Installation Instruction for ProfIC Vario 6 Cation

Manuals |  | MetrohmInstrumentation
Ion chromatography
Industries
Manufacturer
Metrohm

Summary

Importance of the topic


The ProfIC Vario 6 Cation system integrates inline dilution and ultrafiltration to enable reliable ion chromatographic analysis of high concentration samples containing particles or microorganisms. Precise sample pretreatment in one automated workflow reduces manual handling, minimizes contamination risk, and ensures reproducible results in quality control, environmental monitoring, and industrial process analytics.

Objectives and overview


This application bulletin provides step-by-step installation and configuration guidelines for the ProfIC Vario 6 Cation system. Key aims include software setup, hardware assembly, tubing and capillary connections, instrument interconnection, method configuration in MagIC Net, and demonstration of system performance through a representative cation analysis.

Instrumentation used


  • Professional IC Vario ONE system for non-suppressed cation analysis
  • IC Conductivity Detector
  • 858 Professional Sample Processor with inline dilution and inline ultrafiltration modules
  • 800 Dosino dosing unit
  • MagIC Net 3.0 software
  • Metrosep C 4 - 150/4.0 analytical column and guard column


Methodology and installation procedure


The installation begins with MagIC Net software deployment and driver installation for USB-connected instruments. Hardware assembly covers mounting the detector block, sample processor, inline dilution (LQH Station), ultrafiltration cell, and eluent production accessories. Detailed tubing and capillary connections ensure appropriate flow paths: coarse lines for sample feed, finer capillaries for filtrate, and correct use of PEEK and PVDF fittings to prevent leakage. The 940 Professional IC Vario is configured with eluent supply, column inlet, detector in/out lines, and the sample injection valve. Software configuration in MagIC Net involves device recognition, defining rack positions, Dosino settings, method import from example files, method parameter adjustment (column oven at 30 C), system purge, and equilibration.

Main results and discussion


A single 20 mg/L standard containing Li+, Na+, NH4+, K+, Ca2+, and Mg2+ was used for calibration by automatic inline dilution. Linear calibration across 0.2–10 mg/L showed correlation coefficients > 0.99985 for all ions. Precision studies (n=10 for a 10 mg/L equivalent) yielded relative standard deviations below 1.1% for monovalent cations and up to 2.3% for divalent cations. Inline dilution and ultrafiltration proved effective at maintaining system performance and minimizing carry-over.

Benefits and practical applications


  • Fully automated inline dilution and particle removal save time and reduce manual preparation errors
  • Robust performance for samples with high ionic strength or particulate matrices
  • Flexible method configuration via MagIC Net for diverse cationic analyses in environmental, industrial, and QA/QC laboratories


Future trends and possibilities


Integration of eluent production modules directly in MagIC Net could streamline routine maintenance and eluent management. Advancements in inline sample preparation, including additional inline derivatization or pre-concentration steps, may extend the system’s applicability to trace analysis and complex sample matrices. Remote monitoring and control via network solutions will further enhance automation in high-throughput laboratories.

Conclusion


The ProfIC Vario 6 Cation system, with its integrated inline dilution and ultrafiltration capabilities, delivers reproducible, high-quality cationic analyses for challenging sample types. Adherence to the detailed installation and configuration workflow ensures optimal system performance and reliable data in a fully automated environment.

Reference


No external literature references were provided in the source bulletin.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Installation Instruction for ProfIC Vario 6 Anion
Application Bulletin 376 Installation Instruction for ProfIC Vario 6 Anion The ProfIC Vario 6 Anion is a Professional IC Vario system with Inline Dilution and Inline Ultrafiltration for the fully automatic determination of anions or cations with sequential suppression in…
Key words
msm, msmrinsing, rinsingultrafiltration, ultrafiltrationpump, pumpconnected, connectedrotor, rotorperistaltic, peristalticregeneration, regenerationtubing, tubingstation, stationprofic, proficcapillary, capillarystoppers, stopperslabeled, labeledcapillaries
Installation Instruction for ProfIC Vario 6 AnCat
Application Bulletin 378 Installation Instruction for ProfIC Vario 6 AnCat The ProfIC Vario 6 AnCat is a Professional IC Vario system with Inline Dilution and Inline Ultrafiltration for the fully automatic and parallel determination of anions and cations in very…
Key words
msm, msmrinsing, rinsingpump, pumpperistaltic, peristalticexemplary, exemplaryconnected, connectedultrafiltration, ultrafiltrationmcs, mcsrotor, rotorconnection, connectionprocessor, processorregeneration, regenerationstandard, standardcapillaries, capillariesanion
Installation Instruction for ProfIC Vario 2 Cat
Application Bulletin 366 Installation Instruction for ProfIC Vario 2 Cat ProfIC Vario 2 Cat is a very compact and Professional IC Vario system for the fully automatic determination of nonsuppressed cations or anions with Metrohm Inline Ultrafiltration. «ProfIC Vario 2»…
Key words
station, stationhandling, handlingrinsing, rinsingultrafiltration, ultrafiltrationfiltrate, filtrateeluent, eluentscrews, screwsconveying, conveyingliquid, liquidprocessor, processorperistaltic, peristalticcapillaries, capillariesconnected, connectedsample, samplerack
Installation Instruction for ProfIC Vario 2 Anion
Application Bulletin 365 Installation Instruction for ProfIC Vario 2 Anion ProfIC Vario 2 Anion is a Professional IC Vario system for the fully automatic determination of anions or cations with sequential suppression with Metrohm Inline Ultrafiltration. «ProfIC Vario 2» safely…
Key words
msm, msmstation, stationrinsing, rinsingultrafiltration, ultrafiltrationregeneration, regenerationhandling, handlingconnected, connectedrotor, rotorperistaltic, peristalticcapillary, capillarystoppers, stopperslabeled, labeledexemplary, exemplarysuppressor, suppressormcs
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike