Shimadzu IC-A3/-GA3 HPLC Column
Manuals | | ShimadzuInstrumentation
Ion chromatography of inorganic and organic anions is essential for environmental monitoring, water quality testing and industrial process control. The Shim-pack IC-A3 column provides high theoretical plate number and selectivity in non-suppressed anion analysis.
This report presents the design and performance evaluation of the Shim-pack IC-A3 analytical column and its GA3 guard column. It aims to provide practical guidelines for installation, sample preparation, operational parameters and maintenance procedures to ensure reliable and reproducible separations.
The analytical setup comprises a polymer based anion exchange column Shim-pack IC-A3 with a quaternary ammonium group and a guard column Shim-pack IC-GA3. Typical chromatographic conditions are:
Factory testing ensures a nitrate ion theoretical plate number greater than 5000 and consistent retention under the specified conditions. The column tolerates up to 70% organic modifier but high solvent content may increase backpressure. Periodic flushing with dedicated solvents restores performance after contamination. Flushing protocols are provided for ionic hydrophobic and metal-contaminated samples.
The Shim-pack IC-A3 column combines robust polymer chemistry with efficient non-suppressed detection offering:
Advances in suppressed conductivity detection coupling with mass spectrometry and microfluidic formats will enhance sensitivity and throughput. Development of novel stationary phases and greener mobile phases will expand applications in environmental and biomedical fields.
The Shim-pack IC-A3 provides a reliable and versatile platform for anion analysis in non-suppressed ion chromatography. Following the recommended operational and maintenance guidelines ensures stable performance and longevity of the column.
Consumables, LC columns
IndustriesManufacturerShimadzu
Summary
Significance of the Topic
Ion chromatography of inorganic and organic anions is essential for environmental monitoring, water quality testing and industrial process control. The Shim-pack IC-A3 column provides high theoretical plate number and selectivity in non-suppressed anion analysis.
Objectives and Study Overview
This report presents the design and performance evaluation of the Shim-pack IC-A3 analytical column and its GA3 guard column. It aims to provide practical guidelines for installation, sample preparation, operational parameters and maintenance procedures to ensure reliable and reproducible separations.
Methodology and Used Instrumentation
The analytical setup comprises a polymer based anion exchange column Shim-pack IC-A3 with a quaternary ammonium group and a guard column Shim-pack IC-GA3. Typical chromatographic conditions are:
- Mobile phase 8 mmol/L p-hydroxybenzoic acid 3.2 mmol/L Bis-Tris 50 mmol/L boric acid pH 3–8.5
- Flow rate 1.2 mL/min
- Column temperature 40 °C
- Detection Non-suppressed conductivity
- Injection volume 10 μL
Main Results and Discussion
Factory testing ensures a nitrate ion theoretical plate number greater than 5000 and consistent retention under the specified conditions. The column tolerates up to 70% organic modifier but high solvent content may increase backpressure. Periodic flushing with dedicated solvents restores performance after contamination. Flushing protocols are provided for ionic hydrophobic and metal-contaminated samples.
Benefits and Practical Applications
The Shim-pack IC-A3 column combines robust polymer chemistry with efficient non-suppressed detection offering:
- High resolution separation of diverse anions
- Wide pH and organic solvent compatibility
- Simplified maintenance using a guard column and flushing procedures
Future Trends and Potential Applications
Advances in suppressed conductivity detection coupling with mass spectrometry and microfluidic formats will enhance sensitivity and throughput. Development of novel stationary phases and greener mobile phases will expand applications in environmental and biomedical fields.
Conclusion
The Shim-pack IC-A3 provides a reliable and versatile platform for anion analysis in non-suppressed ion chromatography. Following the recommended operational and maintenance guidelines ensures stable performance and longevity of the column.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Shimadzu IC-SA4 HPLC Column
|Shimadzu|Manuals
228-91805A Jul.2023 Shimadzu High Performance Packed Column for HPLC Shim-pack IC-SA4 Instruction Manual NOTE: Compound retention times and peak shape may vary with usage. Before developing an analytical method with the column, verify that ■ Introduction Shim-pack IC-SA4 is a…
Key words
カラム, カラムイオン, イオンサプレッサ, サプレッサピーク, ピークガードフィルタ, ガードフィルタcolumn, columnイオンクロマトグラフィー, イオンクロマトグラフィーメイルナット, メイルナットpeek, peekflushing, flushingguard, guardexit, exitmay, maywaste, wasteパッケー
Shim-pack IC-A1/IC-GA1
2025|Shimadzu|Manuals
228-10327A Jan. 2025 High Performance Packed Column for HPLC Shim-pack IC-A1/IC-GA1 INSTRUCTION MANUAL 1. Introduction The Shim-pack IC-A1/GA1 is an ion chromatography column filled with a polymer gel that contains anionic exchange groups. The Shimpack IC-A1 is an analytical column…
Key words
カラム, カラムイオン, イオンガードカラム, ガードカラムメイルナット, メイルナットイオンクロマトグラフィー, イオンクロマトグラフィーピーク, ピークメタノール, メタノールcleaning, cleaningpeek, peekアミノメタン, アミノメタンメンブランフィルター, メンブランフィルターノンサプレッサ, ノンサプレッササイズ, サイズグレード, グレードcolumn
Shim-pack IC-C1/IC-GC1
2025|Shimadzu|Manuals
228-10323B Jan. 2025 High Performance Packed Column for HPLC Shim-pack IC-C1/IC-GC1 INSTRUCTION MANUAL 1. Introduction Shim-pack IC-C1/GC1 is a polymer gel based column for ion chromatography. Shim-pack IC-C1 is a cation exchange column optimized for non-suppressor ion chromatography. By choosing…
Key words
カラム, カラムイオン, イオンガードカラム, ガードカラムメイルナット, メイルナットイオンクロマトグラフィー, イオンクロマトグラフィーピーク, ピークpeek, peekcolumn, columnメンブランフィルター, メンブランフィルターcleaning, cleaningノンサプレッサ, ノンサプレッササイズ, サイズグレード, グレードmultivalent, multivalenteluent
Shim-pack NovaCore C18-HB [Reversed phase column]
|Shimadzu|Manuals
日本語取扱説明書はP.4以降をご参照ください。 ERAD-0001-9087A High Performance Packed Column for HPLC Shim-pack NovaCore C18-HB [Reversed phase column] 取扱説明 書 INSTRUCTION MANUAL ■ Introduction NOTE Near the pH limit of use, please use a mobile phase containing at least 10% organic solvent. At the…
Key words
カラム, カラムシステム, システムイオンペア, イオンペアメイルナット, メイルナットピーク, ピークcolumn, columnthf, thfmobile, mobilesus, susノイズ, ノイズベースラインドリフト, ベースラインドリフトprecautions, precautionsチューブ, チューブphase, phaseflow