Fast Gradients for RP-HPLC
Technical notes | 2000 | MerckInstrumentation
Reversed-phase high-performance liquid chromatography (RP-HPLC) remains a cornerstone of analytical chemistry in pharmaceutical, environmental and industrial laboratories. The demand for rapid, high-throughput methods has driven the development of fast gradient separations on short columns, offering dramatic reductions in run time, solvent consumption and system backpressure while preserving adequate resolution.
This application note demonstrates how the combination of short (5 cm) columns packed with 5 µm particles and optimized gradient conditions can achieve baseline separations of small molecules and proteins in minutes. Key objectives include:
The study employed the following hardware and consumables:
1. Paraben Test Mix – Six parabens were separated by:
2. Excedrin® Actives – Acetaminophen, caffeine and acetylsalicylic acid were fully resolved on 5 cm C8 in under 1.5 min.
3. OTC Cough/Cold Formulation – Four compounds (phenylpropanolamine, doxylamine, dextromethorphan, acetylsalicylic acid) were eluted in <2 min using a segmented gradient (0→5% B in 0.5 min, 20→40% B in 1.5 min) at 3 mL/min.
4. Protein Standards – Ribonuclease A, cytochrome c, lysozyme and β-lactoglobulin B were baseline resolved on a 5 cm C8 in <3 min using 0–50% B in 2.75 min.
5. Extracolumn and Dwell-Volume Effects – Minimizing tubing ID and length alongside semi-micro flow cells (2.5 µL, 5 mm pathlength) enhanced peak sharpness by 50%. High-pressure mixers with 5 µL or lower dwell volumes are essential for narrow-bore columns to prevent gradient distortion.
Advancements are expected in:
Fast gradient RP-HPLC on short columns delivers significant gains in speed and efficiency without major sacrifices in separation quality. Key to success are appropriate system design choices—column dimensions, tubing ID, mixer dwell volume and optimized gradient profiles—which collectively enable robust high-throughput analyses across diverse applications.
HPLC
IndustriesManufacturerMerck
Summary
Significance of the Topic
Reversed-phase high-performance liquid chromatography (RP-HPLC) remains a cornerstone of analytical chemistry in pharmaceutical, environmental and industrial laboratories. The demand for rapid, high-throughput methods has driven the development of fast gradient separations on short columns, offering dramatic reductions in run time, solvent consumption and system backpressure while preserving adequate resolution.
Goals and Overview of the Study
This application note demonstrates how the combination of short (5 cm) columns packed with 5 µm particles and optimized gradient conditions can achieve baseline separations of small molecules and proteins in minutes. Key objectives include:
- Comparing traditional long-column gradients with fast gradients on 5 cm columns
- Illustrating separations of model compounds: parabens, Excedrin® actives, OTC cough/cold formulations and protein standards
- Addressing system design factors such as extralong column volume and dwell volume on gradient performance
Methodology and Used Instrumentation
The study employed the following hardware and consumables:
- Columns: Discovery C18, C8, RP-AmideC16 and Cyano, lengths of 5, 10 and 12.5 cm, 4.6 mm ID, 5 µm particles
- HPLC systems: low-pressure and high-pressure gradient mixers; narrow-bore tubing (0.005" ID) to minimize extracolumn band broadening
- Static mixers: stainless steel cartridges (2–25 µL) and PEEK housings for gradient dwell-volume adjustment
- Mobile phases: water (A) and acetonitrile (B) with 0.1% formic or trifluoroacetic acid modifiers
- Flow rates: 1.0–4.0 mL/min; column temperature: 35 °C; detection at 220–254 nm
Main Results and Discussion
1. Paraben Test Mix – Six parabens were separated by:
- 15 cm column, 10–80% B in 21 min at 1 mL/min (baseline resolution)
- 5 cm column, same gradient scaled to 7 min at 1 mL/min (3× speedup, baseline resolution)
- 5 cm column, gradient in 1.75 min at 4 mL/min (12× speedup with adequate separation of four major bands)
2. Excedrin® Actives – Acetaminophen, caffeine and acetylsalicylic acid were fully resolved on 5 cm C8 in under 1.5 min.
3. OTC Cough/Cold Formulation – Four compounds (phenylpropanolamine, doxylamine, dextromethorphan, acetylsalicylic acid) were eluted in <2 min using a segmented gradient (0→5% B in 0.5 min, 20→40% B in 1.5 min) at 3 mL/min.
4. Protein Standards – Ribonuclease A, cytochrome c, lysozyme and β-lactoglobulin B were baseline resolved on a 5 cm C8 in <3 min using 0–50% B in 2.75 min.
5. Extracolumn and Dwell-Volume Effects – Minimizing tubing ID and length alongside semi-micro flow cells (2.5 µL, 5 mm pathlength) enhanced peak sharpness by 50%. High-pressure mixers with 5 µL or lower dwell volumes are essential for narrow-bore columns to prevent gradient distortion.
Benefits and Practical Applications
- Up to 12× faster analyses with minimal resolution loss
- Lower solvent usage and operating costs
- Adaptability to a broad range of analytes from small molecules to small proteins
- Enhanced throughput for QC laboratories, drug profiling and proteomic screening
Future Trends and Potential Applications
Advancements are expected in:
- Column chemistries optimized for sub-2 µm particles and ultra-high pressure
- Automated gradient optimization using machine-learning algorithms
- Integrated microfluidic mixers for sub-microliter dwell-volume control
- Applications in real-time process monitoring, metabolomics and high-throughput combinatorial chemistry
Conclusion
Fast gradient RP-HPLC on short columns delivers significant gains in speed and efficiency without major sacrifices in separation quality. Key to success are appropriate system design choices—column dimensions, tubing ID, mixer dwell volume and optimized gradient profiles—which collectively enable robust high-throughput analyses across diverse applications.
References
- Goetzinger W.K., Kyroanos J.N. Fast gradient methodology: balancing productivity and resolution. American Laboratory, 1998, 30, 27–37.
- Weller H.N. et al. Rapid reversed-phase separations of combinatorial libraries. Molecular Diversity, 1997, 3, 61–70.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
New Paradigm for HPLC Method Development
2000|Merck|Technical notes
VOL 18.1 TheReporter www.sigma-aldrich.com/TheReporter New Paradigm for HPLC Method Development Column: Mobile Phase: Discovery, 5cm x 4.6mm ID, 5µm particles A: O.1% HCO2H (pH 2.5) B: MeCN 0 - 4% in 0.8 min., 4 - 30% in 1.7 min. 3mL/min…
Key words
mecn, mecnpenicillin, penicillincyano, cyanolabpro, labprobonded, bondedantibiotics, antibioticsrheodyne, rheodynepseudoephedrine, pseudoephedrinepercent, percentdiscovery, discoveryacetaminophen, acetaminophenmau, mauphases, phaseslactam, lactamampicillin
Optimizing HPLC Separations: Samples with Widely Differing Polarities
2000|Merck|Technical notes
VOL 18.3 TheReporter www.sigma-aldrich.com/TheReporter Optimizing HPLC Separations: Samples with Widely Differing Polarities When confronted with a sample containing widely varying polarities, the LC chromatographer will typically rely on gradient elution. However, if gradient elution is not an option, isocratic elution…
Key words
paclitaxel, paclitaxeldiscovery, discoverydiethylstilbestrol, diethylstilbestrolmin, mintime, timehplc, hplcguard, guardsepara, separacolumns, columnsaldrich, aldrichsigma, sigmaoptimizing, optimizingcyanocobalamin, cyanocobalaminreagents, reagentsriboflavin
Scale-Up of Fast Analytical to Fast Preparative Separations
2000|Merck|Technical notes
VOL 18.10 TheReporter www.sigma-aldrich.com/TheReporter Scale-Up of Fast Analytical to Fast Preparative Separations This article describes a simple, reproducible, and predictable process for scaling up from fast analytical to fast preparative HPLC separations. Chromatographers performing lab scale preparative or larger process…
Key words
preparative, preparativecustom, customsupelcosil, supelcosilsupelco, supelcoresin, resincolumn, columnscale, scaleanalytical, analyticaldiameter, diameterliterature, literatureyour, yourabz, abzmedia, mediaseal, sealuracil
The Effect of Ion Pair Reagent on Reversed-Phase Selectivity of Polypeptide Separations
2002|Merck|Technical notes
VOL 20.3 TheReporter sigma-aldrich.com/TheReporter The Effect of Ion Pair Reagent on Reversed-Phase Selectivity of Polypeptide Separations Reversed-phase high performance liquid chromatography (RP-HPLC) has been a preferred method for resolution of peptides by virtue of its resolving power and suitability of…
Key words
custom, customdiscovery, discoverydsc, dscpore, porebio, bioreversed, reversedphase, phaseamide, amidehplc, hplcwide, widemicrobore, microboreangiotensin, angiotensinreagent, reagentcytochrome, cytochromebüchner