Ion chromatography with non-suppressed detection is a widely used technique for the quantification of inorganic anions and organic acid anions in various fields including environmental monitoring, quality control and industrial analysis. The Shim-pack IC-A1 column offers robust performance across a broad pH range (2–11), enabling flexible method development and reliable separation under both acidic and alkaline conditions.
This manual aims to outline the design features, operational guidelines, and maintenance procedures for the Shim-pack IC-A1 analytical column and its companion Shim-pack IC-GA1 guard column. It provides a comprehensive overview of specifications, installation protocols, eluent selection, sample requirements, cleaning strategies, storage instructions, and troubleshooting tips to ensure optimal column performance and longevity.
The Shim-pack IC-A1 employs a poly(methacrylate) gel matrix functionalized with quaternary ammonium groups for anion exchange. Standard operating conditions include:
The recommended instrumentation setup includes:
At shipment, each Shim-pack IC-A1 meets performance criteria of more than 2 000 theoretical plates for a 10 mg/L nitrate peak (35 µL injection). Performance may vary over time due to contamination by metal ions, proteins, or hydrophobic compounds, which can elevate backpressure and alter retention. Cleaning procedures using appropriate aqueous or organic solvents can partially restore function. Guard column usage effectively prolongs the analytical column’s life.
The method enables:
Emerging directions include integrating non-suppressed IC with mass spectrometric detection, miniaturized column formats for faster analysis, eco-friendly eluents, and automated cleaning protocols. Advances in polymer-based stationary phases may further enhance selectivity and reduce run times.
The Shim-pack IC-A1 and IC-GA1 guard column combination provides a versatile and durable solution for non-suppressed ion chromatography. Adherence to recommended installation, operating, and maintenance procedures ensures consistent separation performance and extended column lifetime.
Consumables, LC columns, Ion chromatography
IndustriesManufacturerShimadzu
Summary
Importance of the Topic
Ion chromatography with non-suppressed detection is a widely used technique for the quantification of inorganic anions and organic acid anions in various fields including environmental monitoring, quality control and industrial analysis. The Shim-pack IC-A1 column offers robust performance across a broad pH range (2–11), enabling flexible method development and reliable separation under both acidic and alkaline conditions.
Objectives and Overview
This manual aims to outline the design features, operational guidelines, and maintenance procedures for the Shim-pack IC-A1 analytical column and its companion Shim-pack IC-GA1 guard column. It provides a comprehensive overview of specifications, installation protocols, eluent selection, sample requirements, cleaning strategies, storage instructions, and troubleshooting tips to ensure optimal column performance and longevity.
Applied Methodology
The Shim-pack IC-A1 employs a poly(methacrylate) gel matrix functionalized with quaternary ammonium groups for anion exchange. Standard operating conditions include:
- Eluent: 2.5 mmol/L phthalic acid + 2.4 mmol/L TRIS, supplemented with 16.2 mmol/L boric acid
- Flow rate: 1.5 mL/min (max 2.0 mL/min)
- Column temperature: 40 °C (max 50 °C)
- pH range: 2–11
- Detection: non-suppressed conductivity
Instrumentation
The recommended instrumentation setup includes:
- HPLC pump capable of stable low-pressure delivery (< 2.0 MPa)
- Injector equipped with a pre-heating coil (P/N 228-45714-91) to minimize temperature fluctuations
- Conductivity detector in non-suppressor mode
- PEEK tubing (0.25–0.3 mm ID, 1.6 mm OD) and male nuts (P/N 228-18565-84)
Main Results and Discussion
At shipment, each Shim-pack IC-A1 meets performance criteria of more than 2 000 theoretical plates for a 10 mg/L nitrate peak (35 µL injection). Performance may vary over time due to contamination by metal ions, proteins, or hydrophobic compounds, which can elevate backpressure and alter retention. Cleaning procedures using appropriate aqueous or organic solvents can partially restore function. Guard column usage effectively prolongs the analytical column’s life.
Benefits and Practical Applications
The method enables:
- Wide eluent pH tolerance for tailored anion separation
- Rapid analysis of inorganic and organic anions in diverse matrices
- Simplified maintenance via routine guard column replacement and targeted cleaning
- Compatibility with aqueous samples containing up to 10% organic solvents
Future Trends and Opportunities
Emerging directions include integrating non-suppressed IC with mass spectrometric detection, miniaturized column formats for faster analysis, eco-friendly eluents, and automated cleaning protocols. Advances in polymer-based stationary phases may further enhance selectivity and reduce run times.
Conclusion
The Shim-pack IC-A1 and IC-GA1 guard column combination provides a versatile and durable solution for non-suppressed ion chromatography. Adherence to recommended installation, operating, and maintenance procedures ensures consistent separation performance and extended column lifetime.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Shim-pack IC-C1/IC-GC1
2025|Shimadzu|Manuals
228-10323B Jan. 2025 High Performance Packed Column for HPLC Shim-pack IC-C1/IC-GC1 INSTRUCTION MANUAL 1. Introduction Shim-pack IC-C1/GC1 is a polymer gel based column for ion chromatography. Shim-pack IC-C1 is a cation exchange column optimized for non-suppressor ion chromatography. By choosing…
Key words
カラム, カラムイオン, イオンガードカラム, ガードカラムメイルナット, メイルナットイオンクロマトグラフィー, イオンクロマトグラフィーピーク, ピークpeek, peekcolumn, columnメンブランフィルター, メンブランフィルターノンサプレッサ, ノンサプレッサcleaning, cleaningグレード, グレードサイズ, サイズmultivalent, multivalenteluent
Shimadzu IC-A3/-GA3 HPLC Column
|Shimadzu|Manuals
228-30003D Feb. 2009 High Performance Packed Column for High Performance Liquid Chromatograph Shim-pack IC-A3/-GA3 Instruction Manual Introduction Shim-pack IC-A3 is an polymer gel based column with an anion exchange group for ion chromatography. It is optimized for non-suppressed ion…
Key words
カラム, カラムイオン, イオンガードカラム, ガードカラムピーク, ピークcolumn, columnpeek, peekイオンクロマトグラフィー, イオンクロマトグラフィーメイルナット, メイルナットエタノール, エタノールメタノール, メタノールノンサプレッサ, ノンサプレッサシリーズ, シリーズグレード, グレードアセトニトリル, アセトニトリルflushing
Shimadzu IC-SA4 HPLC Column
|Shimadzu|Manuals
228-91805A Jul.2023 Shimadzu High Performance Packed Column for HPLC Shim-pack IC-SA4 Instruction Manual NOTE: Compound retention times and peak shape may vary with usage. Before developing an analytical method with the column, verify that ■ Introduction Shim-pack IC-SA4 is a…
Key words
カラム, カラムイオン, イオンサプレッサ, サプレッサピーク, ピークガードフィルタ, ガードフィルタイオンクロマトグラフィー, イオンクロマトグラフィーメイルナット, メイルナットcolumn, columnpeek, peekflushing, flushingguard, guardexit, exitmay, maywaste, wasteパッケー
Shim-pack NovaCore C18-HB [Reversed phase column]
|Shimadzu|Manuals
日本語取扱説明書はP.4以降をご参照ください。 ERAD-0001-9087A High Performance Packed Column for HPLC Shim-pack NovaCore C18-HB [Reversed phase column] 取扱説明 書 INSTRUCTION MANUAL ■ Introduction NOTE Near the pH limit of use, please use a mobile phase containing at least 10% organic solvent. At the…
Key words
カラム, カラムシステム, システムイオンペア, イオンペアメイルナット, メイルナットピーク, ピークcolumn, columnthf, thfsus, susノイズ, ノイズベースラインドリフト, ベースラインドリフトprecautions, precautionsmobile, mobileチューブ, チューブphase, phaseflow