNIFI Software Workflows
Application Note Targeted and Non-Targeted Identification and Characterization of the Forced Degradation Products of Glipizide Using the ACQUITY RDa Detector and UNIFI Software Workflows Chris Henry, Paul D. Rainville Waters Corporation Abstract In this application note we demonstrate the use…
Key words
targeted, targetedglipizide, glipiziderda, rdaforced, forcedunifi, unifidegradation, degradationacquity, acquityworkflows, workflowscharacterization, characterizationnon, nonproducts, productsdetector, detectoridentification, identificationsoftware, softwareusing
DEMONSTRATING AN EASE-OF-USE BENCHTOP TIME OF FLIGHT DETECTOR AS AN EASE-OF-USE SOLUTION FOR ACCURATE MASS MEASUREMENTS IN FORCED DEGRADATION STUDIES
DEMONSTRATING AN EASE-OF-USE BENCHTOP TIME OF FLIGHT DETECTOR AS AN EASE-OF-USE SOLUTION FOR ACCURATE MASS MEASUREMENTS IN FORCED DEGRADATION STUDIES Authors Chris Henry1([email protected]), Paul Rainville2 Affiliations: 1Waters Corporation, Stamford Avenue. Altrincham Road, Wilmslow, Cheshire, SK9 4AX, UK 2 Waters Corporation,…
Key words
glipizide, glipizideimpurity, impuritydegradants, degradantsrda, rdaacquity, acquitystressing, stressingiii, iiienergy, energyfragmentation, fragmentationdegradation, degradationunder, underforced, forcedgenerated, generatedoxidation, oxidationapi
Efficient Identification of Impurities in Nicotine API Using the SmartMS-Enabled ACQUITY RDa Detector
Application Note Efficient Identification of Impurities in Nicotine API Using the SmartMS-Enabled ACQUITY RDa Detector Tirupateswara Rao. B, Raghu Tadala, Dilshad Pullancheri, Bheeshmacharyulu. S, Dr. Padmakar Wagh Waters Corporation Abstract This application note demonstrates the efficient and easy to use…
Key words
nicotine, nicotinerda, rdaacquity, acquityimpurities, impuritiessmartms, smartmscigarettes, cigarettesworkflows, workflowsdetector, detectorditartrate, ditartrateend, endrelated, relatedrecreationally, recreationallygums, gumsdiscovery, discoveryoctadecylsilyl
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike