New Paradigm for HPLC Method Development

Technical notes | 2000 | MerckInstrumentation
HPLC
Industries
Manufacturer
Merck

Summary

Importance of the Topic


Reverse-phase high-performance liquid chromatography (RP-HPLC) remains a cornerstone in pharmaceutical analysis across all drug discovery stages. Establishing robust and selective methods is critical for high-throughput screening, purity assays, metabolite profiling, and regulatory compliance.

Objectives and Overview of the Study


This article introduces a new paradigm for HPLC method development by emphasizing the role of bonded phase chemistry. It compares traditional C18, C8 and cyano phases with a polar-embedded RP-AmideC16 phase and demonstrates how simple column switching accelerates method optimization and enhances selectivity.

Methodology


Identical gradient or isocratic conditions were applied to multiple bonded phases (Discovery RP-AmideC16, C18, C8, Cyano) of 4.6×50–150 mm, 5 µm particle size. Typical mobile phases included aqueous buffers (formic or phosphate acidified to pH 2.5–7) and acetonitrile. Flow rates ranged from 1 to 4 mL/min at 25–35 °C. Detection was performed by UV at 220 or 254 nm. Sample sets comprised over-the-counter cold remedies, fluoroquinolones, β-lactam antibiotics, and basic diphenylmethane drugs.

Used Instrumentation


  • Supelco Discovery HPLC columns: RP-AmideC16, C18, C8, Cyano
  • Mobile phase delivery system compatible with gradient elution
  • UV detection at 220 nm or 254 nm
  • Rheodyne LabPro Column Selector for automated switching among up to six columns

Main Results and Discussion


The polar-embedded RP-AmideC16 phase provided unique selectivity through combined hydrophobic and hydrogen-bond interactions. Key findings include:
  • OTC formulation (pseudoephedrine, acetaminophen, chlorpheniramine, dextromethorphan) was baseline-resolved in < 3 min on RP-AmideC16. C8 and cyano phases showed incomplete separations or poor retention; C18 required longer gradients.
  • Fluoroquinolone antibiotics (ofloxacin, norfloxacin, ciprofloxacin, lomefloxacin) achieved baseline separation of critical pairs on RP-AmideC16 via amide–piperazine hydrogen bonding, whereas C18 co-eluted ofloxacin and norfloxacin.
  • Separation of seven β-lactams (penicillins G, V, amoxicillin, ampicillin, cloxacillin, piperacillin, amoxicillin) demonstrated that C8, C18 and RP-AmideC16 resolve all analogues under a moderate gradient. Cyano required shallower gradient profiles to achieve equivalent resolution.
  • Basic diphenylmethane drugs were separated in < 12 min only on the cyano phase; other phases offered distinct but much longer elution times.
  • Mass overload studies on narrow (< 50 mm) columns highlighted the necessity to verify sample capacity. Dilution experiments on RP-AmideC16 revealed distorted peaks when overloaded, underscoring careful loading control on short/narrow columns.

Benefits and Practical Applications


Column switching among bonded phases provides a rapid, low-cost tool to fine-tune selectivity without extensive solvent optimization. Polar-embedded phases such as RP-AmideC16 offer enhanced resolution for polar and basic analytes, improving throughput in QC, LC–MS workflows, and high-throughput screening.

Future Trends and Possibilities


Continued development of polar-embedded and mixed-mode stationary phases will expand selectivity options for complex analytes. Integration with automated column selectors and advanced modeling software will further accelerate method discovery. Constraints imposed by LC–MS detection will drive innovation toward phases that balance MS compatibility with high resolution.

Conclusion


Emphasizing bonded phase selection alongside mobile phase optimization introduces a powerful paradigm for rapid, efficient HPLC method development. The RP-AmideC16 phase and automated column switching streamline the path to high-resolution separations of challenging pharmaceutical compounds.

References


  1. R.K. Gilpin, L.A. Pachia, Anal. Chem. 71 (1999) 217R–233R.
  2. J. Zweigenbaum et al., Anal. Chem. 71 (1999) 2294–2300.
  3. L. Huber, Biopharm 12 (1999) 64–66.
  4. R.E. Majors, LC-GC 17 (1999) 212–228.
  5. P.T. Jackson, P.W. Carr, Chemtech 29 (1999) 29–37.
  6. L.R. Snyder, J. Chromatogr. B 689 (1999) 105–115.
  7. L.R. Snyder, J.J. Kirkland, J.L. Glajch, Practical HPLC Method Development, Wiley, 1997.
  8. R. Stevenson, American Laboratory 31(23) (1999) 4–8.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Fast Gradients for RP-HPLC
Fast Gradients for RP-HPLC
2000|Merck|Technical notes
VOL 18.6 TheReporter www.sigma-aldrich.com/TheReporter (continued on page 3) LC issue Conditions for A, B & C differ only in column length, flow and gradient. Panel A. Column: Discovery C18, 15cm x 4.6mm ID, 5µ Cat. No.: 504955 Sample: Gradient RP…
Key words
mecn, mecndiscovery, discoverymixer, mixerpercent, percentmau, maufamily, familyminutes, minutescyano, cyanomicro, microvolume, volumehplc, hplcdwell, dwellliterature, literaturestatic, staticcolumns
Use Discovery to Reduce HPLC Method Development Time
595 North Harrison Road Bellefonte, PA 16823-0048 USA Telephone 800-247-6628 • 814-359-3441 Fax 800-447-3044 • 814-359-3044 email: supelco@s ial.com sigma-aldrich.com/supelco Bulletin 937 Use Discovery to Reduce HPLC Method Development Time Following the Discovery method development protocol and using the Column…
Key words
buffer, bufferphenacetin, phenacetinnitrobenzoic, nitrobenzoicparaben, parabendiscovery, discoverycodeine, codeineacid, acidbenzene, benzenecyano, cyanofunctionalized, functionalizedphase, phasepropyl, propylnitrosobenzene, nitrosobenzenebenzyl, benzylhexafluorobenzene
Optimizing HPLC Separations: Samples with Widely Differing Polarities
VOL 18.3 TheReporter www.sigma-aldrich.com/TheReporter Optimizing HPLC Separations: Samples with Widely Differing Polarities When confronted with a sample containing widely varying polarities, the LC chromatographer will typically rely on gradient elution. However, if gradient elution is not an option, isocratic elution…
Key words
paclitaxel, paclitaxeldiscovery, discoverydiethylstilbestrol, diethylstilbestrolmin, mintime, timeguard, guardhplc, hplcsepara, separacolumns, columnsaldrich, aldrichsigma, sigmaoptimizing, optimizingcyanocobalamin, cyanocobalaminreagents, reagentsriboflavin
HPLC Column Comparison Screening Study for Reversed Phase Columns
HPLC Column Comparison Screening Study for Reversed Phase Columns Carmen T. Santasania, David S. Bell, and Tracy L. Ascah-Ross Supelco, 595 North Harrison Road, Bellefonte, PA 16823, USA T403157 GIB Introduction The selection of the proper HPLC column for any…
Key words
phenacetin, phenacetincodeine, codeinecyano, cyanostudy, studyselectivity, selectivityscreening, screeningcolumn, columnphases, phasesalpha, alphaphase, phasechosen, chosenhplc, hplcsystematic, systematicgib, gibpredicted