Analysis of Cetirizine Hydrochloride Using a Core Enhanced Technology Accucore HPLC Column
Applications | 2011 | Thermo Fisher ScientificInstrumentation
Cetirizine hydrochloride is a widely used antihistamine for allergy and hay fever treatment. Reliable and rapid analysis is essential in pharmaceutical quality control. Conventional USP methods employ strong acids, which can damage chromatography equipment and result in longer run times.
This study evaluates the performance of a Thermo Scientific Accucore HILIC 2.6 µm column as an alternative to the USP monograph method. The goal is to develop a non-aggressive, high-efficiency approach for cetirizine hydrochloride analysis.
Sample preparation involved a working standard of 50 µg/mL cetirizine hydrochloride in 5:95 water/acetonitrile (v/v). Chromatographic conditions were:
Cetirizine eluted in less than 1 minute, representing a more than tenfold reduction in analysis time compared to the USP method (≈11 min). System suitability criteria were met with a tailing factor of 1.21, %RSD retention time of 1.39, and %RSD peak area of 0.17. The non-aggressive buffer preserved column and system integrity.
Superficially porous particles in HILIC mode show promise for rapid analysis of other polar pharmaceuticals and complex matrices. Future developments may include coupling with mass spectrometry for trace-level quantification and expanding method scope to related antihistamines.
The Accucore HILIC column offers a robust, rapid, and precise method for cetirizine hydrochloride analysis without aggressive acids. This approach enhances laboratory efficiency and instrument longevity, making it ideal for high-throughput QA/QC environments.
Thermo Fisher Scientific. Analysis of Cetirizine Hydrochloride Using a Core Enhanced Technology Accucore HPLC Column. Application Note ANCCSCETCHYDCHL, 2011.
Consumables, HPLC, LC columns
IndustriesPharma & Biopharma
ManufacturerThermo Fisher Scientific
Summary
Importance of Topic
Cetirizine hydrochloride is a widely used antihistamine for allergy and hay fever treatment. Reliable and rapid analysis is essential in pharmaceutical quality control. Conventional USP methods employ strong acids, which can damage chromatography equipment and result in longer run times.
Objectives and Study Overview
This study evaluates the performance of a Thermo Scientific Accucore HILIC 2.6 µm column as an alternative to the USP monograph method. The goal is to develop a non-aggressive, high-efficiency approach for cetirizine hydrochloride analysis.
Methodology and Instrumentation
Sample preparation involved a working standard of 50 µg/mL cetirizine hydrochloride in 5:95 water/acetonitrile (v/v). Chromatographic conditions were:
- Column: Accucore HILIC 2.6 µm, 50 × 2.1 mm
- System: Thermo Scientific Accela HPLC with UV detection at 230 nm
- Mobile phase: 90:10 acetonitrile/200 mM ammonium acetate, pH 5.0
- Flow rate: 0.4 mL/min
- Column temperature: 30 °C
- Injection volume: 1.0 µL
Main Results and Discussion
Cetirizine eluted in less than 1 minute, representing a more than tenfold reduction in analysis time compared to the USP method (≈11 min). System suitability criteria were met with a tailing factor of 1.21, %RSD retention time of 1.39, and %RSD peak area of 0.17. The non-aggressive buffer preserved column and system integrity.
Benefits and Practical Applications
- High sample throughput due to sub-minute analysis.
- Elimination of corrosive acids reduces maintenance and prolongs equipment life.
- Precision and peak shape meet or exceed USP requirements.
Future Trends and Potential Applications
Superficially porous particles in HILIC mode show promise for rapid analysis of other polar pharmaceuticals and complex matrices. Future developments may include coupling with mass spectrometry for trace-level quantification and expanding method scope to related antihistamines.
Conclusion
The Accucore HILIC column offers a robust, rapid, and precise method for cetirizine hydrochloride analysis without aggressive acids. This approach enhances laboratory efficiency and instrument longevity, making it ideal for high-throughput QA/QC environments.
Reference
Thermo Fisher Scientific. Analysis of Cetirizine Hydrochloride Using a Core Enhanced Technology Accucore HPLC Column. Application Note ANCCSCETCHYDCHL, 2011.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Analysis of Melamine and Cyanuric Acid Using a Core Enhanced Technology Accucore HILIC HPLC column
2011|Thermo Fisher Scientific|Applications
Application Note: ANCCSCETMELCYAN Analysis of Melamine and Cyanuric Acid Using a Core Enhanced Technology Accucore HILIC HPLC column Joanna Freeke, Thermo Fisher Scientific, Runcorn, Cheshire, UK Abstract Key Words • Accucore HILIC • Fused core • Superficially porous • Food…
Key words
accucore, accucorecyanuric, cyanuricscientific, scientifichilic, hilicmelamine, melaminefisher, fisherthermo, thermoacid, acidcore, corehplc, hplcnorth, northamerica, americadomestic, domesticltd, ltdenhanced
Analysis of Catecholamines Using a Core Enhanced Technology Accucore HPLC Column
2011|Thermo Fisher Scientific|Applications
Application Note: ANCCSCETCATECH Analysis of Catecholamines Using a Core Enhanced Technology Accucore HPLC Column Joanna Freeke, Thermo Fisher Scientific, Runcorn, Cheshire, UK Abstract Key Words • Accucore HILIC • Fused core • Superficially porous • Catecholamines • Hormones • Neurotransmitters…
Key words
accucore, accucorescientific, scientificfisher, fisherhplc, hplchilic, hilicthermo, thermocatecholamines, catecholaminescatecholamine, catecholamineneurotransmitters, neurotransmittersnorth, northamerica, americacore, coreseven, sevendomestic, domestichormones
Comparison of Epinephrine and Norepinephrine Analysis Using Core Enhanced Technology Accucore HPLC Columns with HILIC and aQ Stationary Phases
2011|Thermo Fisher Scientific|Applications
Application Note: ANCCSCETHORMONE Comparison of Epinephrine and Norepinephrine Analysis Using Core Enhanced Technology Accucore HPLC Columns with HILIC and aQ Stationary Phases Joanna Freeke and Anila I. Khan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Abstract Key Words • Accucore HILIC…
Key words
accucore, accucoreepinephrine, epinephrinehilic, hilicnorepinephrine, norepinephrinescientific, scientificfisher, fisherhplc, hplcthermo, thermoadrenaline, adrenalinecolumns, columnsneurotransmitters, neurotransmittersnorth, northamerica, americacore, corebackpressures
Analysis of Fenoprofen Using a Core Enhanced Technology Accucore HPLC Column
2011|Thermo Fisher Scientific|Applications
Application Note: ANCCSCETFENOPRO Analysis of Fenoprofen Using a Core Enhanced Technology Accucore HPLC Column Monica Dolci, Thermo Fisher Scientific, Runcorn, Cheshire, UK Abstract Key Words • Accucore RP-MS • Fused core • Superficially porous • Fenoprofen • USP This application…
Key words
fenoprofen, fenoprofenaccucore, accucorescientific, scientificfisher, fisherthermo, thermousp, usphplc, hplcnorth, northamerica, americacore, coredomestic, domesticanccscetfenopro, anccscetfenoproltd, ltdpart, partunited